Download OM 402PWR - Orbit Merret

Transcript
GUARANTEE
Y E A R S
OM 402PWR
4 DIGIT PROGRAMMABLE
AC VOLTMETER/AMMETER
NETS ANALYSER
WATTMETER
SAFETY INSTRUCTIONS
Please, read the enclosed safety instructions carefully and observe them!
These instruments should be safeguarded by isolated or common fuses (breakers)!
For safety information the EN 61 010-1 + A2 standard must be observed.
This instrument is not explosion-safe!
TECHNICAL DATA
Measuring instruments of the OM 402 series conform to the European regulation 89/336/EWG and the Ordinance
168/1997 Coll.
The instruments are up to the following European standards:
EN 55 022, class B
EN 61000-4-2, -4, -5, -6, -8, -9, -10, -11
The instruments are applicable for unlimited use in agricultural and industrial areas.
CONNECTION
Supply of energy from the main line has to be isolated from the measuring leads.
ORBIT MERRET, spol. s r.o.
Vodnanska 675/30
198 00 Prague 9
Czech Republic
Tel: +420 - 281 040 200
Fax: +420 - 281 040 299
e-mail: [email protected]
www.orbit.merret.cz
2
|
INSTRUCTIONS FOR USE OM 402PWR
CONTENTS
1.
2.
3.
4.
Contents
..........................................................................................3
Instrument description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Instrument connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Instrument setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Symbols used in the instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Setting the DP and the (-) sign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Control keys function
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Setting/permitting items into “USER” menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
5.
Setting “LIGHT” menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.0 Description “LIGHT” menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Input to „LIGHT“ menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Setting „Channel I“ - Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Setting „Channel U“ - Voltage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Setting „Channel P“ - Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Setting mathematic functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Setting Limits
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Setting analog output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Selection of programming menu „LIGHT“/„PROFI“ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Restoration of manufacture setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Selection of instrument menu language version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Setting new access password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Instrument identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
6.
Setting “PROFI” menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.0 Description of “PROFI” menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.1
6.2
6.3
6.4
“PROFI” menu - INPUT
6.1.1
Resetting internal values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.1.2
Setting measuring type, range, mode, rate, ... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.1.3
Setting the Real Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.1.4
External input function selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.1.5
Optional accessory functions of the keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
“PROFI” menu - CHANNELS
6.2.1
Setting measuring parameters for „Channel I“ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.2.2 Setting measuring parameters for „Channel U“ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.2.3 Setting measuring parameters for „Channel P“ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.2.4 Setting measuring parameters for „Channel F“ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.2.5 Setting mathematic functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.2.6 Selection of evaluation of min/max. value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
“PROFI” menu - OUTPUT
6.3.1
Setting data logging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3.2 Setting Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3.3 Setting data output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3.4 Setting analog output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3.5 Selection of display projection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
“PROFI” menu - SERVICE
6.4.1
Selection of programming menu „LIGHT“/„PROFI“ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.4.2 Restoration manufacture setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.4.3 Selection of instrument menu language version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.4.4 Setting new access password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.4.5 Instrument identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1
32
33
35
35
37
42
45
46
46
47
51
52
54
56
58
60
62
63
64
64
64
7.
Setting items into “USER” menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
7.0 Configuration “USER” menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
8.
9.
10.
11.
12.
13.
Data protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Error statements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Table of symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Instrument dimensions and instalation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Certificate of guarantee . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
INSTRUCTIONS FOR USE OM 402PWR
|
3
2
INSTRUMENT DESCRIPTION
2.1
Description
The OM 402PWR model is a universal 4-digit panel wattmeter with independent measuring of AC voltage/current, frequency,
power factor and other quantities.
The instrument is based on an 8-bit microcontroller and precise RMS converter, which secures high accuracy, stability and
easy operation of the instrument.
PROGRAMMABLE PROJECTION
Measured quantities
with calculation
Measuring range:
Setting:
Projection:
voltage (VRMS)
current (ARMS)
active power (P)
frequency (Hz)
idle power (Q)
apparent power (S)
power factor (cos ϕ)
adjustable as fixed or with automatic change
manual, it is possible to set arbitrary projection on the display for maximum value of the input signal
in the menu, e.g. input 0...250 V/0...5 A 0...1.500 kW
-9999…9999
LINEARIZATION
Linearization:
by linear interpolation in 50 points (solely via OM Link)
DIGITAL FILTERS
Exponen.average:
Rounding:
from 2…100 measurements
setting the projection step for display
MATHEMATIC FUCTIONS
Min/max. value:
Tare:
Peak value:
Mat. operations:
registration of min./max. value reached during measurement
designed to reset display upon non-zero input signal
the display shows only max. or min. value
polynome, 1/x, logarithm, exponential, power, root, sin x
EXTERNAL CONTROL
Lock:
Hold:
Tare:
Resetting MM:
Memory:
4
|
control keys blocking
display/instrument blocking
tare activation/resetting tare to zero
resetting min/max value
data storage into instrument memory
INSTRUCTIONS FOR USE OM 402PWR
INSTRUMENT DESCRIPTION
2.2
2
Operationn
The instrument is set and controlled by five control keys located on the front panel. All programmable settings of the instrument
are performed in three adjusting modes:
LIGHT
Simple programming menu
- contains solely items necessary for instrument setting and is protected by optional number code
PROFI
Complete programming menu
- contains complete instrument menu and is protected by optional number code
USER
User programming menu
- may contain arbitrary items selected from the programming menu (LIGHT/PROFI), which determine
the right (see or change)
- acces without password
All programmable parameters are stored in the EEPROM memory (they hold even after the instrument is switched off).
Complete instrument operation and setting may be performed via OM Link communication interface,
which is a standard equipment of all instruments.
The operation program is freely accessible (www.orbit.merret.cz) and the only requirement is the
purchase of OML cable to connect the instrument to PC. It is manufactured in version RS 232 and USB and is compatible
with all ORBIT MERRET instruments. Another option for connection is with the aid of data output RS 232 or RS 485 (without
the need of the OML cable).
The program OM LINK in „Basic“ version will enable you to connect one instrument with the option of visualization and
archiving in PC. The OM Link „Standard“ version has no limitation of the number of instruments connected.
2.3
Options
Comparators are assigned to monitor one, two, three or four limit values with relay output. The user may select limits regime:
LIMIT/DOSING/FROM-TO. The limits have adjustable hysteresis within the full range of the display as well as selectable
delay of the switch-on in the range of 0...99,9 s. Reaching the preset limits is signalled by LED and simultaneously by the
switch-on of the relevant relay.
Data outputs are for their rate and accuracy suitable for transmission of the measured data for further projection or directly
into the control systems. We offer an isolated RS232 and RS485 with the ASCII or DIN MessBus protocol.
Analog outputs will find their place in applications where further evaluating or processing of measured data is required in
external devices. We offer universal analog output with the option of selection of the type of output - voltage/current. The
value of analog output corresponds with the displayed data and its type and range are selectable in Menu.
Measured data record is an internal time control of data collection. It is suitable where it is necessary to register measured
values. Two modes may be used. FAST is designed for fast storage (40 records/s) of all measured values up to 8 000
records. Second mode is RTC, where data record is governed by Real Time with data storage in a selected time segment
and cycle. Up to 250 000 values may be stored in the instrument memory. Data transmis sion into PC via serial interface
RS232/485 and OM Link.
INSTRUCTIONS FOR USE OM 402PWR
|
5
3
INSTRUMENT CONNECTION
The instrument supply leads should not be in proximity of the incoming low-potential signals.
Contactors, motors with larger input power should not be in proximity of the instrument.
The leads into the instrument input (measured quantity) should be in sufficient distance from all power leads and appliances.
Provided this cannot be secured it is necessary to use shielded leads with connection to ground (bracket E).
The instruments are tested in compliance with standards for use in industrial area, yet we recommend to abide by the above
mentioned principles.
1
2
+
-
3
4
5
5
6
7
E
D
Input 2 - U
C
GND - U
B
GND - I
A
Input 1 - I
III
GND
AO-I
AO-U
IV V VI
11 12 13
I
10
Input 3 - U
II
9
L4
L3
8
RxD/L+
TxD/LGND
14 15 16 17
L2
L1
POWER SUPPLY
EXT. 1
EXT. 2
EXT. 3
N
E
L
!
Clips GND - U and GND - I have galvanic connection
MEASURING RANGES
Range/Input
Input 1 - I
Range/Input
Input 2 - U
Input 3 - U
Range “K”
0…60/150/300 mV
Range “S”
0…10 V
0…120 V
Range “P”
0…1 (2,5)/5 A
Range “U”
0…250 V
0…450 V
6
|
INSTRUCTIONS FOR USE OM 402PWR
INSTRUMENT CONNECTION
3
E
D
Input 2 - U
C
GND - U
GND - I
Input 1 - I
III
Input 3 - U
B
11 12 13
A
10
I
9
II
8
IV V VI
CONNECTION FOR MEASURING ON ONE PHASE
1
2
3
4
5
5
6
7
L
N
LOAD
14 15 16 17
E
D
Input 2 - U
C
GND - U
GND - I
Input 1 - I
III
Input 3 - U
B
11 12 13
A
10
I
9
II
8
IV V VI
CONNECTION FOR MEASURING ON ONE PHASE
1
2
3
4
5
5
6
7
L1
LOAD
14 15 16 17
LOAD
L2
LOAD
L3
N
!
In case of connecting clips GND - U or GND - I to a
phase, on this potential is subsequently the connector OM Link and secondary inputs.
It is necessary to abide by the precautions of hazardous contact with active parts of the instrument in
compliance with relevant standards and regulations
INSTRUCTIONS FOR USE OM 402PWR
|
7
4
INSTRUMENT SETTING
• For expert users
• Complete instrument menu
PROFI
• Access is password protected
Setting
• Possibility to arrange items of the
„User“ menu
• Tree menu structure
• For trained users
LIGHT
• Only items necessary for instrument
setting
• Access is password protected
Setting
• Possibility to arrange items of the
„User“ menu
• Linear menu structure
USER
• For user operation
• Menu items are set by the user
(Profi/Light) as per request
Setting
• Access is not password protected
• Optional menu structure either tree
(PROFI) or linear (LIGHT)
8
|
INSTRUCTIONS FOR USE OM 402PWR
INSTRUMENT SETTING
4.1
4
Setting
The instrument is set and controlled by five control keys located on the front panel. All programmable settings of the instrument
are performed in three adjusting modes:
LIGHT
Simple programming menu
- contains solely items necessary for instrument setting and is protected by optional number code
PROFI
Complete programming menu
- contains complete instrument menu and is protected by optional number code
USER
User programming menu
- may contain arbitrary items selected from the programming menu (LIGHT/PROFI), which determine
the right (see or change)
- acces without password
All programmable parameters are stored in the EEPROM memory (they hold even after the instrument is switched off).
Complete instrument operation and setting may be performed via OM Link communication interface, which is a standard
equipment of all instruments.
The operation program is freely accessible (www.orbit.merret.cz) and the only requirement is the purchase of OML cable
to connect the instrument to PC. It is manufactured in version RS 232 and USB and is compatible with all ORBIT MERRET
instruments.
Another option for connection is with the aid of data output RS 232 or RS 485 (without the need of the OML cable).
Scheme of processing the measured signal
Channel I
Input I
Converter
Input U
Min/Max
FILTER
Linearization
Channel U
Relays
FILTER
Linearization
FILTER
Linearization
FILTER
Linearization
Channel P
Analog
Channel F
Data
Memory
MAT.FCE.
Linearization
Display
INSTRUCTIONS FOR USE OM 402PWR
|
9
4
INSTRUMENT SETTING
Setting and controlling the instrument is performed by means of 5 control keys located on the front panel. With the aid of
these keys it is possble to browse through the operation menu and to select and set required values.
Measured value (red/green LED)
Indication of channel (green LED)
I
U
P
F
Current
Voltage
Power
Frequency
the last two places may display
measuring units
Ralay status (red LED)
ON the digit is lit
OFF the digit is not lit
OFF the digit is flashing
limits with restriction
(hysteresis, delay)
Function (green LED)
I
U
P
F
1
2
85.42
3
4
M
M
T
Min/max. value
Tare
T
Symbols used in the instructions
values preset from manufacture
DEF
42
MIN
CONECT.
