Download Waters CapLC Autosampler
Transcript
Chapter 2 Waters Systems Waters 2790/2795 Column Page Figure 2.33 Column page Position This field allows the column to be selected for the method. The options available will depend on the column setup on the Waters 2790/2795 Separations Module. If only one column is installed then this box will display Column 1 and cannot be changed. For other configurations this box will allow the selection of a column (between 1 and 6 depending on configuration) or No change from the drop down list box. Selecting a numbered column will use this column for the method, selecting No Change will use the column defined in the last method used to acquire a sample. See the Waters 2790/2795 Separations Module Operator’s Guide for more information on the column selection valve. Equilibrium Time Enter the time required to reach equilibrium (i.e. run in initial conditions), before performing an injection, after a column change. Range: 0.00 to 999.99 minutes. Temperature Set Enter the target operating temperature for the optional column heater. This value must be at least 5 °C above ambient. Range: 20 to 60 °C. Temperature Limit Enter the maximum deviation in column temperature allowed. If this is exceeded the current acquisition will stop and the LC Status error light, on the MassLynx screen, will turn red. Range: ±1 to ±20 °C. Waters 2790/2795 Rapid Equilibration Page Path Select the path to be used for flushing solvent during rapid equilibration from the drop down list box. Waste, Off or Column 1 to Column 6 depending on the instrument configuration. Flow Enter the system equilibration flow rate. Range: 0.00 to 10.00 ml/min. 2-39