7.3. Manual Self-Test of the Alarm System You can manually self-test test the alarm system by the following steps. Select a monitor that is being monitored at the NTS and bring it up in the large view area on the NTS Turn SpO2 alarms on (via the Parameter Setup Menu) At the monitor, unplug the SpO2 sensor After 10 seconds, the NTS should sound a low grade alarm tone, flash the SpO2 parameter box (white), and display the message “SpO2 unplugged” Note: If using the Direct Connection option, you must first connect the VitalPoint® PRO monitor to the laptop running the NTS software using the Direct Connect Cable part number VS-940. VitalPoint® NOMAD™ Telemetry System User's Guide DL-01008 Rev.1 7-16