Download SY14B Man Dec05.DOC - Hoffer Flow Controls, Inc
Transcript
8.3 UUT SYSTEM OPERATION CHECK Perform a complete test of the UUT transfer standard using the following procedure. 1. 2. 3. • Perform a visual inspection of all cabling. Examine flowmeter and/or replace the flowmeter bearings. Verify proper flowmeter pickup coil resistance with an ohm meter. Connect the CALIBRATOR and UUT system using standard trade practices and the figure below. SY14B-VR2-FM cable, connect from Transfer Standard Flowmeter to the UUT input of CALIBRATOR Hoffer Master Flowmeter connection UUT Transfer Standard Temperature probe connection Hoffer Master Temperature Probe connection CALIBRATOR UUT Transfer Standard Control Valve Fluid flow retrun to tank Hoffer Master Flowmeter Fluid flow from pump Transfer Standard Temperature Probe Hoffer Master Temperature Probe Transfer Standard Flowmeter Transfer Standard Blow-Down Valve. Figure 8-1 Connection diagram for Flowmeter Calibration • Turn the power on for both the CALIBRATOR and UUT systems. • On both the CALIBRATOR and UUT systems use the PRINT key at the label field “TEST TYPE” to select SENSOR CHECK. Press the START key on both systems to enter the SENSOR CHECK test mode. • On both systems use the scroll key to display PRINT SETUP. Press the PRINT keys on both systems to print the setups. Verify that the CALIBRATOR configuration matches the configuration of the UUT system. Pay particular attention to the UNIT OF MEASUREMENT and SYSTEM OF MEASUREMENT settings. If the configuration of both units do not match, enter the CONFIGURATION mode of the CALIBRATOR and set its configuration to match the configuration of the UUT. WARNING: If the SY-14B units and System of Measure option in the CONFIGURATION Mode do not match the display units of the UNIT UNDER TEST(UUT), inaccurate test results will occur. • SY-14B Continue with the next section SYSTEM COOL DOWN. 50 HP-220