Download USER-MANUAL - Tamson
Transcript
9.2 Code E61 Message on display What does it mean Defect / calibration PT100 E70 E50 E82 E4 E5 E6 What action to take Check wiring and element. Ohmic value at 0°C=100R and at 100°C=138R. Replace PT100 Error on operational parameters EEprom error Error on input during calibration EEprom error EEprom error Watchdog error Auto - reset after 2 s Auto - reset after 2 s Auto - reset after 2 s Auto - reset after 2 s 10 SPECIFICATION 10.1 Technical specifications overview Item Ordering code 230V 115V Power Materials Range Reading Setting ± Stability ± Heating Bath volume Pump Pump Opening Depth bath Length Width Height Weight CE Unit 50 or 60Hz 50 or 60Hz [Watt] [°] [°K] [W] [L] TC16 00T0671 TC40 00T0681 TC58 00T0691 00T0861 00T0851 00T0881 1600 2800 2800 Used inside bath: stainless steel 304, PTFE Ambient .. 250°C/..482°F Standard °C, °F on request 0.1 0.02 0.02 0.02 1200** 2700 2700 16 40 58 [mBar] [L/min] [mm] [mm] Max pressure 300* Max flow 7.5* 180 *210 420 *275 200 200 [mm] [mm] [mm] [Kg] 455 705 295 375 440 440 21 30 Conforms to CE regulation 420 *275 300 705 375 590 35 * Optional pump with pressure max 1bar, flow max 16ltrs/min, reference: 24T0392. See specification sheet "pumpgraph ** Additional heater 1500W optional to heat larger loads when circualting Tamson Instruments B.V. P.M. Tamson Instruments bv van 't Hoffstraat 12 2665 JL Bleiswijk THE NETHERLANDS Rev. 3.0 UK 10-2010 T. 31 (0) 10 522 43 73 F. 31 (0) 10 521 19 41 www.tamson.com [email protected] VAT: NL 80 66 34 984 B01 Bank account no.: Rabobank 16 01 00 046 Chamber of commerce 27 16 95 41 IBAN Code: NL95 RABO 0160100046 Page 18/19