Download Amendment III - UNOPS-IPO-WBMSC-EQUIP-18
Transcript
• NEEDLE ELECTRODE FOR OPEN SURGERY Working length should be 40mm Unit should be CE/USFDA approved Safety and other standards: Interference immunity in accc.w.IEC 801 specs MDD Class IIb Mark of conformity CD 0297 in conformity w.93/42/EEC Patient leakage current Within the limit values specified in DIN IEC 601,part 1 Housing leakage current Within the limit values specified in DIN IEC 601 Part1 Schedule 16 - Liquid Nitrogen Cryosurgery Unit Sr. SPECIFICATION No 1. Should be a liquid nitrogen Cryosystem for use in Gynaecology/dermatology/ Ophthalmology/ Proctology etc. 2. Should be equipped with; i. Exchangeable jets ii. Plastic fingertip trigger iii. Safety and regulation valve iv. Apparatus cover with control valve, vacuum isolated outside tube and inside suction plastic pipe v. Dewar vessel vacuum isolated vi. Cryogun of capacity 500 ml vii. With a liquid nitrogen storage tank of capacity of 20 litres viii. Repumping apparatus with a controllable valve in the head of the apparatus to maintain in the storage tank the pressure of 20-30 kPa required for withdrawal of liquid nitrogen from the storage tank to the cryogun. 3. Should be supplied with the following sterilizable probes: 4. 5. For Dermatology; I. One set of Jets of Diameter: 0.4 mm, 0.6 mm & 0.8 mm and II. one set of long jets of diameter 0.9 mm/Length 100 mm, diameter 0.9 mm/Length 70 mm and diameter 0.9 mm/Length 40 mm III. one set of contact tips(Spray tips) of diameter 20 mm, 10 mm, 5 mm,3mm & 2 mm equipped with a tailpipe to provide freezing without direct contact of the tissue with the liquid nitrogen. Including 2 years warranty followed by 5 years of CAMC. Amendment – III: [UNOPS/IPO/WBMSC/EQUIP/18/2012] Page 47 of 60