The operating parameters are summarized below in Table 4.
Table 4.
3.1.4 Bladder Status
The bladder status indication portion of the device has been changed from using a
pressure transducer to a stretch sensor. This change was made because the stretch
sensor was readily available and was more ideal for the status indication. Cost
was also a deciding factor between the two status indication methods. Also using
the stretch sensor, in replacement of the pressure transducer reduces the amount
of power supply that would be needed to operate the device, and would eliminate
the need for signal amplification that was necessary to operate the pressure
transducer. Although the pressure transducer would give a more precise reading
of the bladder status regarding the accuracy of the volume percent of fluid in the
bladder, the stretch sensor would be sufficient as the brain itself does not usually
recognize the volume of the bladder until it is approximately 50 % full. A urine
bag shown below with stretch sensor placed across was used to measure the
resistance across the bag as it was filled with fluid for purposes of testing the
device and establishing ranges of resistance values to corresponding with each
volume percentage. Resistance ranges were measured with the urine bad was 0%,