30
symbol indicates a flashing light (symbol)
inverted triangle indicates the item that can be placed in USER menu
broken line indicates a dynamic item, i.e. it is displayed only in particular selection/version
after pressing the key the set value will not be stored
after pressing the key the set value will be stored
continues on page 30
Setting the decimal point and the minus sign
DECIMAL POINT
Its selection in the menu, upon modification of the number to be adjusted it is performed by the control key
beyond the highest decade, when the decimal point starts flashing . Positioning is performed by / .
with transition
THE MINUS SIGN
Setting the minus sign is performed by the key on higher decade. When editing the item substraction must be made from
the current number (e.g..: 013 > , on class 100 > -87)
10
|
INSTRUCTIONS FOR USE OM 402PWR
INSTRUMENT SETTING
4
Control keys functions
Key
Measurement
Menu
Setting numbers/selection
access into USER menu
exit menu
quit editing
programmable key function
back to previous level
move to higher decade
programmable key function
move to previous item
move down
programmable key function
move to next item
move up
programmable key function
confirm selection
confirm setting/selection
numeric value is set to zero
+
+
access into LIGHT/PROFI menu
+
direct access into PROFI menu
+
configuration of an item for
“USER” menu
+
determine the sequence of items
in “USER - LIGHT” menu
Setting items into „USER“ menu
• in LIGHT or PROFI menu
• no items permitted in USER menu from manufacture
• on items marked by inverted triangle
legend is flashing - current setting is displayed
---
NO
YES
SHOW
+
NO
YES
SHOW
return to item
---
item will not be displayed in USER menu
item will be displayed in USER menu with the option of setting
item will be solely displayed in USER menu
INSTRUCTIONS FOR USE OM 402PWR
|
11
5
SETTING
5.0
light
Setting “LIGHT”
LIGHT
Simple programming menu
- contains only items necessary for instrument setting and is protected by optional number code
• For capable users
LIGHT
• Only items necessary for instrument
setting
• Access is password protected
SETTING
• Possibility to arrange items of the
„User“ menu
• Linear menu structure
Preset from manufacture
Password
Menu
USER menu
Setting the items
Preset
“0”
LIGHT
off
DEF
“Power”
PROJECTION OF MEASURING UNITS ON THE DISPLAY
Multiplying constant
Channel/Unit
“Channel I” - Current
“Channel U” - Voltage
“Channel P” - Active power
“Channel Fr” - Frequency
“S” - Apparent power
“Q” - Idle power
“Cos Fi” - Power factor
12
|
mili (0,001)
without (1)
kilo (1000)
mA
mV
mW
Hz
mVA
mVr
-
A
V
W
Hz
VA
VAr
-
kA
kV
kW
Hz
kVA
kVr
-
INSTRUCTIONS FOR USE OM 402PWR
light
5
!
142. 8 W
Upon delay exceeding 60 s the programming mode
is automatically discontinued and the instrument
itself restores the measuring mode
+
0
PASSW.
Range “Channel I”
MODE I
SETTING
Access password
AC Filter
Range “Channel I”
I R.3
MODE U
U INP.3
Multiplying const. “Channel I”
Projection “Channel I”
MAX I
5
Projection “Channel U”
PREF. I
450
Calculation of 3-phase output
PREF. U
NO
Select input for “Mat. functions”
PREF. P
FORM. I
0000. o o
Decimal point for “Channel U”
- NULL
Multiplying const. “Channel P”
3F. GRID
NO
Decimal point for “Channel I”
- NULL
Multiplying const. “Channel U”
MAX U
AC. FIL.
FORM. U
00000. o
Decimal point for “Channel P”
- NULL
Multiplying const. “Mat. functions”
FORM. P
000000
Decimal point for “Mat. functions”
INP. M
FIL. P
PREF. M
- NULL
LIM. L.1
20
LIM. L. 2
40
LIM. L.3
60
LIM. L.4
80
FORM. M
FLOA. P.
Option - comparator
Option - Analog output
TYP A. O .
I 20
Menu type
MENU
OM 402. . . .
0
CALIB.
YES
New password
ENGL.
142. 8 W
N. PASS.
MAX A. O .
100
Return to manufacture setting
Return to manufacture calibration
LIGHT
Language selection
LANG.
MIN A. O .
SETTIN.
POWER
Identification
0
IDENT.
YES
Return to measuring mode
INSTRUCTIONS FOR USE OM 402PWR
|
13
5
light
SETTING
142. 8 W
PASSW.
Entering access password
for access into the menu
0
PASSW.
Access into instrument
menu
PAS = 0
- access into menu is unrestricted, after
releasing keys you automaticaly move to first
item of the menu
PAS > 0
- access into menu is protected ny number
code
Example
Set “Password” = 42
1
2
32
42
MODE I
I R.1
MODE I
I R. 2
I R.3
02
Select measuring range of
the instrument
- settting of the input range depends on the
ordered measuring range
12
22
MODE I
MODE I
0
Menu
Range „K“
Range „P“
I R.1
0…60 mV
0…1 A
I R.2
0…150 mV
0…2,5 A
I R.3
0…300 mV
0…5 A
DEF = I R.3 (Range “P”)
Example
Ronge: 0…150 mV
I R.3
14
|
INSTRUCTIONS FOR USE OM 402PWR
I R. 2
MODE U
light
5
U INP. 2
U INP.3
MODE U
Select measuring range of
the instrument
- setting of the input range depends on the
ordered measuring range
MODE U
MODE U
SETTING
Menu
Range „S“
Range „U“
U INP.2 0…10 V
0…250 V
U INP.3 0…120 V
0…450 V
DEF = U INP.3
Example
Range 0…250 V
U INP. .3
AC. FIL.
YES
AC. FIL.
U INP. 2
AC. FIL.
NO
Input filter function suppressing the DC
component
NO
- allows for measurement solely of the AC
component
- measures both the DC and the AC
component of the input signal
YES
DEF = NO
Filter is on
- measures solely the AC component of input
signal
Example
Measurrement of alternating component only
NO
Filter is off
YES
MAX I
INSTRUCTIONS FOR USE OM 402PWR
|
15
5
SETTING
MAX I
light
MAX I
Setting projection for
maximum input current
5
Setting display projection
for maximum value of
- position of the DP does not affect display
projection
input current
- range of the setting -999…9999
DEF = 5
Example
Projection for 150 mV > Max = 3500
5
PREF. I
4
3
0
00
000
100
200
300
400
0500
1500
2500
3500
PREF. I
m-MILI
-NULL
K-KILO
Select the multiplying
constant
- the constant allows for another mathematical
calculation with the option of extended
projection of measuring units, display
projection
INSTRUCTIONS FOR USE OM 402PWR
K-KILO
- NULL
Constant 0,001
prefix “m” added
Constant 1
without prefix
Constant 1000
prefix “K” added
Example
Division constant 1000 > PREF. I = K-KILO
-NULL
m-MILI
K-KILO
DEF = 1 (- NULL)
|
1
500
PREF. I
16
2
FORM. I
light
FORM. I
000000
00000. o
0000. o o
000. o oo
5
FLOA. P.
FORM. I
SETTING
Setting projection of the
decimal point
DEF = 0000.oo
- positioning of the DP is set here in the
measuring mode
Example
Projection of DP on display > 00000.o
0000. o o
MAX U
450
MAX U
00000. o
MAX U
Setting projection for
maximum input voltage
Setting display for
maximum value of input
- position of the DP does not affect display
projection
current
- range of the setting is -999…9999
DEF = 450
Example
Projection for 450 V > MAX U = 1000
450
450
460
470
480
490
200
100
000
500
400
300
0000
1000
PREF. U
INSTRUCTIONS FOR USE OM 402PWR
|
17
5
light
SETTING
PREF. U
m-MILI
PREF. U
-NULL
K-KILO
Select multiplying constant
m-MILI
- constant allows for another mathematic
recalculation with option of extended
projection of measuring units, display
projection
- NULL
K-KILO
Constant 0,001
prefix “m” added
Constant 1
without prefix
Constant 1000
prefix “K” added
DEF = 1 (- NULL)
Example
Division constant 1000 > PRE. I = K-KILO
-NULL
FORM. U
K-KILO
000000
00000. o
FORM. U
0000. o o
FLOA. P.
FORM. U
000. o oo
Setting projection of the
decimal point
DEF = 00000.o
- positioning of the DP is set here in the
measuring mode
Example
Projection of DP on display > 000000
00000. o
18
|
INSTRUCTIONS FOR USE OM 402PWR
000000
3F. GRID
light
3F. GRID
NO
3F. GRID
SETTING
5
YES
Select calulation for
3-phase network
NO
DEF = NO
YES
Calculation is inactive
Calculation is active
- value is calculated with presumption of
balanced take-off in all phases
- for P, S Q the value is multiplied by 3x
Example
Calculation for 3-phase network > 3F. GRID = YES
NO
PREF. P
m-MILI
PREF. P
YES
-NULL
PREF. P
K-KILO
Select multiplying constant
m-MILI
- constant allows for another mathematic
recalculation with option of extended
projection of measuring units, display
projection
- NULL
K-KILO
Constant 0,001
prefix “m” added
Constant 1
without prexix
Constant 1000
prefix “K” added
DEF = 1 (- NULL)
Example
Division constant 1000 > PREF. I = K-KILO
-NULL
K-KILO
FORM. U
INSTRUCTIONS FOR USE OM 402PWR
|
19
5
light
SETTING
FORM. P
000000
00000. o
0000. o o
FLOA. P.
000. o oo
FORM. P
Setting projection of the
decimal point
DEF = 000000
- positioning of the DP is set here in the
measuring mode
Example
Projection of DP on display > 00000.o
000000
INP. M.
00000. o
NO
FIL. I
INP. M.
FIL. U
....
S
Select input for calculation
of mathematic function
INPUT M
- select value from which mathematic function
will be calculated
DEF = FIL. P
Input Mat. functions from “Channel U” > INP. M=FIL. U
FIL. P
20
|
INSTRUCTIONS FOR USE OM 402PWR
FIL. U
COS. FI
INP. M.
Q
INP. M.
Menu
Description
BAN
Mat. functions off
FIL. I
From „Chan. I“ after filtration
FIL. U
From „Chan. U“ after filtration
FIL. P
From „Chan. P“ after filtration
FIL. F
From „Chan. F“ after filtration
MIN
From „Min. value“
MAX
From „Max. value“
S
From „Apparent power“
Q
From „Idle power“
COS. FI
From „Cos Fi“
Example
light
PREF. M.
m-MILI
PREF. M.
-NULL
SETTING
5
K-KILO
Select multiplying constant
m-MILI
- constant allows for another mathematic
recalculation with option of extended
projection of measuring units, display
projection
- NULL
K-KILO
Constant 0,001
prefix “m” added
Constant 1
without prexix
Constant 1000
prefix “K” added
DEF = 1 (- NULL)
Example
Division constant 1000 > PREF. I = K-KILO
-NULL
FORM. M.
000000
K-KILO
00000. o
FORM. M
0000. o o
00. o ooo
0. o oooo
FLOA. P.
FORM. M
000. o oo
Setting projection of the
decimal point
DEF = FLOA. P.
- positioning of the DP is set here in the
measuring mode
Example
Projection of DP on display > 000000
PLOV. T
000000
FORM. P
*subsequent item on the menu depends on instrument equipment
INSTRUCTIONS FOR USE OM 402PWR
|
21
5
SETTING
LIM L. 1
light
LIM L.1
Setting boundary for
limit 1
20
- contingent modification of hysteresis or
delay may be performed in “PROFI” menu
Setting boundary for
limit 1
Displayed only with options > Comparators
- range of the setting is -99999…999999
- default “Hystresis”=0 “Delay”=0
DEF = 20
Example
Setting limit 1 > L 1 = 32
20
LIM L. 2
21
22
32
- contingent modification of hysteresis or
delay may be performed in “PROFI” menu
Setting boundary for
limit 2
- range of the setting is -99999…999999
- default “Hysteresis”=0 “Delay”=0
DEF = 40
Example
Setting limit 2 > L 2 = 53.1
40
MENU
Setting boundary for
limit 2
40
LIM L. 2
22
41
41
31
231
331
431
531
000531
000531.
00053.1
MENU
031
131
0531
00531
* subsequent item on the menu depends on instrument
equipment
!
Items for “Limits” and “Analog output” are accessible only if incorporated in the instrument.
