Download DAWN HELEOS Hardware Manual
Transcript
Chapter 4: Using the Display Window 4.6.2 Audio Alarm To turn off, enable, or disable the audio alarm, display the Alarm Panel, Tab to the Audio Alarm field, then Enter to toggle the option. Note: Even when the audible alarm is turned off, the back panel alarm output will remain active. Here is a list of the alarms and their meanings: 4-12 • Overheat (not in Ambient version) is triggered if the read head ever exceeds 220°C • • • Vapor: Vapor sensor detected a leak. • • External: Back panel external alarm input is activated. • N2 pressure (not in Ambient version) is triggered when the temperature is set to less than 20°C, but the nitrogen pressure is less than 20psi. In this case, the alarm activates and resets the system temperature to 20.5°C. This prevents condensation from damaging the optics if the nitrogen connection is not made, or if the tank runs empty. • Flow Cell temperature lock shows if the flow cell temperature is locked Leak: Liquid sensor detected a leak. Recycle Open: Back panel recycle valve input requested the recycle valve be actuated Comet failed to actuate. This can occur when the cable is disconnected, or if the transducer needs replacement. MWH Rev. A