Download - Bio-Rad
Transcript
SYSTEM OPERATION MODES OF OPERATION Table 7-1. Valve Setup Information Valve Type Valve Name/ Function Position Names Notes SVT3-2 Fraction Collector Low pressure Diverter solenoid valve 1 Waste 2 Collect Functions as a fraction collection diverter, determined by the actual fraction collection parameters chosen. Note: Indicate the location of the valve cable connection. Inlet A Named by user Inlet B Named by user When used before the inlet to Pump A or B, the valve enables buffer selection. The buffer name specified for each position will appear in the Protocol screen’s “Isocratic Flow”, “Linear Gradient”, and “Change Valve” dialog box. Refer to Chapter 8, Sample Loading, for examples. When the Maximizer is Aux Pump Inlet used, inlets A and B become A1, A2, B1, B2. User Assigned Name Named by user Used for auxiliary pump load selection to select one of two solutions. Refer to Chapter 8, Sample Loading, for examples. Named by user When used for a purpose other than described above. The name specified for each position will appear in the Protocol screen’s “Change Valve” dialog box. Named by user When used before the inlet to Pump A or B, the valve provides preparative sample loading or buffer selection. The buffer or sample name specified for each position will appear in the Protocol screen’s “Isocratic Flow”, “Linear Gradient”, and “Change Valve” dialog box. Refer to Chapter 8, Sample Loading, for examples. SV5-4 Inlet A Low pressure solenoid valve Inlet B Named by user Note: Indicate the location of the valve cable connection. Aux Pump Inlet Named by user When the Maximizer is used, inlets A User Assigned and B become Name A1, A2, B1, B2. Used for auxiliary pump load selection to select one of four solutions. Refer to Chapter 8, Sample Loading, for examples. Named by user When used for a purpose other than that described above. The name specified for each position will appear in the Protocol screen’s “Change Valve” dialog box. AVR7-3 Sample Inject High pressure valve 1 Load Sample 2 Inject Sample 3 Purge For automatically loading a sample. Note: Indicate the location of the valve cable connection Named by user When used for a purpose other than described above. The name specified for each position will appear in the Protocol screen’s “Change Valve” dialog box. Refer to Section 4, Advanced System Applications, Chapters 8 through 10. User Assigned Name 7-7