Introducing the Finnigan AQA
_______________________________________________________________ The Data System
Full Scan Mode
There are three different types of Full Scan acquisition. These are:
•
Centroid
•
Profile
•
MCA (Multi Channel Analysis)
In all full scan acquisitions raw data is collected over the whole scan range
defined by the start and end mass.
Centroid
During centroid acquisitions, Xcalibur automatically determines and saves
the mass center of the acquired profile peak returned from the detector.
Hence, the previously large number of points that described the mass
spectral peak are reduced to a single centroid stick (see Figure 1-26) for
each ion mass recorded. This has the advantage of reducing the quantity of
data recorded to the hard disk and improving processing speeds.
265
100
267
263
%
0
m/z
120
140
160
180
200
220
240
260
280
Figure 1-26. Full scan centroid spectrum of pentachlorophenol
_______________________ Finnigan AQA Hardware Manual _______________________ 1-35