Download 7025 User Manual
Transcript
Specifications and Environmental Conditions 2 2.1 Specifications and Environmental Conditions • temperature range: 5–50°C (41–122°F) Specifications • ambient relative humidity: 15–50% See Table 1. 2.2 • pressure: 75kPa - 106kPa Environmental Conditions Although the instrument has been designed for optimum durability and trouble-free operation, it must be handled with care. The instrument should not be operated in an excessively dusty or dirty environment. Maintenance and cleaning recommendations can be found in the Maintenance Section of this manual. The instrument operates safely under the following conditions: • mains voltage within ±10% of nominal • vibrations in the calibration environment should be minimized • altitude does not effect the performance or safety of the unit 2.3 Warranty The 7025 bath is covered by a 2 year warranty that takes effect 10 days after the product is shipped. The manufacturer will provide parts and labor without charge for repair or replacement of the instrument due Table 1 Specifications Operating Range: Temperature Stability: Temperature Gradients: Set-point Accuracy: –20 to 110°C ±0.005°C or better ±0.01°C max between any two points in the work area ±0.5°C or better ±5°C Cut-out Accuracy: Exterior Dimensions: Width: 20.2 inches Front to back: 15.88 inches Height: 12.2 inches to working surface, 15.5 inches to top of stirring motor Power Requirements: 115 VAC (±10%), single phase, 60 Hz [230 VAC (±10%), 50 Hz Optional] Weight: 55 lbs Fluid Volume: Approximately 2.1 gallons Working Area: 5 3/4" W x 3 1/4" Front to Back x 7 1/2" Deep 10 Manual Rev. 722602 Hart Scientific