Download User manual
Transcript
RMC COMMAND DESCRIPTION TEXT INDEX RMC 003 Return the current of the used motors. PARAMETER DESCRIPTION # Digits Minimum Value Maximum Value 0 2 0 1 Set motor. 0 = RA, 1 = DE RETURN DESCRIPTION # Digits Value 0 2 0-255 PALOMAR identifier 1 4 0-2800 Current in use in mA DETAIL Command used to know the current in use on the selected motor. EXAMPLE RMC,0CR 0,2000CR OFF COMMAND TEXT OFF PARAMETER # Digits DESCRIPTION INDEX 004 Minimum Value Turn off the PALOMAR. DESCRIPTION Maximum Value RETURN DESCRIPTION # Digits Value 0 2 0-255 PALOMAR identifier 1 3 OFF DETAIL By this command you can turn off the PALOMAR by remote connection EXAMPLE OFFCR 0,OFFCR THS COMMAND DESCRIPTION TEXT INDEX THS 005 Temperature threshold to start on the fan PARAMETER DESCRIPTION # Digits Minimum Value Maximum Value 1 2 0 99 Temperature in degrees centigrade RETURN DESCRIPTION # Digits Value 0 2 0-255 PALOMAR identifier 1 2 CODE Operation result DETAIL Set the temperature threshold measured on the inner heat sink in the order to turn on the fan EXAMPLE THS,55CR 0,ACKCR 62 PALOMAR User Manual Rev. 1.5.1 – © DTA 2009-2010 All Rights Reserved