NIBP
CAS 750 Monitor
Warning: In some cases, rapid, prolonged cycling of an oscillometric, noninvasive blood pressure monitor cuff has been associated with any or all of the following: ischemia, purpura, or neuropathy. Apply the oscillometric cuff appropriately, according to
instructions, and check the cuff site and cuffed extremity regularly
when blood pressure is measured at frequent intervals or over extended periods of time.
Entering STAT Mode
To activate the STAT mode, proceed as follows:
•
Enter the NIBP Menu as described in this section on page 79.
Note: The monitor must be in the Manual NIBP Cycle mode when
selecting STAT mode.
•
Select an alternate Initial Inflation Pressure if necessary.
•
Move the cursor to STAT and then select ON.
•
Return to the Main Screen by pressing the RETURN key. The
message “STAT” appears in the NIBP window.
•
Press the NIBP key to begin the first measurement.
Note: If the Initial Inflation Pressure was altered, the first measurement will be made using that setting. After the first measurement, the monitor will switch to the Adaptive mode and adapt the
cuff inflation pressure automatically based on the previously
measured systolic value (approximately 30 mmHg above).
Between readings, the NIBP window will display the remaining
seconds until the next measurement.
Note: Selecting STAT mode will override any time interval selected for the NIBP measurement cycle.
Exiting STAT Mode
The monitor will stop the STAT mode automatically after 5 minutes
and returns to the Manual NIBP mode.
Page 78
•
Press the NIBP key to terminate the current NIBP measurement.
•
Enter the NIBP Menu as described in this section on page 79.
•
Select STAT mode OFF.
21-02-0289 REV. 00 09/06