Download Version 3.1 System Manual
Transcript
USING HeRO® ES Alarms The HeRO system includes a visual high score alarm that is designed to draw attention to patients with elevated risk. The alarms activate when a patient’s HeRO Score rises to, or beyond, a threshold value of 2.0. The alarm state is indicated by one of the two Alarm Symbols shown below. State Off Active Appearance Symbol not shown Symbol shown with exclamation points flashing Paused Symbol shown with exclamation points solid An active alarm may be paused by tapping (or clicking on) the flashing symbol or the surrounding area. The alarm will now display the Paused symbol, which will remain visible until the alarm resets. When the patient’s score drops below the threshold, an Active or Paused alarm will reset to the Off state, and the alarm symbol will disappear. The alarm is automatically re-armed such that it will activate if the threshold value is exceeded in the future. The alarm system in HeRO is implemented in software and does not require separate testing. If HeRO is operating, the alarm system is functional. - 20 -