Sika TP281300
Operating manual
•
Operating the TP281300
2-wire circuited resistance thermometers measuring an
actual temperature.
Connect the resistance thermometer as shown below:
5.2.1 Testing temperature switch
Caution:
Burning heat: Let the TP281300 cool down after use
(see paragraph „Cool down heating block“).
Note:
To obtain an exact measuring result the specimen has to
fit properly in the bore of the heating block or of the
transition casing.
Prerequisites for exact results:
• The outer diameter of the test piece
matches the inner diameter of the
bore hole
(--->Sect. 3.4.1 "Transition casings").
• The calibrating instrument is in
vertical position.
• Unbended specimen.
• Insert the specimen in dry state
directly into the bore of the heating
block or in that one of the appropriate
transition casing.
• The measuring element of the sensor
is located in the region of the
homogeneous zone* of the calibrator
heating block.
Note:
Before starting the measurement the specimen must have
obtained the temperature of the block.
Give the specimen enough time for coming up to the block
temperature level.
3235004E.DOC
August 2009
21