22
|
INSTRUCTIONS FOR USE OM 402PWR
light
LIM L.3
60
LIM L.3
SETTING
Setting boundary for
limit 3
- contingent modification of hysteresis or
delay may be performed in “PROFI” menu
Setting boundary for
limit 3
DEF = 60
Example
Setting limit 3 > L 3 = 85
60
61
62
63
65
75
85
MENU
80
LIM L.4
Displayed only with options > Comparators
- range of the setting is -99999…999999
- default “Hysteresis”=0 “Delay”=0
LIM L.4
5
64
65
* subsequent item on the menu depends on instrument
equipment
Setting boundary for
limit 4
- contingent modification of hysteresis or
delay may be performed in “PROFI” menu
Setting boundary for
limit 4
- range of the setting is -99999…999999
- default “Hysteresis”=0 “Delay”=0
DEF = 80
Example
Setting limit 4 > L 4 = 103
80
81
82
83
03
003
103
MENU
83
93
* subsequent item on the menu depends on instrument
equipment
INSTRUCTIONS FOR USE OM 402PWR
|
23
5
SETTING
TYP. A. O .
light
0-20mA
E. 4-20
4-20mA
0-5mA
0-2 V
0-5 V
0-10 V
Setting the type of analog
output
Displayed only with options > Analog output
TYP. A. O .
Menu
Range
0-20mA
0…20 mA
E. 4-20mA
4…20 mA
4-20mA
4…20 mA
0-5mA
0…5 mA
0-2 V
0…2 V
0-5 V
0…5 V
0-10 V
0…10 V
Description
with indication of error statement (<3,6 mA)
DEF = 4…20 mA
Example
Type of analog output - 0...10 V > TYP. A.O. = 0-10 V
4-20mA
MIN A. O .
0-5mA
Assigning the display
value to the beginning of
the AO range
0
MIN A. O .
0-2 V
Assigning the display
value to the beginning of
0-5 V
0-10 V
MIN A. O .
DEF = 0
the AO range
- range of the setting is -999…9999
Example
Display value for the beginning of the AO range > MIN A.O. = 0
0
MAX A. O .
!
Items for “Limits” and “Analog output” are accessible only if incorporated in the instrument.
24
|
INSTRUCTIONS FOR USE OM 402PWR
light
MAX A. O .
100
MAX A. O .
SETTING
5
Assigning the display
value to the end of the
AO range
Assigning the display
value to the end of the
DEF = 100
AO range
Displayed only with options > Analog output
- range of the setting is -999…9999
Example
Display value for the end of the AO range > MAX A.O. = 120
100
100
110
120
MENU
INSTRUCTIONS FOR USE OM 402PWR
|
25
5
SETTING
MENU
light
LIGHT
MENU
PROFI
Setting the menu type
LIGHT/PROFI
LIGHT > menu LIGHT, a simple menu,
which contains only the most essential items
necessary for instrument setting
> linear tree structure
PROFI > menu PROFI, a complete menu for
complete instrument setting
> tree menu structure
DEF = LIGHT
Example
Menu LIGHT > MENU = LIGHT
LIGHT
CALIB.
CALIB.
YES
CALIB.
Restoration of manufacture
calibration
Prior to execution of any modifications you
will be asked to confirm your selection.
(YES)
- in the event of error calibration it is feasible
to restore manufacture calibration.
Example
Restoration of manufacture setting > CALIIB.
CALIB.
SETTIN.
YES
POWER VOLTAGE
SETTIN.
SETTIN.
CURREN.
Restoration of manufacture
instrument setting
- in the event of error setting or calibration it
is possible to return to manufacture setting.
Prior to execution of changes you will be
asked to confirm your selection (YES)
- reading manufacture calibration and
primary setting of items in menu (DEF)
26
|
INSTRUCTIONS FOR USE OM 402PWR
POWER
VOLTAG.
Setting for measurement of
power output - “Channel P”
Setting for measurement of
voltage - “Channel U”
Setting for measurement of
current - “Channel I”
Example
Restoration of manufacture setting > SETTIN.
POWER
- presetting “source” for further evaluation
(e.g.: select “VOLTAGE” > evaluation of
limits, analog, min/max. value, etc... from
“Channel U”
CURREN.
Restore manufacture
Presetting constants and evaluation
SETTIN.
LANG.
light
LANG.
CZECH
LANG.
SETTING
ENGL.
Selection of language in
instrument menu
- selection of language version of the
instrument menu
DEF = ENGL.
Example
Language selection - ENGLISH > LANG. = ENGL.
CZECH
N. PASS.
0
N. PASS.
5
ENGL.
N. PASS.
Setting new access
password
Setting new access
password
- access password for menu LIGHT/PROFI
- range of the number code 0…9999
- upon setting the password to “000” the
access to menu LIGHT/PROFI is free without
prompt to enter it
- in the event of loss universal password
“8177” may be used
DEF = 0
Example
New password - 341 > N.PASS. = 341
0
1
01
11
21
31
41
041
141
241
341
IDENT.
INSTRUCTIONS FOR USE OM 402PWR
|
27
5
light
SETTING
instrument type
IDENT.
OM 402PWR
YES
IDENT.
Instrument SW version
- the display shows the type of instrument
indication, SW number, SW version and
current input setting (Mode)
142. 8 W
28
|
Return to measuring mode
INSTRUCTIONS FOR USE OM 402PWR
SW number
61-
SW version
139
setting the input type
4-20mA
- if SW version contains a letter in first
position, then it is a customer SW
- after the identification is completed the
menu is automatically exited and the
instrument restores the measuring mode
light
SETTING
INSTRUCTIONS FOR USE OM 402PWR
|
5
29
6
SETTING
6.0
profi
Setting “PROFI”
PROFI
Complete programming menu
• contains complete instrument menu and is protected by optional number code
• designed for expert users
• preset from manufacture is menu LIGHT
• For expert users
• Complete instrument menu
PROFI
• Access is password protected
SETTING
• Possibility to arrange items of the
„User“ menu
• Tree menu structure
Switching over to “PROFI” menu
+
• temporary switch-over to PROFI menu, which is suitable to edit a few items
• after quitting PROFI menu the instrument automatically switches to LIGHT menu
• access is password protected (if it was not set under item N. PASS. =0)
+
• access into LIGHT menu and transition to item „MENU“ with subsequent selection of
„PROFI“ and confirmation
• after re-entering the menu the PROFI type is active
• access is password protected (if it was not set under item N. PASS. =0)
30
|
INSTRUCTIONS FOR USE OM 402PWR
profi
SETTING
6
PROJECTION OF MEASURING UNITS ON THE DISPLAY
Multiplying constant
Channel/Unit
“Channel I” - Current
“Channel U” - Voltage
“Channel P” - Active power
“Channel Fr” - Frequency
“S” - Apparent power
“Q” - Idle power
“Cos Fi” - Power factor
mili (0,001)
bnull (1)
kilo (1000)
mA
mV
mW
Hz
mVA
mVr
-
A
V
W
Hz
VA
VAr
-
kA
kV
kW
Hz
kVA
kVr
-
INSTRUCTIONS FOR USE OM 402PWR
|
31
6
profi
SETTING
6.1
Setting “PROFI” - INPUT
The primary instrument parameters are
set in this menu
INPUTS
CLEAR
CHANNE.
CONFIG.
CLEAR
OUTPUT.
RTC
CONFIG.
SERVIC.
EXT. IN.
RTC
KEYS
EXT. IN.
KEYS
Resetting internal
values
Selection of measuring
range and parameters
Setting date and time for
option with RTC
Setting external inputs
functions
Assigning further
functions to keys on the
instrument
6.1.1
Resetting internal values
CLEAR
INPUTS
CLEAR
CL. TAR.
TARE I
CHANNE.
CONFIG.
CL. M.M.
TARE U
OUTPUT.
RTC
CL. MEM.
TARE P
SERVIC.
EXT. IN.
TARE F
KEYS
ALL
TARE I
TARE U
TARE P
TARE F
ALL
INPUTS
CLEAR
CL. TAR.
CHANNE.
CONFIG.
CL. M.M.
OUTPUT.
RTC
CL. MEM.
SERVIC.
EXT. IN.
KEYS
CL. M. M .
Resetting internal values
Tare resetting
channel - Current
Tare resetting
channel - Voltage
Tare resetting
channel - Power
Tare resetting
channel - Frequency
Tare resetting on
all channels
Resetting min/max value
- resetting memory for the storage of
minimum and maximum value achieved
during measurement
CL. MEM.
Resetting the instrument
memory
- resetting memory with data measured in
the “FAST” or “RTC” modes
- not in standard equipment
32
|
INSTRUCTIONS FOR USE OM 402PWR
profi
6.1.2a
SETTING
Select measuring mode - Measuring channel I
MODE I
CLEAR
MODE I
I R.1 .
CHANNE. CONFIG.
INPUTS
MODE U
I R. 2.
RTC READ/S
I R.3 .
OUTPUT.
SERVIC.
6.1.2b
6
Select measuring range
- Current
- selection of measuring range depends on
the ordered version
DEF
I R.1 .
EXT. IN.
AC. FIL.
KEYS
SWITCH
I R. 2.
TIM. SW.
I R.3 .
Range 1
0…60 mV/1 A
Range 2
0…150 mV/2,5 A
Range 3
0…300 mV/5 A
Select measuring range - Measuring channel U
MODE U
CLEAR
MODE I
U INP. 2.
CHANNE. CONFIG.
INPUTS
MODE U
U INP.3 .
OUTPUT.
SERVIC.
DEF
EXT. IN.
AC. FIL.
KEYS
SWITCH
- selection of measuring range (input)
depends on the ordered version
U INP.3
Voltage input 2
0 …10 V/250 V
Voltage input 2
0 …120 V/450 V
READ/S
Selection of measuring
rate
U INP. 2
RTC READ/S
Select measuring range
- Voltage
TIM. SW.
6.1.2c
Selection of measuring rate
INPUTS
CLEAR
MODE I
5. 0
CHANNE. CONFIG.
MODE U
2.5
RTC READ/S
1. 2
OUTPUT.
SERVIC.
EXT. IN.
AC. FIL.
KEYS
SWITCH
TIM. SW.
0. 6
DEF
5. 0
2.5
1. 2
0. 6
5,0 measurements/s
2,5 measurements/s
1,2 measurements/s
0,6 measurements/s
INSTRUCTIONS FOR USE OM 402PWR
|
33
6
profi
SETTING
6.1.2d
Select function of input filter
AC. FIL.
INPUTS
CLEAR
MODE I
NO
CHANNE. CONFIG.
MODE U
YES
OUTPUT.
SERVIC.
DEF
- allows for measurement of alternating
component of the input signal
RTC READ/S
EXT. IN.
AC. FIL.
KEYS
SWITCH
Input filter function
- supressing DC
component
NO
Filter is off
- measures both DC and AC components of
the input signal
TIM. SW.
YES
Filter is on
- measures only AC component of the input
signal
6.1.2e
Select projection of measuring channels
SWITCH
INPUTS
CLEAR
CHANNE. CONFIG.
OUTPUT.
SERVIC.
MODE I MANUAL
MODE U
AUTOM.
RTC READ/S
EXT. IN.
MODE
KEYS
AC. FIL.
SWITCH
TIM. SW.
DEF
Select switching of
projection of measuring
channels
- in this menu it s possible to select
switching for projection of measuring
channels I, U, P, F, MF
- selection of quantities for switching can
be set separately for individual channels
(e.g. for Channel I >> “CHANNELS >
CHAN. I > SWITCH.”)
MANUAL
Manual switching of
projection
- projection may be switched by selected
button (“INPUT > BUTTON”) or external
input (“INPUT > EXT. IN.”)
AUTOM.
CLEAR
MODE I
CHANNE. CONFIG.
INPUTS
MODE U
OUTPUT.
SERVIC.
RTC READ/S
EXT. IN.
KEYS
|
- switching time is adjustable in item “TIM.
SW.”
TIM. SW.
MODE
Setting switching interval
for projection
AC. FIL.
- time is adjustable in range 0,5…99,9 s
SWITCH
- DEF = 2
TIM. SW.
34
Switching of projection is
automatic
INSTRUCTIONS FOR USE OM 402PWR
profi
6.1.3
SETTING
Setting real time clock
RTC
CLEAR
TIME
CHANNE. CONFIG.
INPUTS
DATE
OUTPUT.
RTC
SERVIC.
EXT. IN.
TIME
Setting the real time
clock (RTC)
Time setting
- format 23.59.59
DATE
KEYS
6.1.4a
6
Date setting
- format DD.MM.YY
Select function of external input
EXT. IN.
