CAS 750 Monitor
Appendix
Power
External Power:
AC: 100-240 VAC, 50/60 Hz, 0.5 A
Fuse Rating – T500mAL250V or
T1.25AL250V (two provided)
Refer to monitor labeling for actual fuse rating
DC (optional): +12 VDC; 7W
Fuse Rating – Littelfuse Type 154, Slo-Blo,
5.0A, 125VAC or approved equivalent (one
provided)
Chassis Leakage Current: 100 µA (maximum)
Battery:
NiMH battery pack (user removable).
Charge Time: 3-5 hours
Operating Time: 3 hrs (minimum)
Serial Interface
Interface Type:
Speed:
Signal Level:
Data Length:
Start Bit:
Stop Bit:
Parity:
Flow Control:
Bi-directional Serial Communication
9600 for Printer
115200 for CAS Serial Protocol
RS232C
8 bits
1 bit
1 bit
None
None
Nurse Call Interface
Relay contacts:
Max. Switching Power:
Max. Delay:
Floating, one opening and one closing contact.
0.3 A at 120 VAC or 1.0 A at 30 VDC
< 0.5 seconds.
Standards
Units comply with the following requirements:
ANSI/AAMI SP-10: 2002
CE marking according to Directive 93/42/EEC
IEC 60601-1, EN 60601-1-2
IEC 60601-2-27, IEC 60601-2-30, IEC 60601-20-49
EN 864, EN 865
UL classified – UL 60601-1, CAN/CSA C22.2 No. 60601.1
All units covered by U.S. patent 4,796,184 and 5,022,403. Other patents
pending.
21-02-0289 REV. 00 09/06