Download User Manual, AED10 Automatic External
Transcript
61 6 Troubleshooting This chapter provides information on how to troubleshoot situations and conditions that arise during the operation of the Welch Allyn AED 10 and gives answers to frequently asked questions. Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Attaching Electrode Pads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Analyzing Interrupted. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . No Shock Delivered . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Defibrillator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Other Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Frequently Asked Questions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61 61 62 62 63 63 63 64 Overview If the built-in sensors in the Welch Allyn AED 10 detect a problem prior to or during operation, the unit provides a voice or screen prompt indicating the problem. Use the information in the following tables to troubleshoot. The tables list fault indicators and determine possible corrective actions. Note In some situations, the operator will be instructed to change the battery or defibrillation electrode pads. It is important to always have spare batteries and other accessories available. Attaching Electrode Pads Indicator/Possible Cause Corrective Action Electrode pad does not adhere properly to the patient. Wipe moisture from chest and/or shave excessive hair from chest. Electrode pads are dry, damaged, or out-of-date. Replace the electrode pads. Improper pad placement or pads touching each other. Check pad placement; make sure pads are in the correct location. Inadequate connection to AED 10 Check for proper insertion of defibrillation pads connector into AED 10. Pads connector not connected or properly inserted into the connector socket Push pads connector firmly into the connector socket. Poor defibrillation pad contact with the patient’s bare chest. Press firmly on electrode pads to improve adherence to patient's skin.