CLEAR
EXT. 1
NO
CHANNE. CONFIG.
INPUTS
EXT. 2
HOLD
OUTPUT.
RTC
EXT. 3
LOCK. K.
SERVIC.
EXT. IN.
M. HOLD
B. PASS.
KEYS
TARE I
TARE U
TARE P
TARE F
TAR.ALL
TAR.AKT.
CLR. MM
CLR.T. I
CLR.T. U
CLR.T. P
CLR.T. F
C..T. ALL
C.T. AKT
SWCH.1.
SXCH. 2.
SWCH. 3
NO
HOLD
External input function
selection
Input is off
Activation of HOLD
Locking keys on the
instrument
Activation of locking
B. PASS.
access into programming
menu LIGHT/PROFI
Tare activation for
TARE I
“Channel I”
Tare activation for
TARE U
“Channel U”
Tare activation for
TARE p
“Channel P”
Tare activation for
TARE F
“Channel F”
Tare activation on all
TAR. ALL
channels
Tare activation on active
TAR. AKT.
channel
Resetting
CLR. MM
min/max value
Clear Tare
CLR. T. I
for “Channel I”
Clear tare
CLR. T. U
for “Channel U”
Tare clear
CLR. T. p
for “Channel P”
LOCK. K.
SAVE
INSTRUCTIONS FOR USE OM 402PWR
|
35
6
profi
SETTING
Table with control of external inputs
Channel
EXT. 1
EXT. 2
I
0
0
U
0
1
P
1
0
F
1
1
MF
0
0
1
cos 02
0
1
1
Min
1
0
1
Max
1
1
1
CLR. T. F
EXT. 3
C.T. ALL
C.T. AKT.
SWCH. 1
SWCH. 2
Clear tare for “Channel
F”
Clear tare on all
channels
Clear tare on active
channel
Successive switching of
channels projection
BCD switching of
channels projection -
Ext. 1, 2
- control see table
- after selecting this option setting for
“INPUT 2” is automatically banned
SWCH. 3
BCD switching of
channels projection -
EXT. 1, 2, 3
- control see table
- after selecting this option setting for “EXT.
2” and “EXT. 3”
Activation of measured
SAVE
data record in
instrument memory (not in standard
equipment)
*
- DEF EXT. 1 > HOLD
- DEF EXT. 2 > LOCK. K.
Setting procedure is identical for EXT. 2 and EXT. 3
- DEF EXT. 3 > SWCH. 1
6.1.4b
Selection of function “HOLD”
M. HOLD
INPUTS
CLEAR
CHANNE. CONFIG.
OUTPUT.
RTC
SERVIC.
EXT. IN.
EXT. 1
DISPL.
EXT. 2 DIS.+A.O.
EXT. 3 D.+A.O.+L.
M. HOLD
ALL
KEYS
DEF
DISPL.
DIS.+AO.
D.+AO.+L.
|
INSTRUCTIONS FOR USE OM 402PWR
“HOLD” locks only the
value displayed
“HOLD” locks the value
displayed and on AO
“HOLD” locks the value
displayed, on AO and
limit evaluation
ALL
36
Selection of function
“HOLD”
“HOLD” locks the entire
instrument
profi
6.1.5a
SETTING
Optional accessory functions of the keys
FN. LE.
CLEAR
LEFT
FN. LE.
NO
CHANNE. CONFIG.
INPUTS
DOWN
TMP. LE.
CLR. MM
MNU. LE.
CLR.T. I
OUTPUT.
RTC
UP
SERVIC.
EXT. IN.
ENTER
KEYS
CLR.T. P
NO
CLR.T. F
CLR. MM
C.T. ALL
CLR. T. I
CLR. T. U
MENU
TEMP. N.
CLR. T. p
TARE I
CLR. T. F
TARE U
C.T. ALL
TARE P
C.T. AKT.
TARE F
TAR.ALL
TAR.AKT.
SWCH. 1
!
Presetting measurement > POWER
Current (Chan. I)
Frequency (Chan. F)
Voltage (Chan. U)
cos fi (Mat. Fn.)
Presetting measurement > VOLTAGE
LEFT
UP
DOWN
ENTER
Current (Chan. I)
Frequency (Chan. F)
Power (Chan. P)
cos fi (Mat. Fn.)
Presetting measurement > CURRENT
LEFT
UP
DOWN
ENTER
MENU
Power (Chan. P)
Frequency (Chan. F)
Voltage (Chan. U)
cos fi (Mat. Fn.)
!
Key has no further
function
Resetting
min/max value
Clear tare for “Channel
I”
Clear tare for “Channel
U”
Clear tare for “Channel
P”
Clear tare for “Channel
F”
Clear tare on all
channels
Clear tare on active
channel
Direct access into menu
on selected item
- after confirmation of this selection the
“MNU. LE.” item is displayed on superior
menu level, where required selection is
performed
TEMP. N.
Preset values of the control keys DEF :
Assigning further
functions to instr. keys
- „FN. LE.“ > executive functions
- „TMP. LE.“ > temporary projection of
selected values
- „MNU. LE.“ > direct access into menu on
selected item
CLR.T. U
C.T. AKT
LEFT
UP
DOWN
ENTER
6
Temporary projection of
selected values
- after confirmation of this selection the
item “TMP. LE.” is displayed on superior
menu level, where required selection is
performed
TARE I
TARE U
TARE p
TARE F
TAR. ALL
TAR. AKT.
SWCH. 1
Tare activation for
“Channel I”
Tare activation for
“Channel U”
Tare activation for
“Channel P”
Tare activation for
“Channel F”
Tare activation on all
channels
Tare activation on active
channel
Successive switching of
channels projection
Setting is identical for LEFT, DOWN, UP and ENTER
INSTRUCTIONS FOR USE OM 402PWR
|
37
6
profi
SETTING
142. 8
+
PASSW.
Access password
0
INPUTS
CLEAR
CHANNE.
TARE I ALL
CL. TAR.
CHAN. I
MAX. I
0
FILTER
CL. M. M .
CON. F.I.
CL. MEM.
I R.1 I R.2 I R.3
MODE I
MODE U
U INP. 2 U INP. 3
READ/S
0.6 1.2 2.5 5.0
CHAN. U
AC FIL.
YES NO
SWITCH
MANUAL AUTOM.
CHAN. P
2
TIME
0
DATE
0
FORM. I
000000 00000.o FLOAT. P.
SWCH. I
YES NO
SAVE I
YES NO
Further setting is identical with setting “Channel I”
YES NO
3F. GRID.
Further setting is identical with setting “Channel I”
CHAN. F
TIM. SW.
FILTER
EXT. IN.
EXT. 1
MAT. FN.
EXT. 3
OFF SAVE
DISPL. ALL
MIN. MAX.
LEFT
ENTER
38
|
NO POLIN ROOT SIN X
MATH. F.
OFF SAVE
OFF SAVE
NO FIL. I FIL. U S Q COS. FI
INP. F.
CON. A
0
CON. F
0
PREF. M.
KEYS
0
Further setting is identical with setting “Channel I”
EXT. 2
M. HOLD
NO AVER. ROUND
MOD. F. F
CON. F. F.
RTC
0
m-MILI -NULL K-KILO
PREF. I
CONFIG.
NO AVER. ROUND
MOD. F.I
m-MILI -NULL K-KILO
FORM. M.
000000 00000.o FLOAT. P.
SWCH. M.
YES NO
SAVE M.
YES NO
INP. M. M .
NO CHAN. I FIL. I MAT. FN. COS.
FN. LE.
NO CLR. MM CLR. T.I. MENU TMP. H. TARE I TARE U TARE P TARE P TARE F TAR.ALL
TMP. LE.
NO CHAN. I. FIL. I CHAN. U.. FIL. U CHAN. P. FIL. P TIME DATE TARE I TARE U MNU. LE.
LIM. 1 LIM. 2. LIM. 3 LIM. 4
INSTRUCTIONS FOR USE OM 402PWR
profi
SETTING
OUTPUT.
LIMITS
SERVIC.
LIM 1
LIM 4
INP. L1
NO CHAN. I FIL. I MAX
MOD. L.1
HYSTER FROM.. . DOSING
TYP L.1
CLOSE OPEN
LIM. L.1
20
HYS. L1.
0
ON L.1
0
OFF L.1
0
PER. L.1
0
TIM. L.1
0
LIGHT PROFI
MENU
RESTOR.
CALIB.
YES
SETTIN.
POWER
VOLTAG.
CURREN.
CZECH ENGL.
MENU
N. PASS.
DATA
IDENT.
ASCII M. BUS
NO CHAN. I FIL. I MAT. FN. COS. 02 MIN MAX
INP. A. O .
0-20mA E 4-20 4-20mA 0-5mA 0-2V 0-5V 0-10V
TYP. A. O .
MIN A. O .
0
MAX A. O .
5 / 450 / 2250
PERM.
BRIGHT
02
L TAR.AKT. SWCH. 1
OM 402PWR. . . .
0
PROT.
DISP.
0
600 1200 2400 115200 2320400
DAUB
ADDR.
AN. OUT.
6
CHAN. I FIL. I MAT. FN. COS. 02 MIN MAX
0% 25% 50% 75% 100%
!
Upon delay exceeding 60 s the programming mode
is automatically discontinued and the instrument
itself restores the measuring mode
TARE P TAR F.
INSTRUCTIONS FOR USE OM 402PWR
|
39
6
profi
SETTING
6.1.5b
Optional accessory functions of the keys - Temporary projection
TMP. LE.
CLEAR
LEFT
FN. LE.
NO
CHANNE. CONFIG.
INPUTS
DOWN
TMP. LE.
CHAN. I
OUTPUT.
RTC
UP
FIL. I
SERVIC.
EXT. IN.
ENTER
CHAN. I
KEYS
FIL. U
CHAN. U
FIL. P
CHAN. F
FIL. F
MAT. FN.
COS. 02
MIN.
MAX
LIM 1
LIM 2
LIM 3
LIM 4
TIME
DATE
TARE I
TARE U
TARE P
TARE F
!
Setting is identical for LEFT, DOWN, UP and ENTER
!
“COS. 02” is derived from “COS. FI.” when the
range -1…1 is tranferred to 0…2 for evaluation
40
|
INSTRUCTIONS FOR USE OM 402PWR
Temporary projection of
selected item
- “Temporary” projection of selected value
is displayed for the time of keystroke
- “Temporary” projection may be switched
+ “Selected
to permanent by pressing
key”, this holds until the stroke of any key
Temporary projection
is off
Temporary projection of
CHAN. I
“Channel I” value
Temporary projection of
FIL. I
“Channel I” value after
processing digital filters
Temporary projection of
CHAN. U
“Channel U” value
Temporary projection of
FIL. U
“Channel U” value after
processing digital filters
Temporary projection of
CHAN. P
“Channel P” value
Temporary projection of
FIL. P
“Channel P” value after
processing digital filters
Temporary projection of
CHAN. F
“Channel F” value
Temporary projection of
FIL. F
“Channel F” value after
processing digital filters
Temporary projection of
MAT. FN.
“Mathematic functions”
value
Temporary projection of
COS. O2
auxiliary channel cos fi
(range 0-2)
Temporary projection of
MIN
“Min. value”
Temporary projection of
MAX
“Max. value”
Temporary projection of
LIM 1
“Limit 1” value
Temporary projection of
LIM 2
“Limit 2” value
Temporary projection of
LIM. 3
“Limit 3” value
NO
profi
SETTING
TARE F
Temporary projection of
“Limit 4” value
Temporary projection of
“TIME” value
Temporary projection of
“DATE” value
Tare value for
“Channel I”
Tare value for
„Channel U”
Tare value for “Channel
P”
Tare value for
“Channel F”
MNU. LE.
Assigning access to
selected menu item
LIM. 4
TIME
DATE
TARE I
TARE U
TARE p
6.1.5c
6
Optional accessory functions of the keys - Direct access to item
CLEAR
LEFT
FN. LE.
LIM 1
CHANNE. CONFIG.
INPUTS
DOWN
MNU. LE.
LIM 2
OUTPUT.
RTC
UP
LIM 3
SERVIC.
EXT. IN.
ENTER
LIM 4
LIM 1
LIM 2
LIM 3
KEYS
LIM 4
Direct access to item
“LIM 1”
Direct access to item
“LIM 2”
Direct access to item
“LIM 3”
Direct access to item
“LIM 4”
!
