MVE Biological Systems TEC 3000 Technical Freezer Manual 7.2.2.2. Enable / Disable Auto Fill Control Auto Fill Control can be enabled or disabled in the Advanced Level Menus. NOTE: Security Level 3 or higher is required to enable/disable Auto Fill. 1. Press “SETUP” Controller will prompt for a password. A flashing cursor on the entry screen will make it clear which digit is being changed. Use the “▲/▼” keys to scroll to the appropriate number. Press “ENTER” to advance the cursor to the next position. 2. Press “ENTER” The display will read “Temperature Menus”. 3. Press “SETUP” Until the display reads “Liquid Level Menus”. 4. Press “ENTER” The display will read “High Level Alarm”. 5. Press “SETUP” Until the display reads “Advanced Level Menus”. 6. Press “ENTER” The display will read Auto Fill Control”. Use the “▲/▼” keys to adjust and press “ENTER” to save the change. Higher User Level Required use ▲▼ to ENTER Password 0000 Press ENTER for Temperature Menus or press SETUP for next menu Press ENTER for Liquid Level menus or press SETUP for next menu High Level Alarm 8.0 in Use ▲▼ to adjust Press ENTER to save Press ENTER for Advanced Level menus or press SETUP for next menu Auto Fill Control ENABLED Use ▲▼ to adjust Press ENTER to save 13289499 Rev F • 53