Download SBE 9plus Manual
Transcript
Section 4: Routine Maintenance and Calibration Section 4: Routine Maintenance and Calibration This section reviews corrosion precautions, connector mating and maintenance, plumbing air valve maintenance, conductivity cell storage and cleaning, pressure sensor maintenance, sensor calibration, and TC Duct disassembly. The SBE 9plus accuracy is sustained by care and calibration of the sensors and establishment of proper handling practices. Corrosion Precautions Rinse the SBE 9plus with fresh water after use and prior to storage. Aluminum Main Housing The housing is insulated from the stainless steel guard cage and seacable power circuits, preventing heavy corrosion of the housing. Avoid direct attachment of metal objects to the housing. All stainless steel screws that are exposed to salt water have been generously lubricated with Blue MolyTM, a molybdenum lubricant containing nickel power and zinc oxide. After each cruise, remove these screws and re-lubricate. This compound is electrically conductive; use care to ensure it does not get on PCBs. The SBE 9plus has three large zinc anodes screwed into the main housing top end cap. Check the anodes periodically to verify that they are securely fastened and have not been eaten away. Once a year, remove the stainless steel hose clamps and PVC mounting chocks to clean the entire housing surface; this will prevent long-term breakdown of the surface. We recommend periodic replacement of the clamps; use Teflon tape or similar material between the clamps and the anodized housing surface. Optional Titanium Main Housing All exposed metal and hardware are titanium. No corrosion precautions are required, but avoid direct electrical connection of the titanium to dissimilar metal hardware. Ring-shaped anode Modular Sensors and Auxiliary Equipment The modular temperature and conductivity sensors, as well as auxiliary sensors manufactured by Sea-Bird that have an aluminum housing (SBE 18, SBE 27, etc.), have a ring-shaped zinc anode. Check the anodes periodically to verify that they are securely fastened and have not been eaten away. Auxiliary equipment powered from the SBE 9plus should ideally have a floating housing relative to their power circuits. It is permissible to have CTD power and/or signal common lines connected to auxiliary equipment housing. However, do not connect the auxiliary equipment housing to the 9plus 15-volt power source, as this can destroy the conductivity cell electrodes. 28