Setting is identical for LEFT, DOWN, UP and ENTER
INSTRUCTIONS FOR USE OM 402PWR
|
41
6
profi
SETTING
6.2
Setting “PROFI” - CHANNELS
The primary instrument parameters are
set in this menu
INPUTS
CHAN. I
CHANNE.
CHAN. U
CHAN. I
OUTPUT.
CHAN. P
CHAN. U
SERVIC.
CHAN. F
CHAN. P
MAT. FN.
CHAN. F
MIN. MAX
MAT. FN.
MIN. MAX
Setting parameters of
measuring “Channel I”
Setting parameters of
measuring “Channel U”
Setting parameters of
measuring “Channel P”
Setting parameters of
measuring “Channel F”
Setting parameters of
mathematic functions
Selection of access
and evaluation of Min/
max value
6.2.1a
Display projection for “Channel I”
MAX. I
42
INPUTS
CHAN. I
MAX. I
CHANNE.
CHAN. U
FILTER
OUTPUT.
CHAN. P
PREF. I
SERVIC.
CHAN. F
FORM. I
MAT. FN.
SWCH. I
MIN. MAX
SAVE I
|
INSTRUCTIONS FOR USE OM 402PWR
Setting display projection
for minimum value of
input signal
- range of the setting is -99999…999999
- DEF = 5
profi
6.2.1b
SETTING
6
Digital filters
MOD. F. I
INPUTS
CHAN. I
MAX. I
MOD. F.I
NO
CHANNE.
CHAN. U
FILTER
CON. F.I
AVER.
OUTPUT.
CHAN. P
PREF. I
FLOAT.
SERVIC.
CHAN. F
FORM. I
EXPON.
MAT. FN.
SWCH. I
ROUND
MIN. MAX
SAVE I
DEF
Selection of digital
filters
- at times it is useful for better user
projection of data on display to modify it
mathematically and properly , wherefore
the following filters may be used:
NO
AVER.
Filters are off
Measured data
average
- arithmetic average from given number
(„CON.F. A.“) of measured values
- range 2…100
FLOAT.
Selection of floating filter
- floating arithmetic average from given
number („CON.F. A.“) of measured data
and updates with each measured value
- range 2…30
EXPON.
Selection of exponential
filter
- integration filter of first prvního grade with
time constant („CON.F. A.“) measurement
- range 2…100
ROUND
Measured value
rounding
- is entered by any number, which
determines the projection step
(e.g: “CON.F. A.”=2,5 > display 0,
2.5, 5,...)
CON. F. A.
Setting constants
- this menu item is always displayed after
selection of particular type of filter
- DEF = 2
INSTRUCTIONS FOR USE OM 402PWR
|
43
6
profi
SETTING
6.2.1d
Select multiplying constant
PREF. I
INPUTS
CHAN. I
MAX. I
m-MILI
CHANNE.
CHAN. U
FILTER
- NULL
OUTPUT.
CHAN. P
PREF. I
K-KILO
SERVIC.
CHAN. F
FORM. I
MAT. FN.
SWCH. I
MIN. MAX
SAVE I
6.2.1d
DEF
Select multiplying
constant
- constant allows for another mathematic
recalculation with option of extended
projection of measuring units
K-KILO
Constant 0,001
prefix “m” added
Constant 1
without prefix
Constant 1000
prefix “K” added
FORM. I
Selection of decimal
point
m-MILI
- NULL
Projection format - positioning of decimal point
INPUTS
CHAN. I
CHANNE.
CHAN. U
FILTER 00000. o
MAX. I
OUTPUT.
CHAN. P
PREF. I 0000. o o
SERVIC.
CHAN. F
FORM. I 000. o oo
MAT. FN.
SWCH. I
MIN. MAX
SAVE I
000000
FLOA. P.
- the instrument allows for classic projection
of a number with positioning of the DP
as well as projection with floating DP,
allowing to display a number in its most
exact form „FLOAT.P.“
000000.
00000. o
0000. o o
000. o oo
FLOA. P.
Setting DP - XXXXXX.
Setting DP - XXXXX.x
Setting DP - XXXX.xx
Setting DP - XXX.xxx
Floating DP
DEF “Channel I” - 000.ooo
DEF “Channel U” - 00000.o
DEF “Channel P” - 000000
DEF “Channel F” - 0000.oo
44
|
INSTRUCTIONS FOR USE OM 402PWR
profi
6.2.1e
SETTING
Select channel projection during switching
SWCH. I
INPUTS
CHAN. I
MAX. I
YES
CHANNE.
CHAN. U
FILTER
NO
OUTPUT.
CHAN. P
PREF. I
SERVIC.
CHAN. F
FORM. I
MAT. FN.
SWCH. I
MIN. MAX
SAVE I
6.2.1f
DEF
Select channel projection
during switching
- setting in this item enables the user to
select individual measuring channels,
which will be displayed during switching
function channels „SWCH. 1“
YES
NO
Projection permitted
Projection prohibited
Selection of storing data into instrument memory
SAVE I
INPUTS
CHAN. I
MAX. I
YES
CHANNE.
CHAN. U
FILTER
NO
OUTPUT.
CHAN. P
PREF. I
SERVIC.
CHAN. F
FORM. I
MAT. FN.
SWCH. I
MIN. MAX
SAVE I
6.2.2a
6
DEF
Selection of storing data
into instrument memory
- by selection in this item you allow to
register values into instrument memory
- another setting in item “OUTPUT. >
MEMORY” (not in standard experiment)
YES
NO
Measured data are
stored in the memory
Measured data are not
stored
Display projection for “Channel U”
MAX. U
INPUTS
CHAN. I
MAX. U
Setting display projection
for minimum value of
input signal
- range of the setting is -99999…999999
CHANNE.
CHAN. U
FILTER
OUTPUT.
CHAN. P
PREF. U
SERVIC.
CHAN. F
FORM. U
!
MAT. FN.
SWCH. U
Further setting is identical with setting “Channel I”
MIN. MAX
SAVE U
- DEF = 450
INSTRUCTIONS FOR USE OM 402PWR
|
45
6
profi
SETTING
6.2.3a
Setting parameters for “Channel P”
3F. SIT
INPUTS
CHAN. I
3F. SIT
YES
CHANNE.
CHAN. U
FILTER
NO
OUTPUT.
CHAN. P
PREF. P
SERVIC.
CHAN. F
FORM. P
MAT. FN.
SWCH. P
MIN. MAX
SAVE P
NO
DEF
YES
Calculation of 3-phase
power
Calculation is off
Calculation is on
- value is calculated with presumption of
balanced take-off on all phases
- for P, S Q the value is multiplied by 3x
!
Further setting is identical with setting “Channel I”
6.2.4a
Setting parameters for “Channel F”
MOD. F. F
INPUTS
CHAN. I
FILTER
MOD. F. F
NO
CHANNE.
CHAN. U
FORM. F
CON. F. F.
AVER.
OUTPUT.
CHAN. P
SWCH. F
FLOAT.
SERVIC.
CHAN. F
SAVE F
EXPON.
MAT. FN.
DEF
Setting digital filters
- sometimes, it is useful for better user
projection of data on display to aptly
modify it mathematically and to that
purpose the following filters may be used
ROUND
NO
MIN. MAX
AVER.
FLOAT.
EXPON.
ROUND
CON F. F.
Filters are off
Averaging of measured
value
Select floating filter
Select exponential filter
Rounding measured
value
Setting constant
!
Further setting is identical with setting “Channel I”
46
|
INSTRUCTIONS FOR USE OM 402PWR
profi
6.2.5a
SETTING
6
Mathematic functions - select input
INP. M.
INPUTS
CHAN. I
INP. M.
NO.
CHANNE.
CHAN. U
MAT. F.
FIL. I
OUTPUT.
CHAN. P
CON. A
FIL. U
SERVIC.
CHAN. F
CON. B
FIL. P
MAT. FN.
CON. C
FIL. F
MIN. MAX
CON. D
MIN.
CON. E
MAX
CON. F
S
PREF. M.
Q
FORM. M.
COS. FI
- select value from which mathematic
function is calculated
DEF
SAVE M.
FIL. I
Mathematic functions
are off
From “Channel I” after
modification by digital
FIL. U
From “Channel U” after
modification by digital
FIL. P
From “Channel P” after
modification by digital
FIL. F
From “Channel F” after
modification by digital
NO
filter
filter
filter
DESC. M.
SWCH. M.
Select input for
calculation of mat.
function
filter
MIN
MAX
S
Q
COS. FI
From “Min. value”
From “Max. value”
From “Apparent power”
From “Idle power”
From “cos fi”
INSTRUCTIONS FOR USE OM 402PWR
|
47
6
profi
SETTING
6.2.5b
Mathematic functions
Select mathematic
functions
MATH. F.
INPUTS
CHAN. I
INP. M.
OFF
CHANNE.
CHAN. U
MATH. F.
POLIN.
OUTPUT.
CHAN. P
CON. A
1/MUL.
SERVIC.
CHAN. F
CON. B
LOGAR.
MAT. FN.
CON. C
EXPON.
MIN. MAX
CON. D
POWER
CON. E
ROOT
KON. F
SIN X
PREF. M.
FORM. M.
DESC. M.
DEF
POLIN
Ax5 + Bx 4 + Cx3 + Dx 2 + Ex + F
1/x
1/MUL.
A B C
D E
+
+
+
+ +F
x5 x4 x 3 x 2 x
Logarithm
LOGAR.
⎛ Bx + C ⎞
A × ln ⎜
⎟+ F
⎝ Dx + E ⎠
Exponential
EXPON.
SWCH. M.
SAVE M.
Mathematic functions
are off
Polynome
OFF
A× e
⎛ Bx +C ⎞
⎜
⎟
⎝ Dx+ E ⎠
+F
Power
POWER
A × ( Bx + C )(Dx + E ) + F
Root
ROOT
Bx + C
A×
+F
Dx + E
Sin x
SIN X
A sin x + B sin x + C sin 3 x + D sin 2 x
5
4
+ E sinx + F
CON. -
Setting constants for
calculation of mat.
functions
- this menu is displayed only after selection
of given mathematic function
48
|
INSTRUCTIONS FOR USE OM 402PWR
profi
6.2.5c
SETTING
6
Mathematic functions - select multiplying constant
PREF. M.
INPUTS
CHAN. I
INP. M
m-MILI
CHANNE.
CHAN. U
MATH. F.
- NULL
OUTPUT.
CHAN. P
CON. A
K-KILO
SERVIC.
CHAN. F
CON. B
MAT. FN.
CON. C
MIN. MAX
CON. D
DEF
- constant allows for another mathematic
recalculation with option of extended
projection of measuring units
K-KILO
Constant 0,001
prefix “m” added
Constant 1
without prefix
Constant 1000
prefix “K” added
FORM. M.
Selection of decimal
point
m-MILI
- NULL
CON. E
Select multiplying
constant
CON. F
PREF. M.
FORM. M.
DESC. M.
SWCH. M.
SAVE M.
6.2.5d
Mathematic functions - decimal point
INPUTS
CHAN. I
CHANNE.
CHAN. U
MATH. F. 00000. o
INP. M.
OUTPUT.
CHAN. P
CON. A 0000. o o
SERVIC.
CHAN. F
CON. B 000. o oo
MAT. FN.
CON. C 00. o ooo
MIN. MAX
CON. D 0. o oooo
CON. E
CON. F
000000
FLOA. P.
- the instrument allows for classic projection
of a number with positioning of the DP
as well as projection with floating DP,
allowing to display a number in its most
exact form „FLOA.P.“
000000.
00000. o
DEF
0000. o o
000. o oo
PREF. M.
FORM. M.
00. o ooo
DESC. M.
0. o oooo
SWCH. M.
SAVE M.
FLOA. P.
Setting DP - XXXXXX.
Setting DP - XXXXX.x
Setting DP - XXXX.xx
Setting DP - XXX.xxx
Setting DP - XX.xxxx
Setting DP - X.xxxxx
Floating DP
INSTRUCTIONS FOR USE OM 402PWR
|
49
6
profi
SETTING
6.2.5e
Mathematic functions - measuring units
DESC. M.
INPUTS
CHAN. I
INP. M.
CHANNE.
CHAN. U
MATH. F.
OUTPUT.
CHAN. P
CON. A
SERVIC.
CHAN. F
CON. B
MAT. FN.
