Download User Manual - Stat-Med
Transcript
ZIPPERS AND PULL-TABS • Inspect zipper coils for any kinks or misalignment (See Fig. 25). Test that all zippers open easily and close securely and there are no gaps or openings when pressure is applied along the entire length of each zipper. If a zipper has been damaged (See Fig. 26A), it may not fully close. • ALWAYS use the zipper pull-tabs when opening and closing panels (See Fig. 27). NEVER rip the panels open, as this may damage the fabric or the zippers (See Fig. 27A). • NEVER use the bed if a zipper slider is damaged or the pull-tab is missing or broken. Call Posey at 1.800.44.POSEY (1.800.447.6739) for repair instructions. • NEVER use the bed if a zipper coil is kinked or misaligned. Crown (Tab) Holder Fig. 25. Kinks in zipper Opened Up PullTab Body Fig. 26. Undamaged Zipper Slider Fig. 26A. Zipper Slider Damaged from ripping the access panel open Fig. 27. ALWAYS use the zipper pull-tabs when opening and closing the access panels Fig. 27A. Do.2 NOT rip panels open as this will damage the zipper slider (See Fig. 26A) and prevent the zipper from closing securely NEVER leave the patient’s bedside until all access panel zippers are securely closed. Test the entire length of each zipper by pressing against the panel near the zipper to make sure it is securely closed and the access panel will not open when pressure is applied (See Fig. 28). A failure to heed this warning could result in serious injury or death from a fall or unassisted bed exit. See “Check the Zippers” section, Page 10. Fig. 28. ALWAYS test that the zipper is securely closed by pressing against the entire length of the closed zipper to make sure it will not open when pressure is applied 22 Posey Bed 8060 2