Finnigan GasBench Measurement Procedures for Real Samples II ________________________________________________ Water Equilibration (H/D) 5.7 Water Equilibration (H/D) H/D Equilibration in Brief Figure 5-29. Sample Preparation for H/D Equilibration In case of any hydrogen equilibration perform the following steps keeping the tray at room temperature (see Sample Tray Temperature Control on page 5-31): • Insert the sample into the vials and insert the catalytic platinum sticks. • Flush all samples with 2 % H2 in He. Run a flushing sequence. See Creating a New Sequence on page 3-34. The equilibration is finished within 40 min. It is not necessary to wait additional time. • Exchange the rinsing needle with the sampling needle. There are various needle sets using the same needle type: one set of needles exists for flushing (that is rinsing needle) and another one for measuring (that is sampling needle). The rinsing needle is used to rinse the vials: the recurrent capillary must be broken off at 20 cm to let the rinsing agent pass into ambient air. In case of the sampling needle, the recurrent capillary leads into GasBench II. • Run a mesurement sequence. See Creating a New Sequence on page 3-34. Thermo ____________________ Finnigan GasBench II Operating Manual __________________ 5-33 ELECTRON CORPORATION