Download Model 2001 Pulse Oximeter User`s Manual
Transcript
SpO2 Sensors Y-Sensor 2. If you are using an adhesive wrap, remove both sides of the release liner. Face the blue side of the wrap toward the skin and wrap around the site (Velcro tab may be removed and replaced to allow excess foam to be cut as necessary). Secure with the Velcro® tab. WARNING: Do not wrap the tape around the limb so tightly that circulation is restricted. Inspect the site often for adequate circulation—at least once every four hours. When applying sensors take note of patient’s physiological condition. For example, burn patients may exhibit more sensitivity to heat and pressure and therefore additional consideration such as more frequent site checks may be appropriate. 3. Ensure the sensor heads are directly opposite each other through the tissue. This prevents the sensor from being placed on a site too thick for proper operation. Neonatal hand Neonatal/pediatric foot Pediatric toe Adult/pediatric finger Adult toe WARNING: Treat foam wrap in accordance with hospital protocol for single-patient use. Check site regularly to ensure adequate circulation and proper sensor positioning. Y-Sensor Application using the Ear Clip 1. Remove center strip from the Y-Sensor. Sensor heads Center strip (remove) Strip removed Rev. 01 Model 2001 User’s Manual 25