GENERAL ABOUT THE SERVICE MODE
Chapter C - SERVICE MODE
Table 11: Other service modes
MODE
SERVICE MODE FUNCTION
DISPLAY/RANGE
page
n.109
Unit type setting to enable or disable the IMUX and the
door features
0 = Compact i
1 = Compact e
28 = Cart
31 = Independent Cart
44 = no Instrument Console
C-38
n 110
Show IC39 outputs
For factory use
C-38
n 111
Show inputs IN3 (IC39)
For factory use
C-39
n 112
Show inputs IN4
For factory use
C-39
n 113
Show side delivery arm emergency switches signals
---- = no side delivery arm
DxUy; x = downwards switch
y = upwards switch
0 = switch not activated
1 = switch activated
C-39
n 114
Show bowl position
0 = bowl not in rest position
1 = bowl in rest position
C-39
n 115
Show glass identification signal
0 = glass is not in holder
1 = glass is in holder
C-39
n.116
Select glass filling mode
0 = no identification
1 = glass identification activated
(default)
2 = automatic glass filling
C-39
n 117
Show active time of drive air pulse width modulation
For factory use
C-39
n 118
Show active time of spray water pulse width modulation
For factory use
C-39
n 119
Show active time of spray air pulse width modulation
For factory use
C-39
n 120
Show active instrument power consumption
nn W
C-40
n.121
Select the length of the automatic legrest
0 = 510 (default)
1 = 460
2 = 390
C-40
n.122
Factory test
-
C-40
n.123
Factory test
-
C-40
n.124
Set the type of Planmeca Compact i
0, 1
C-40
n 125
Show Water management system status
nnn
C-40
n 126
Show Water management system inputs and outputs
¦¦¦¦¦¦¦¦
C-41
n.127
Water management system water flow adjustment
C-41
n.129
Adjustment of the upper limit of pre-programmed chair
positions
C-41
n.130
Disable WCS
0 = WCS in use (default)
1 = Starting of WCS disabled
2 = WCS totally disabled
C-41
n.131
WCS: adjustment of the syring cleaning solution filling
time
20 - 180, (default = 40)
C-41
Technical Manual