alignment of the wavefront sensor optics (with the field stop closed down enough to
create a point source at the input of the WFS), masking of Rayleigh light or close
binaries, and masking of ghosts or stray light (with the field stop set at the size of
the subaperture, 6 arcseconds). The grid of the lenslet array/detector subapertures
must be aligned with the grid of MEMs actuators in the AO relay.
The WFS mount shown in Figure 48 meets all of these requirements. The bottom
linear stage provides focus adjustment for the entire wavefront sensor package. The
linear stage perpendicular to the bottom stage allows changing between two (with
room for an upgrade to three) subaperture modes. Each of the three optical relays
sits in v-grooves that travel perpendicular to the optical axis to center the relay
between the iris and detector. The iris is a motorized, adjustable iris that can close
down completely. The Scimeasure “Big Joe” wavefront sensor camera is mounted
within a flexure mount and can move small amounts in the x and y directions
through the adjustment of New Focus picomotors to align the detector with the
lenslet array. The entire package is mounted on feet that allow adjustment in height
and tip/tilt, and nudgers provide adjustment in x (parallel to the bench) to allow
alignment with the MEMs grid.
Figure 48 The Shane AO WFS mechanical package.