Section 4: Routine Maintenance and Calibration
Section 4: Routine Maintenance
and Calibration
This section reviews corrosion precautions, connector mating and
maintenance, plumbing air valve maintenance, conductivity cell storage and
cleaning, pressure sensor maintenance, sensor calibration, and TC Duct
disassembly. The SBE 9plus accuracy is sustained by care and calibration of
the sensors and establishment of proper handling practices.
Corrosion Precautions
Rinse the SBE 9plus with fresh water after use and prior to storage.
Aluminum Main Housing
The housing is insulated from the stainless steel guard cage and seacable
power circuits, preventing heavy corrosion of the housing. Avoid direct
attachment of metal objects to the housing.
All stainless steel screws that are exposed to salt water have been generously
lubricated with Blue MolyTM, a molybdenum lubricant containing nickel
power and zinc oxide. After each cruise, remove these screws and re-lubricate.
This compound is electrically conductive; use care to ensure it does not
get on PCBs.
The SBE 9plus has three large zinc anodes screwed into the main housing top
end cap. Check the anodes periodically to verify that they are securely
fastened and have not been eaten away.
Once a year, remove the stainless steel hose clamps and PVC mounting chocks
to clean the entire housing surface; this will prevent long-term breakdown of
the surface. We recommend periodic replacement of the clamps; use Teflon
tape or similar material between the clamps and the anodized housing surface.
Optional Titanium Main Housing
All exposed metal and hardware are titanium. No corrosion precautions are
required, but avoid direct electrical connection of the titanium to dissimilar
metal hardware.
Ring-shaped anode
Modular Sensors and Auxiliary Equipment
The modular temperature and conductivity sensors, as well as auxiliary
sensors manufactured by Sea-Bird that have an aluminum housing (SBE 18,
SBE 27, etc.), have a ring-shaped zinc anode. Check the anodes periodically to
verify that they are securely fastened and have not been eaten away.
Auxiliary equipment powered from the SBE 9plus should ideally have a
floating housing relative to their power circuits. It is permissible to have CTD
power and/or signal common lines connected to auxiliary equipment housing.
However, do not connect the auxiliary equipment housing to the 9plus 15-volt
power source, as this can destroy the conductivity cell electrodes.