Download PsimetriQ manual
Transcript
PPA55xx “KinetiQ” user manual lighting ballasts can be easily addressed with special modes application in the firmware of the instrument and dedicated hardware functions including torque/speed inputs and ac line sync. The whole operation of the instrument may be controlled remotely via a serial interface (RS232/USB), or optionally LAN interface or GPIB interface. The voltage and current channels are identical except for the voltage attenuator and current shunt at the very front end. Each channel consists of a selection switch for external or internal attenuator/shunt followed by a high impedance buffer, bandwidth control then a series of gain stages leading to an A/D converter. Selection of the input gain and the sampling of the A/D converter are under the control of the DSP. There is an autozero switch at the front end for dc accuracy. The analogue circuitry is optimised for high linearity over a wide dynamic range and high frequency performance. Both input channels are fully isolated with very good CMRR and noise rejection. The current shunt is of a proprietary design which gives very wide bandwidth with minimum phase shift. The voltage attenuator is of a proprietary design which has a wide bandwidth response matched to that of the current shunt. Both the voltage and current channels are calibrated digitally so there are no physical adjustments to be made. 1-2