CON. C
MIN. MAX
CON. D
Setting projection of
description for “MAT.FN”
- projection of mesured data may be
extended (at the expense of the number
of displayed places) by two characters for
description
- description is set by shifted ASCII code,
when two first places show the set
description and two last characters their
code in period 0…95
- description is cancelled by code 00
- DEF = no description
CON. E
CON. F
!
PREF. M.
Table of signs on page 71
FORM. M.
DESC. M.
SWCH. M.
SAVE M.
6.2.5f
Select channel projection during switching
SWCH. M.
INPUTS
CHAN. I
INP. M.
YES
CHANNE.
CHAN. U
MATH. F.
NO
OUTPUT.
CHAN. P
CON. A
SERVIC.
CHAN. F
CON. B
MAT. FN.
CON. C
MIN. MAX
CON. D
CON. E
CON. F
PREF. M.
FORM. M.
DESC. M.
SWCH. M.
SAVE M.
50
|
INSTRUCTIONS FOR USE OM 402PWR
DEF
Select channel projection
during switching
- setting in this item enables the user to
select individual measuring channels,
which will be displayed during switching
channel functions „PREP. 1“
YES
NO
Projection permtted
Projection prohibited
profi
6.2.5f
SETTING
6
Mathematic functions - selection of storing data into instrument memory
SAVE M.
INPUTS
CHAN. I
INP. M.
YES
CHANNE.
CHAN. U
MATH. F.
NO
OUTPUT.
CHAN. P
CON. A
SERVIC.
CHAN. F
DEF
- by selection in this item you allow to
register values into instrument memory
- another setting in item “OUTPUT. >
MEMORY” (not in standard experiment)
NO
Measured data are
stored in the memory
Measured data are not
stored
INPUTS
Selection of evaluation
of min/max value
YES
MAT. FN.
MIN. MAX
Selection of storing data
into instrument memory
CON. F
PREF. M.
FORM. M.
DESC. M.
SWCH. M.
SAVE M.
6.2.6
Selection of evaluation of min/max value
INPUTS
CHAN. I
CHANNE.
CHAN. U
INPUTS
CHAN. I
NO
OUTPUT.
CHAN. P
FIL. I
NO
SERVIC.
CHAN. F
CHAN. U
CHAN. I
MAT. FN.
FIL. U
FIL. I
MIN. MAX
CHAN. P
FIL. P
CHAN. F
- selection of value from which the min/
max value will be calculated
DEF
CHAN. U
FIL. U
FIL. F
CHAN. P
MAT. FN.
FIL. P
COS. 02
CHAN. F
FIL. F
MAT. FN.
COS. 02
Evaluation of min/max
value is off
From “Channel I”
From “Channel I” after
digital filters processing
From “Channel U”
From “Channel U” after
digital filters processing
From “Channel P”
From “Channel P” after
digital filters processing
From “Channel F”
From “Channel F” after
digital filters processing
From “Mathematic
functions”
From auxiliary channel
„COS FI“
INSTRUCTIONS FOR USE OM 402PWR
|
51
6
profi
SETTING
6.3
Setting „PROFI“ - OUTPUTS
In this menu it is possible to set parame
ters of the instrument output signals
INPUTS MEMORY
CHANNE.
LIMITS
OUTPUT.
DATA
SERVIC.
AN. OUT.
MEMORY
LIMITS
DATA
Setting data logging into
memory
Setting type and
parameters of limits
Setting type and
parameters of data
output
DISP.
AN. OUT.
Setting type and
parameters of analog
output
DISP.
6.3.1a
Selection of mode of data logging into instrument memory
REWRIT.
INPUTS MEMORY
52
REWRIT.
YES
CHANNE.
LIMITS
START
NO
OUTPUT.
DATA
STOP
SERVIC.
AN. OUT.
PERIOD.
DISP.
TRIGER
|
Setting display projection
and brightness
INSTRUCTIONS FOR USE OM 402PWR
DEF
Selection of the mode of
data logging
- selection of the mode in the event of full
instrument memory
Rewriting values
prohibited
Rewriting values
YES
permitted, the oldest get
rewritten by the latest
NO
profi
6.3.1b
SETTING
6
Setting data logging into instrument memory - RTC
START
INPUTS MEMORY
- time format HH.MM.SS
REWRIT.
CHANNE.
LIMITS
START
OUTPUT.
DATA
STOP
SERVIC.
AN. OUT.
PERIOD.
Start of data logging into
instrument memory
STOP
Stop data logging into
instrument memory
- time format HH.MM.SS
PERIOD.
Period of data logging
into instrument memory
- determines the period in which values will
be logged in an interval delimited by the
time set under items START and STOP
DISP.
- time format HH.MM.SS
- records are made on a daily basis in
selected interval and period
- item not displayed if “STORE” is selected
in menu (Input > EXT. IN.)
6.3.1c
Setting data logging into instrument memory - FAST
TRIGER
INPUTS MEMORY
REWRIT.
10%
CHANNE.
TRIGER
50%
LIMITS
OUTPUT.
DATA
90%
SERVIC.
AN. OUT.
ROLL
DEF
Setting logging data into
inst. memory
- logging data into inst. memory is
governed by the folowing selection, which
determines how many percent of the
memory is reserved for data logging prior
to initiation of trigger imputse
- initiation is on ext. input or control key
DISP.
Reser. of 10 % memory
prior init. of data logging
Reser. of 50 % memory
50%
prior init. of data logging
Reser. of 90 % memory
90%
prior init. of data logging
After initiation of data
ROLL
logging the memory is
cyclically transcribed
10%
INSTRUCTIONS FOR USE OM 402PWR
|
53
6
profi
SETTING
6.3.2a
Selection of input for limits evaluation
INP. L1
INPUTS MEMORY
LIM 1
INP. L1
CHAN. I
CHANNE.
LIMITS
LIM 2
MOD. L.1
FIL. I
OUTPUT.
DATA
LIM 3
TYP. L.1
CHAN. U
CHAN. I
SERVIC.
AN. OUT.
LIM 4
LIM. L.1
FIL. U
FIL. I
HYS. L.1
CHAN. P
ON L.1
FIL. P
OFF L.1
CHAN. F
PER. L.1
FIL. F
CHAN. P
TIM. L.1
MAT. FN.
FIL. P
DISP.
COS. 02
- selection of value from which the limit will
be evaluated
CHAN. U
DEF
FIL. U
CHAN. F
MIN
MAX
FIL. F
MAT. FN.
COS. 02
MIN.
MAX.
!
Setting is identical for LIM 2, LIM 3 and LIM 4
!
“COS. 02” is derived from “COS. FI.” when the
range -1…1 is tranferred to 0…2 for evaluation
54
|
INSTRUCTIONS FOR USE OM 402PWR
Selection evaluation
of limits
From “Channel I”
From “Channel I” after
digital filters processing
From “Channel U”
From “Channel U” after
digital filters processing
From “Channel P”
From “Channel P” after
digital filters processing
From “Channel F”
From “Channel F” after
digital filters processing
From “Mathematic
functions”
From auxiliary channel
„COS FI“
From “Min. value”
From “Max. value”
profi
6.3.2b
SETTING
6
Selection of type of limit
MOD. L.1
INPUTS MEMORY
LIM 1
INP. L1
HYSTER
CHANNE.
LIMITS
LIM 2
MOD. L.1
FROM. .
OUTPUT.
DATA
LIM 3
TYP L.1
DOSING
SERVIC.
AN. OUT.
LIM 4
LIM. L.1
DISP.
DEF
HYSTER
Selection the type of limit
Limit is in mode “Limit,
hysteresis, delay”
- for this mode the parameters of “LIM. L.”
are set, at which the limit will shall react,
“HYS. L.” the hysteresis range around the
limit (LIM ±1/2 HYS) and time “TIM. L.”
determining the delay of relay switch-on
HYS. L.1
FROM.
ON L.1
Frame limit
- for this mode the parameters are set for
interval “ON. L.” the relay switch-on
and “OFF. L.” the relay switch-off
OFF L.1
PER. L.1
TIM. L.1
DOSING
Dose limit
(periodic)
- for this mode the parameters are set for
“PER. L.” determining the limit value as
well as its multiples at which the output is
active and “TIM. L.” indicating the time
during which is the output active
!
Setting is identical for LIM 2, LIM 3 and LIM 4
6.3.2c
Selection of type of output
TYP. L.1
INPUTS MEMORY
LIM 1
INP. L.1
CLOSE
CHANNE.
LIMITS
LIM 2
MOD. L.1
OPEN
OUTPUT.
DATA
LIM 3
TYP L.1
SERVIC.
AN. OUT.
LIM 4
LIM. L.1
DISP.
HYS. L.1
ON L.1
DEF
CLOSE
OPEN
Selection of type of
output
Output switches on when
condition is met
Output switches off when
condition is met
!
Setting is identical for LIM 2, LIM 3 and LIM 4
OFF L.1
PER. L.1
TIM. L.1
INSTRUCTIONS FOR USE OM 402PWR
|
55
6
profi
SETTING
6.3.2d
Setting values for limits evaluation
LIM. L1
INPUTS MEMORY
LIM 1
INP. L.1
CHANNE.
LIM 2
MOD. L.1
LIMITS
OUTPUT.
DATA
LIM 3
TYP L.1
SERVIC.
AN. OUT.
LIM 4
LIM. L.1
DISP.
Setting limit for switch-on
- for type “HYSTER”
HYS. L.1
Setting hysteresis
- for type “HYSTER”
- indicates the range around the limit (in
both directions, LIM. ±1/2 HYS.)
HYS. L.1
ON. L.1
ON L.1
Setting the outset of the
interval of limit switch-on
- for type “FROM”
OFF L.1
OFF. L.1
PER. L.1
Setting the end of the
interval of limit switch-on
- for type “FROM”
TIM. L.1
PER. L.1
Setting the period of limit
switch-on
- for type “DOSING”
!
TIM. L.1
- for type “HYSTER” and “DOSING
Setting is identical for LIM 2, LIM 3 and LIM 4
6.3.3a
Selection of data output baud rate
BAUD
INPUTS MEMORY
BAUD
600
CHANNE.
ADDR.
1200
DATA AD.-MOD.
2400
1200
4800
2400
OUTPUT.
SERVIC.
LIMITS
AN. OUT.
DISP.
PROT.
9600
600
DEF
4800
19200
38400
9600
57600
19200
115200
38400
230400
57600
115200
230400
56
|
Setting the time switch-on
of the limit
INSTRUCTIONS FOR USE OM 402PWR
Selection of data output
baud rate
Rate - 600 Baud
Rate - 1 200 Baud
Rate - 2 400 Baud
Rate - 4 800 Baud
Rate - 9 600 Baud
Rate - 19 200 Baud
Rate - 38 400 Baud
Rate - 57 600 Baud
Rate - 115 200 Baud
Rate - 230 400 Baud
profi
6.3.3b
SETTING
6
Setting instrument address
ADDR.
Setting instrument
address
INPUTS MEMORY
BAUD
- setting in range 0…31
CHANNE.
ADDR.
- DEF = 00
OUTPUT.
SERVIC.
LIMITS
DATA AD.-MOD.
AN. OUT.
PROT.
AD.-MOD.
DISP.
Setting instrument
address - MODBUS
- setting in range 1…247
- DEF = 1
6.3.3c
Selection of data output protocol
PROT.
INPUTS MEMORY
BAUD
ASCII
CHANNE.
ADDR.
M.BUS
DATA AD.-MOD.
MODBUS
OUTPUT.
SERVIC.
LIMITS
AN. OUT.
DISP.
PROT.
DEF
ASCII
M. BUS
MODBUS
Selection of the type of
analog output
Data protocol
ASCII
Data protocol
DIN MessBus
Data protocol
MODBUS-RTU
- option is available only for RS 485
INSTRUCTIONS FOR USE OM 402PWR
|
57
6
profi
SETTING
6.3.4a
Selection of input for analog output
INP. A. O .
INPUTS MEMORY
INP. A. O .
NO
CHANNE.
LIMITS
TYP A. O .
CHAN. I
OUTPUT.
DATA
MIN A. O .
FIL. I
NO
SERVIC.
AN. OUT.
MAX A. O .
CHAN. U
CHAN. I
DISP.
- select value from which analog output
will be evaluated
FIL. U
FIL. I
CHAN. P
FIL. P
CHAN. F
DEF
CHAN. U
FIL. U
FIL. F
MAT. FN.
COS. 02
MIN
CHAN. P
FIL. P
CHAN. F
MAX
FIL. F
MAT. FN.
