STOP 03
In event of non-compatibility of the nominal value
stored in the software for the pressure of the device
in range 1 - 5.
Contact the responsible Technical Service.
STOP 04
In event of non-compatibility of the maximum
voltage value for the device stored in the software
in the range 1 – 5.
Contact the responsible Technical Service.
In event of non-compatibility of the maximum
current value for the device stored in the
software in the range 1 – 5.
STOP 05
Contact the responsible Technical Service.
Exceeding of the limit value of the specified
voltage in the different ranges. Contact the
STOP 06
responsible Technical Service.
Exceeding of the limit value for the specified
current in the different ranges. This resulted in
overloading of the drive system. WAL
applicator system blocked or faulty.
STOP 07
First switch the device off and back on again via the
main switch.
If the fault message is displayed again
Change WAL applicator system
If the fault message is displayed once again
contact the responsible Technical Service.
CAUTION!
In the event of malfunctioning, operation of the device must be immediately terminated by
pressing the EMERGENCY OFF switch.
The manufacturer
Human Med AG
Wilhelm-Hennemann-Strasse 9
19061 Schwerin
Tel.: +49 /(0)385 / 3957011
Fax.: +49/(0)385 / 3957010
must be notified of the fault.
Review 2009/05/04