Download TABLE OF REVISIONS - Med

Transcript
1.8.1
In the laboratory
When the platform is taken from a place of storage, it is essential to
check its correct operation in the laboratory before taking it to its
installation site.
1. Open the door and check by seeing the connections between the fixed
connectors on the side of the casing and the PM36 electronic card and
the memory unit.
2. Check the state of the fuses F1 and F2. These are not delayed fuses
marked 3.15 A.
3. Connect the sensors.
4. The platform is booted up by plugging the solar panel connector. Even
if there is no light sun, this connection allows the battery to supply the
electronics,
5. Check through the operator interface. The PM36 platform is ready to
be installed.
6. Sensors calibration : the user must verify the right calibration of its
sensors.
1.8.2
Calibration of sensors
After PM36 control, it is important to check the quality of the
measurements coming from sensors.
The user get from constructors specific datasheets on sensor and also a
calibration sheet.
Rain-gauge Qualimetrics 6011B :
The documentation includes :
- a user’s manual which describes the standard installation,
connection, calibration and maintenance of the sensor.
- a copy of the passed test record.
PM36 for Hydrology and Water Quality Meas.
User Manual
Document : 99D9009a.doc
Rev : A 28/39