COS. 02
MIN
!
MAX
“COS. 02” is derived from “COS. FI.” when the
range -1…1 is tranferred to 0…2 for evaluation
58
|
INSTRUCTIONS FOR USE OM 402PWR
Select evaluation of
analog output
AO evaluation is off
From “Channel I”
From “Channel I” after
digital filters processing
From “Channel U”
From “Channel U” after
digital filters processing
From “Channel P”
From “Channel P” after
digital filters processing
From “Channel F”
From “Channel F” after
digital filters processing
From “Mathematic
functions”
From auxiliary
channel„COS FI“
AO evaluation
from “Min.value”
AO evaluation
from “Max.value”
DEF “Channel I” - FIL. I
DEF “Channel U” - FIL. U
DEF “Channell P” - FIL. P
profi
6.3.4b
SETTING
Selection of the type of analog output
TYP A. O .
INPUTS MEMORY
INP. A. O . 0-20mA
CHANNE.
LIMITS
TYP A. O . Er 4-20
OUTPUT.
DATA
MIN A. O . 4-20mA
SERVIC.
AN. OUT.
MAX A. O .
DISP.
0-5mA
0-20mA
DEF
E 4-20
0-10V
Selection of the type of
analog output
Type - 0…20 mA
Type - 4…20 mA
- with indication of error statement
(< 3,0 mA)
0-2V
0-5V
4-20mA
0-5mA
0-2V
0-5V
0-10V
6.3.4c
6
Type - 4…20 mA
Type - 0…5 mA
Type - 0…2 V
Type - 0…5 V
Type - 0…10 V
Setting the analog output range
AN. OUT.
INPUTS MEMORY
INP. AO.
CHANNE.
LIMITS
TYP. A. O .
OUTPUT.
DATA
MIN A. O .
SERVIC.
AN. OUT.
MAX A. O .
DISP.
Setting the analog
output range
- analog output is isolated and its value
corresponds with displayed data. It is fully
programmable, i.e. it allows to assign the
AO limit points to two arbitrary points of
the entire measuring range
MIN A. O .
Assigning the display
value to the beginning of
the AO range
- range of the setting is -99999…999999
- DEF = 0
MAX A. O .
Assigning the display
value to the end of the
AO range
- range of the setting is -99999…999999
- DEF = 100
INSTRUCTIONS FOR USE OM 402PWR
|
59
6
profi
SETTING
6.3.5a
Selection of input for display projection
PERM.
INPUTS MEMORY
CHANNE.
LIMITS
OUTPUT.
DATA
SERVIC.
PERM.
CHAN. I
BRIGHT
FIL. I
- selection of value which will be shown on
the instrument display
CHAN. U
CHAN. I
AN. OUT.
FIL. U
FIL. I
DISP.
CHAN. P
FIL. P
DEF
CHAN. U
FIL. U
CHAN. F
FIL. F
CHAN. P
MAT. FN.
FIL. P
COS. 02
CHAN. F
MIN.
FIL. F
MAX.
MAT. FN.
!
COS. 02
“COS. 02” is derived from “COS. FI.” when the
range -1…1 is tranferred to 0…2 for evaluation
MIN.
MAX.
6.3.5b
INPUTS MEMORY
CHANNE.
LIMITS
PERM.
100%
BRIGHT.
75%
OUTPUT.
DATA
50%
SERVIC.
AN. OUT.
25%
DISP.
0%
DEF
From “Channel I” after
digital filters processing
From “Channel U”
From “Channel U” after
digital filters processing
From “Channel P”
From “Channel P” after
digital filters processing
From “Channel F”
From “Channel F” after
digital filters processing
From “Mathematic
functions”
From auxiliary channel
„COS FI“
From “Min. value”
From “Max. value”
0%
Display is off
- after keystroke display turns on for 10 s
50%
75%
100%
INSTRUCTIONS FOR USE OM 402PWR
Selection of display
brightness
- by selecting display brightness we may
appropriately react to light conditions in
place of instrument location
25%
|
From “Channel I”
Selection of display brightness
BRIGHT
60
Selection display
projection
Display brightness - 25 %
Display brightness - 50 %
Display brightness - 75 %
Display brightness - 100 %
profi
SETTING
INSTRUCTIONS FOR USE OM 402PWR
|
6
61
6
profi
SETTING
6.4
Setting “PROFI” - SERVICE
The instrument service functions are set
in this menu
INPUTS
MENU
CHANNE.
RESTOR.
OUTPUT.
LANG.
SERVIC.
N. PASS.
MENU
RESTOR.
calibration
IDENT.
LANG.
N. PASS.
IDENT.
6.4.1
INPUTS
MENU
LIGHT
CHANNE.
RESTOR.
PROFI
OUTPUT.
LANG.
SERVIC.
N. PASS.
DEF
IDENT.
!
Change of setting is valid upon next access into
menu
INSTRUCTIONS FOR USE OM 402PWR
Selection of menu type LIGHT/PROFI
- enables setting the menu complexity
according to user needs and skills
LIGHT
Active LIGHT menu
- simple programming menu, contains only
items necessary for configuration and
instrument setting
- linear menu > items one after another
PROFI
|
Language version of
instrument menu
Setting new access
password
Instrument identification
Selection of type of programming menu
MENU
62
Selection of menu type
LIGHT/PROFI
Restore instrument
manufacture setting and
Active PROFI menu
- complete programming menu for expert
users
- tree menu
profi
6.4.2
SETTING
6
Restoration of manufacture setting
Restoration of
manufacture calibration
of the instrument
CALIB.
INPUTS
MENU
CALIB.
CHANNE.
RESTOR.
SETTIN.
OUTPUT.
LANG.
SERVIC.
N. PASS.
YES
- in the event of error setting or calibration
it is possible to return to manufacture
calibration. Prior to execution of changes
you will be asked to confirm your
selection „YES“
IDENT.
SETTIN.
INPUTS
MENU
CALIB.
POWER
CHANNE.
RESTOR.
SETTIN.
VOLTAG.
OUTPUT.
LANG.
SERVIC.
N. PASS.
Return to manufacture
setting of the instrument
- reading manufacture setting
(items marked DEF)
- presetting “source” for further evaluation
(e.g.: selection “VOLTAGE” > evaluation
of limits, analog, min/max. value, etc...
from “Channel U”
- prior to execution of changes you will be
asked to confirm your selection „YES“
CURREN.
IDENT.
POWER
VOLTAG.
CURREN.
Setting for power
measuring - “Channel P”
Setting for voltage
measuring - “Channel U”
Setting for measuring
current “Channel I”
Restore
Jobs performed
cancels USER menu rights
deletes table of items order in USER - LIGHT menu
adds items from manufcture to LIGHT menu
deletes data stored in FLASH
cancels or linearization tables
clears tare
clears conduct resistances
restore manufacture calibration
restore manufacture setting
Calibration
Setting
INSTRUCTIONS FOR USE OM 402PWR
|
63
6
profi
SETTING
6.4.3
Selection of instrument menu language version
LANG.
INPUTS
MENU
CZECH
CHANNE.
RESTOR.
ENGL.
OUTPUT.
LANG.
SERVIC.
N. PASS.
DEF
CZECH
ENGL.
Selection of instrument
menu language version
Instrument menu is in
Czech
Instrument menu is in
English
IDENT.
6.4.4
Setting new access password
Setting new password
for access to LIGHT
and PROFI menu
N. PASS.
INPUTS
MENU
CHANNE.
RESTOR.
OUTPUT.
LANG.
SERVIC.
N. PASS.
IDENT.
6.4.5
- this selection enables changing number
code that blocks the access into LIGHT
and PROFI Menu.
- range of the number code is 0…9999
- universal password in the event of loss is
„8177“
Instrument identification
IDENT.
INPUTS
MENU
CHANNE.
RESTOR.
OUTPUT.
LANG.
SERVIC.
N. PASS.
IDENT.
64
|
INSTRUCTIONS FOR USE OM 402PWR
Projection of instrument
SW version
- display shows type identification of the
instrument, SW number, SW version
and current input setting (Mode)
- if the SW version reads a letter on first
position, it is a customer SW
profi
SETTING
INSTRUCTIONS FOR USE OM 402PWR
|
6
65
7
user
SETTING
7.0
Setting items into “USER” menu
• USER menu is designed for users who need to change only several items of the setting without the option to change the
primary instrument setting (e.g. repeated change of limit setting)
• there are no items from manufacture permitted in USER menu
L1
• on items indicated by inverse triangle
• setting may be performed in LIGHT or PROFI menu, with the USER menu then overtaking the given menu structure
USER
• For user operation
• Menu items are set by the user
(Profi/Light) as per request
Setting
• Access is not password protected
Setting
flashing legend - current setting is displayed
---
NO
YES
SHOW
66
|
+
NO
YES
SHOW
item will not be displayed in USER menu
item will be displayed in USER menu with editing option
item will be solely displayed in USER menu
INSTRUCTIONS FOR USE OM 402PWR
return to item
---
user
SETTING
7
Setting sequence of items in “USER” menu
In compiling USER menu from active LIGHT menu the items (max. 10) may be assigned a sequence, in which they will be
projected in the menu
setting projection sequence
---
+
0
Example:
Into USER menu were selected these items
(keys + ) > CL. TAR., LIM 1, LIM 2, LIM 3, for which we have preset this sequence (keys
CL. TAR.
LIM 1
LIM 2
LIM 3
+
):
5
0 (sequence not determined)
2
1
Upon entering USER menu
(key ) items will be projected in the following sequence: LIM 3 > LIM 2 > CL.TAR. > LIM 1
INSTRUCTIONS FOR USE OM 402PWR
|
67
8
DATA PROTOCOL
The instruments communicate via serial line RS232 or RS485. For communication they use the ASCII protocol. Communication runs in the following format:
ASCII:
8 bit, no parity, one stop bit
DIN MessBus: 7 bit, even parity, one stop bit
The transfer rate is adjustable in the instrument menu. The instrument address is set in the instrument menu in the range of
0 ÷ 31. The manufacture setting always presets the ASCII protocol, rate of 9600 Baud, address 00. The type of line used
- RS232 / RS485 - is determined by an output board automatically identified by the instrument.
The commands are described in specifications you can find at na www.orbit.merret.cz/rs or in the OM Link program.
DETAILED DESCRIPTION OF COMMUNICATION VIA SERIAL LINE
Type
232
Event
Confirmation of data acceptannce
(PC) - OK
Confirmation of data acceptance
(PC) - Bad
485
485
Data transmission (instrument)
232
485
Data solicitation (PC)
Protocol
ASCII
MessBus
ASCII
MessBus
A
A
<CR>
#
A
A
<CR>
<SADR> <ENQ>
ASCII
>
D
(D)
(D)
(D)
(D)
(D)
(D)
(D)
(D)
(D)
<CR>
<SADR>
D
(D)
(D)
(D)
(D)
(D)
(D)
(D)
(D)
(D)
<ETX>
ASCII
>
D
(D)
(D)
(D)
(D)
(D)
(D)
(D)
(D)
(D)
<CR>
MessBus
<SADR>
D
(D)
(D)
(D)
(D)
(D)
(D)
(D)
(D)
(D)
<ETX>
<BCC>
<DLE>
1
(D)
<CR>
(D)
<CR>
MessBus
<NAK>
485
232
ASCII
#
A
A
N
P
(D)
(D)
(D)
(D)
(D)
(D)
MessBus
<STX>
$
N
P
(D)
(D)
(D)
(D)
(D)
<ETX>
<BCC>
ASCII
#
A
A
N
P
(D)
(D)
(D)
(D)
(D)
(D)
MessBus
<SADR>
$
N
P
(D)
(D)
(D)
(D)
(D)
<ETX>
<BCC>
OK
!
A
A
<CR>
Bad
?
A
A
<CR>
ASCII
232
<SADR> <ENQ>
Command confirmation (instrument)
ASCII
MessBus
485
Messbus
No - data is transmitted permanently
OK
!
A
A
<CR>
Bad
?
A
A
<CR>
OK
<DLE>
1
Bad
<NAK>
A
A
<CR>
A
A
<CR>
Instrument identification
#
A
A
1Y
<CR>
HW identification
#
A
A
1Z
<CR>
One-time transmission
#
A
A
7X
<CR>
Repeated transmission
#
A
A
8X
<CR>
68
|
485
!
?
