Download Waters CapLC Autosampler
Transcript
Chapter 2 Waters Systems Ratio Denominator Wavelength Enter the denominator wavelength (in nanometers) for the analog output channel. Ratio Denominator Wavelength must be within the wavelength range specified by the Start Wavelength and End Wavelength parameters on the 996 PDA page. Bandwidth Enter the spectral bandwidth of the analog output channel. The range is 1.2 to 24.0 nm. Filter Type Select Hamming or Single Pole from the dropdown list box. The Hamming filter is designed to create the same degree of peak-height degradation as the Single Pole filter for the same response time, but enhances filtering of high-frequency noise. Filter Response Enter the response time for the filter. The range is 0 to 5 seconds. Offset If required enter an offset to the analog output channel. The range is –0.2 to 2.0 AU. Threshold Enter a threshold above which the ratio (Wavelength / Ratio Denominator Wavelength) must be to be valid data. The range is –0.1 to 2.0 AU. Note: If no ratio is plotted one or both channels are below the current Threshold and a lower Threshold value should be entered. Waters 486 UV Detector This page is used to set parameters specific to the UV detector, to access it select View, Waters486 UV Detector or press the toolbar button. Figure 2.16 486 UV Detector Configuration page Wavelength (µ) Enter the wavelength to monitor. Polarity Select the polarity of the output signal from the drop down list box. 2-22