Manual revision 006
Notes:
• The DC and GetCC responses
contain the same information, but in
different formats.
• Dates shown are when calibrations
were performed.
Appendix III: Command Summary and Details
HydroCAT (SDI-12 & RS-232)
Status Commands (continued)
DC
Display calibration coefficients, which are
initially factory-set and should agree with
Calibration Certificates shipped with
HydroCAT. Optional oxygen sensor calibration
coefficients in response are coefficients
programmed into integrated Hydro-DO.
Example: HydroCAT with a pressure and oxygen sensors (user input in bold, command used to modify parameter in
parentheses).
DC
HydroCAT-SDI12 V2.13.0 32345
temperature: 04-Aug-15
TA0 = 6.947802e-05
TA1 = 2.615233e-04
TA2 = -1.265233e-06
TA3 = 1.310479e-07
conductivity: 04-Aug-15
G = -1.036689e+00
H = 1.444342e-01
I = -3.112137e-04
J = 3.005941e-05
CPCOR = -9.570001e-08
CTCOR = 3.250000e-06
WBOTC = 1.968100e-05
pressure S/N 2478619, range = 2901 psia, 03-Aug-15
PA0 = 0.000000e+00
PA1 = 0.000000e+00
PA2 = 0.000000e+00
PTCA0 = 0.000000e+00
PTCA1 = 0.000000e+00
PTCA2 = 0.000000e+00
PTCB0 = 0.000000e+00
PTCB1 = 0.000000e+00
PTCB2 = 0.000000e+00
PTEMPA0 = 0.000000e+00
PTEMPA1 = 0.000000e+00
PTEMPA2 = 0.000000e+00
POFFSET = 0.000000e+00
oxygen S/N 12, 28-Aug-15
TAU_20 = 4.000000e+00
OXA0 = 1.051300e+00
OXA1 = -1.500000e-03
OXA2 = 4.161926e-01
OXB0 = -2.325492e-01
OXB1 = 1.692931e+00
OXC0 = 8.966704e-02
OXC1 = 3.617471e-03
OXC2 = 5.112384e-05
OXTA0 = 6.517293e-04
OXTA1 = 2.533749e-04
OXTA2 = 3.140482e-07
OXTA3 = 1.064506e-07
OXE = 1.100000e-02
71
[TCalDate=]
[TA0=]
[TA1=]
[TA2=]
[TA3=]
[CCalDate=]
[CG=]
[CH=]
[CI=]
[CJ=]
[CPCor=]
[CTCor=]
[CWBOTC=]
[PRange= (psi), PCalDate=]
[PA0=]
[PA1=]
[PA2=]
[PTCA0=]
[PTCA1=]
[PTCA2=]
[PTCB0=]
[PTCB1=]
[PTCB2=]
[PTempA0=]
[PTempA1=]
[PTempA2=]
[POffset= (decibars)]
[OxCalDate=]
[OxTau20=]
[OxA0=]
[OxA1=]
[OxA2=]
[OxB0=]
[OxB1=]
[OxC0=]
[OxC1=]
[OxC2=]
[OxTA0=]
[OxTA1=]
[OxTA2=]
[OxTA3=]
[OxE=]