Command confirmati (instrument) - Bad
<BCC>
<EADR> <ENQ>
Confirmation of address (instrument)
Command confirmation (inst.) - OK
#
No - data is transmitted permanently
MessBus
Sending address (PC) prior command
Command transmission (PC)
Transmitted data
MessBus
INSTRUCTIONS FOR USE OM 402PWR
DATA PROTOCOL
LEGEND
#
A
RELAY, TARE
35
A
23H
0…31
Command beginning
Two characters of instrument address
(sent in ASCII - tens and units, e.g. “01“,
“99“ universal
Tare
Change
relay 3/4
0
0
0
1
0
0
0
R
0
1
0
0
S
1
1
0
0
T
0
0
1
0
Data - usually characters “0“…“9“, “-“,
“.“; (D) - dp. and (-) may prolong data
U
1
0
1
0
Relay and tare status
V
0
1
1
0
Carriage return
<SP>
32
20H
Space
Number and command - command code
30H…3FH
Relay 2
0
0DH
D
Relay 1
P
13
N, P
Sign
Q
<CR>
R
!
33
21H
Positive confirmation of command (ok)
W
1
1
1
0
?
63
3FH
Negative confirmation of command
(point)
p
0
0
0
1
q
1
0
0
1
>
62
3EH
Beginning of transmitted data
r
0
1
0
1
<STX>
2
02H
Beginning of text
s
1
1
0
1
<ETX>
3
03H
End of text
t
0
0
1
1
u
1
0
1
1
<SADR>
addresa +60H
Prompt to send from address
<EADR>
addresa +40H
Prompt to accept command at address
v
0
1
1
1
<ENQ>
5
05H
Terminate address
w
1
1
1
1
<DLE>1
16
49
10H
31H
Confirm correct statement
<NAK>
21
15H
Confirm error statement
<BCC>
8
Check sum -XOR
Relay status is generated by command #AA6X
<CR>.
The instrument immediately returns the value in the
format >HH <CR>, where HH is value in HEX format
and range 00H…FFH. The lowest bit stands for „Relay
1“, the highest for „Relay 8“
INSTRUCTIONS FOR USE OM 402PWR
|
69
9
ERROR STATEMENTS
ERROR
CAUSE
ELIMINATION
E. D. Un.
Number is too small (large negative) to be
displayed
change DP setting, channel constant setting
Number is too large to be displayed
change DP setting, channel constant setting
Number is outside the table range
increase table values, change input setting (channel constant setting)
Number is outside the table range
increase table values, change input setting (channel constant setting)
E. I. Un.
Input quantity is smaller than permitted input
quantity range
change input signal value or input (range) setting
E. I. Ov.
Input quantity is larger than permitted input
quantity range
change input signal value or input (range) setting
A part of the instrument does not work properly
send the instrument for repair
Data in EEPROM corrupted
perform restoration of manufacture setting, upon
repeated error statement send instrument for
repair
Data in EEPROM outside the range
perform restoration of manufacture setting, upon
repeated error statement send instrument for
repair
Memory was empty
(presetting carried out)
upon repeated error statement send instrument for
repair, possible failure in calibration
E. D. Ov.
E. T. Un..
E. T. Ov.
E. HW.
E. EE
E.. DATA
E. CLR.
70
|
INSTRUCTIONS FOR USE OM 402PWR
TABLE OF SIGNS
10
The instrument allows to add two descriptive characters to the classic numeric formats (at the expense of the number of displayed places). The setting is performed by means of a shifted ASCII code. Upon modification the first two places display the
entered characters and the last two places the code of the relevant symbol from 0 to 95. Numeric value of given character
equals the sum of the numbers on both axes of the table.
Description is cancelled by entering characters with code 00
INSTRUCTIONS FOR USE OM 402PWR
|
71
11
TECHNICAL DATA
Watch-dog:
Calibration:
INPUT
Range is adjustable depending on the order
Voltage:
0…10 V
0…120 V
0…250 V
0…450 V
150 kOhm
930 kOhm
730 kOhm
930 kOhm
Input 2 - U
Input 3 - U
Input 2 - U
Input 3 - U
Current:
0...60 mV
0...150 mV
0...300 mV
0...1 A
0...2,5 A
0...5 A
21 kOhm
21 kOhm
21 kOhm
< 150 mV
< 150 mV
< 150 mV
Input 1 - I
Input 1 - I
Input 1 - I
Input 1 - I
Input 1 - I
Input 1 - I
Input frequency:
0…400 Hz (up to amplitude 8 V)
Measured quantities Voltage (VRMS)
Current (ARMS)
Active power (P)
Frequency (Hz)
swith calclation
Idle power (Q)
Apparent power (S)
Power factor (cos ϕ)
PROJECTION
Display:
Projection:
Decimal point:
Brightness:
999999, intensive red or green
14-ti segment LED, digit height 14 mm
±9999 (-99999…999999)
adjustable - in menu
adjustbale - in menu
INSTRUMENT ACCURACY
TK:
100 ppm/°C
Accuracy:
±0,3 % of range + 1 digit
U, I, P, F
±0,6 % of range + 1 digit
S
±0,9 % of range + 1 digit
Q, Cos Fi
Above accuracies apply for rate 2,5 meas./s
Rate:
0,6…5 measurements/s
Overload capacity: 10x (t < 100 ms) not for 450 V and 5 A,
2x (long-term)
Linearisation:
by linear interpolation in 50 points
- solely via OM Link
Digital filters:
Averaging, Floating average, Exponential filter, Rounding
Functions:
Tare - display resetting
Hold - stop measuring (at contact)
Lock - control key locking
MM - min/max value
Mathematic functions
OM Link:
company communication interface for setting, operation
and update of instrument SW
72
|
INSTRUCTIONS FOR USE OM 402PWR
COMPARATOR
Type:
Mode:
Limita:
Hysteresis:
Delay:
Outputs:
Relay:
reset after 400 ms
at 25°C and 40 % of r.h.
digital, adjustable in menu
Hysteresis, From, Dosing
-99999…999999
0…999999
0…99,9 s
2x relays with switch-on contact (Form A)
(230 VAC/30 VDC, 3 A)*
2x relays with switch-off contact (Form C)
(230 VAC/50 VDC, 3 A)*
2x SSR (250 VAC/ 1 A)*
2x/4x open collector (30 VDC/100 mA)
2x bistabil relays (250 VAC/250 VDC, 3 A/0,3 A)*
1/8 HP 277 VAC, 1/10 HP 125 V, Pilot Duty D300
DATA OUTPUTS
Protocols:
ASCII, DIN MessBus, MODBUS, PROBUS
Data format:
8 bit + no parity + 1 stop bit (ASCII)
7 bit + even parity + 1 stop bit (MessBus)
Rate:
600…230 400 Baud
RS 232:
isolated, two-way communication
RS 485:
isolated, two-way communication,
addressing (max. 31 instruments)
PROFIBUS
Data protocol SIEMENS
ANALOG OUTPUTS
Type:
isolated, programmable with resolution of max.10 000
points, analog output corresponds with displayed data, type
and range are adjustable
Non-linearity:
0,2 % of range
TC:
100 ppm/°C
Rate:
response to change of value < 150 ms
Voltage:
0…2 V/5 V/10 V
Curernt:
0…5/20 mA/4…20 mA
- compensation of conduct to 500 Ohm/12 V
or 1 000 Ohm/24 V
MEASURED DATA RECORD
Type RTC:
time-controlled logging of measured data into instrument
memory, allows to log up to 250 000 values
Type FAST:
fast data logging into instrument memory, allows to log up
to 8 000 values at a rate of 40 records/s
Transmission:
via data output RS 232/485 or via OM Link
* values apply for resistance load
TECHNICAL DATA
11
POWER SUPPLY
Options:
10…30 V AC/DC, 10 VA, isolated,
- fuse inside (T 4000 mA)
80…250 V AC/DC, 10 VA, isolated
- fuse inside (T 630 mA)
MECHANIC PROPERTIES
Material:
Noryl GFN2 SE1, incombustible UL 94 V-I
Dimensions:
96 x 48 x 120 mm
Panel cut-out:
90,5 x 45 mm
OPERATING CONDITIONS
Connection:
connector terminal board, conductor
cross-section <1,5 mm2 /<2,5 mm2
Stabilisation period: within 15 minutes after switch-on
Working temp.:
0°…60°C
Storage temp.:
-10°…85°C
Cover:
IP65 (front panel only)
Construction:
safety class I
Overvoltage category: EN 61010-1, A2
Insulation resistance: for pollution degree II, measurement category III
instrum.power supply > 670 V (PI), 300 V (DI)
Input/output > 300 V (PI), 150 (DI)
EMC:
EN 61000-3-2+A12; EN 61000-4-2, 3, 4, 5, 8, 11;
EN 550222, A1, A2
PI - Primary insulation, DI - Double insulation
INSTRUCTIONS FOR USE OM 402PWR
|
73
12
INSTR. DIMENSIONS AND INSTALLATION
Front view
Panel cut
1
90,5 mm
2
45 mm
48 mm
96 mm
T
Side view
119 mm
13,5 mm
Panel thickness: 0,5...20 mm
1
Instrument installation
2
1. insert the instrument into the panel cut-out
2. fit both travellers on the box
3. press the travellers close to the panel
2
1
1
2
74
|
INSTRUCTIONS FOR USE OM 402PWR
Instrument disassembly
1. slide a screw driver under the traveller wing
2. turn the screw driver and remove the traveller
3. take the instrument out of the panel
CERTIFICATE OF GUARANTEE
Product
13
OM 402PWR
Type
....................
Manufacturing No.
....................
Date of sale
....................
A guarantee period of 60 months from the date of sale to the user applies to this instrument.
Defects occuring during this period due to manufacture error or due to material faults shall be eliminated free of charge.
For quality, function and construction of the instrument the guarantee shall apply provided that the instrument was connected
and used in compliance with the instructions for use.
The guarantee shall not apply to defects caused by:
- mechanic damage
- transportation
- intervention of unqualified person incl. the user
- unavoidable event
- other unprofessional interventions
The manufacturer performs guarantee and post.guarantee repairs unless provided for otherwise.
Stamp, signature
Y E A R S
INSTRUCTIONS FOR USE OM 402PWR
|
75
DECLARATION OF CONFORMITY
Company:
ORBIT MERRET, spol. s r.o.
Klánova 81/141, 142 00 Prague 4, Czech Republic, IDNo: 00551309
Manufactured:
ORBIT MERRET, spol. s r.o.
Vodňanská 675/30, 198 00 Prague 9, Czech Republic
declares at its full responsibility that the product presented hereunder meets all technical requirements, is safe for use when
utilised under the terms and conditions determined by ORBIT MERRET, spol.s r.o. and that our company has taken all measures
to ensure conformity of all products of the type listed hereunder, which are being brought out to the market, with technical
documentation and requirements of the appurtenant statutory orders.
Product:
4-digit programmable panel instrument
Type:
OM 402
Version:
UNI, PWR
Conformity is assessed pursuant to the following standards:
El. safety:
EN 61010-1
EMC:
EN 50131-1, chapter 14 and chapter 15
EN 50130-4, chapter 7
EN 61000-4-11
EN 50130-4, chapter 8
EN 61000-4-11
EN 50130-4, chapter 9
EN 61000-4-2
EN 50130-4, chapter 10
EN 61000-4-3
EN 50130-4, chapter 11
EN 61000-4-6
EN 50130-4, chapter 12
EN 61000-4-4
EN 50130-4, chapter 13
EN 61000-4-5
EN 50130-5, chapter 20
prEN 50131-2-1, par. 9.3.1
EN 61000-4-8
EN 61000-4-9
EN 61000-3-2 ed. 2:2001
EN 61000-3-3: 1997, Cor. 1:1998, Z1:2002
EN 55022, chapter 5 and chapter 6
and Ordinance on:
El. safety:
No. 168/1997 Coll.
EMC:
No. 169/1997 Coll.
The evidence are the protocols of authorized and accredited organizations:
VTÚE Praha, experimental laboratory No. 1158, accredited by ČIA
VTÚPV Vyškov, experimental laboratory No. 1103, accredited by ČIA
Place and date of issue:
Prague, 18. March 2006
Miroslav Hackl v.r.
Company representative
Mode of asses. of conformity §12, par. 4 b, d Act No. 22/1997 Coll.
TECHDOK - OM 402PWR - 2007 - 2v0 - en - V3
76
|
INSTRUCTIONS FOR USE OM 402PWR