Download User`s Manual - Particulate Systems
Transcript
Figure 4.46 Filling the Cell with Monitoring Particles Solution ・ Fill the monitoring particle solution slowly to prevent introducing air bubbles into the cell. ・ The Flat surface cell is coating the surface by Acrylamide. Since there is a possibility that coating may separate, a Flat surface cell recommends use at normal temperature. Washing the Flat Surface Cell Be careful not to damage the Flat Surface Cell during handling. Do not wash the cell in an ultrasonic cleaner. Wash the cell as described below. To wash the Flat Surface Cell: 1. Wash the Flat Surface Cell with purified water. Do not wash the cell in an alkaline solution. This may cause the coating to peel. 2. Wrap the cell with lens paper, and store it in purified water. 3. Wipe off water drops around the glass cell with lens paper. 4. When washing parts other than the Flat Surface Cell, such as the O-rings, immerse the parts in a neutral detergent and insert into an ultrasonic washer. Rinse them well with purified water to remove the detergent completely. 146