38 and cannot be change. If you inadvertently trigger on noise, change to a higher scale. ERRATIC TRIGGERING? For a few detector heads, in electrically noisy environments, it is possible that the 842-PE will inadvertently trigger on the noise. In that case, increase the trigger level to 3% or higher if necessary. It is always good practice to reduce electrical noise generation or shield the detector and meter when measuring very low pulse energies. 2.3.3.6 Reference Values Comparisons of your power measurements to certain user-defined reference values are easy to make. If your current measurements lie within your established Upper and Lower bounds, then they meet the standards. If they go above or below the bounds, then there could be a quality problem, which triggers a fail message. The Reference Values option has two modes: Threshold mode and Pass/Fail mode. Display Scale Settings Ctrl ? Graphical Mode Full Display Mode Analog Out Reset Threshold Off Pass/Fail Fig. 2-14 Pass/Fail menu Threshold mode: Use this mode to set or change the upper and lower bounds, reference values and to restore the last values. To set or change a setting, use the arrow keys and scroll to Set Upper Bound, Set Lower Bound, Close, Set Reference Values, or Restore Last Values. Highlight your choice and press the ↵ Enter key. You may select Upper Bound only, Lower Bound only or both.