Rosetta-ALICE User Manual
time-resolved acquisitions using relatively bright targets in the Alice FOV. At the conclusion of this
acquisition period the total amount of generated data can be further reduced by selecting only events that
have occurred within up to eight separate windows for downlink.
Count Rate Mode. In this mode, the acquisition memory is again used as a one-dimensional linear
array of 32,768 entries. The count rate mode is designed to periodically (configurable between 3 ms and
12 s) collect the total detector array count rate sequentially in the linear memory array, as if the entire
instrument were an FUV photometer. This mode allows for high count rates from the detector (up to 10
kHz), without rapid fill up of the array. It does not, however, retain any spatial or spectral information for
broadband photometric studies. Depending on the required periodic acquisition rate, total acquisition
durations of up to 98 seconds to 100 hours are possible.
3.1.7
Detector Acquisition Durations
The actual start and end time of any acquisition is reported in the header of the science data dump,
both specified in instrument (spacecraft) MET. The Alice post-processing chain (i.e. Lima, Mike and
November) includes this information in the produced FITS files and calculates the acquisition duration
from these numbers as the difference between the start and stop times. For histogram and pixel list
acquisitions this calculated number represents the best available knowledge on the actual duration of the
acquisition.
The acquisition duration is not always exactly the same as the commanded acquisition time. This is
easy to understand for special cases where an acquisition was terminated early either by user command or
because a safety condition occurred, but also for the nominal acquisitions there will be a difference
between commanded and actual acquisition time. The reason for this is that the instrument software
controls the acquisition hardware in such a way that also adds a kind of granularity to the acquisition
duration, which is dependant on the specific acquisition mode. For the four different cases the
consequences of this quantization will be discussed.
These timing measurements apply to each of the separate acquisitions within a MAD commanded
(multi) acquisition sequence. The start time of an acquisition within a MAD multi acquisition sequence
with a fixed cycle duration is checked every 0.5 seconds. This means that the start of an exposure within
a multiple acquire cycle may occur up to 0.5 seconds later than expected from the commanded cycle
duration. This start time error is not cumulative though; the start times of the multiple exposures are all
related to the initial acquisition command start time.
3.1.7.1 Histogram Acquisition
During a histogram acquisition, in parallel to the histogram acquisition a coarse count rate data set of
up to 100 elements is acquired. This count rate data is included in the header of the data dump and the
Alice post-processing chain places this data in the second extension of the generated FITS file. The
minimum time resolution for this count rate acquisition is 100 ms for acquisition durations less than 10
seconds and the maximum is 25.5 seconds for acquisitions of 2550 seconds or more. For shorter
acquisitions, the 100-element count rate buffer is only partially filled and for acquisitions longer than
2550 seconds (42.5 minutes) only the initial part of the acquisition is covered in the count rate
measurements. This count rate measurement interval also determines the granularity of the histogram
acquisition end time check. This means that for short histogram acquisitions the commanded end time is
checked every 100 ms but for longer acquisitions the check is performed less frequently. The worst case
situation occurs for acquisitions longer than 2550 seconds for which the termination condition is only
checked once every 25.5 seconds, this means that worst case such an acquisition could continue up to
25.5 seconds longer than commanded. Except for the very short acquisitions (less than 10 seconds),
where the acquisition duration may last up to 100 ms longer than planned, the maximum acquisition
extension is always less than 0.4% of the commanded acquisition duration. In all cases the reported start
-40-