Malfunction
Probable reason
Corrective actions
Check the electrode cable’s
All (or one) of the ECG
The electrodes (or the
connection. Check the electrode
signals are not displayed on electrode of this lead) are
the screen.
not connected to the patient. cable’s fixation in the electrode.
PC-ECG module failure.
Check the connection between the
electrode and the patient.
If the malfunction is not corrected
contact the manufacturer
representative.
All (or one) of the ECG
signals are very noisy.
The electrodes (or one
electrode) are not greased
with the electroconductive
paste/gel.
Grease the electrodes (or one
needed electrode) with the
electroconductive paste/gel.
Operate the system in the
prescribed conditions.
Read the Warning about
conditions of system’s operating.