Download User manual
Transcript
Alert Icon Type Description Possible Cause Action Automatic Off Status Displayed when The mask has been therapy ends due to removed. automatic off function. Put your mask back on, confirm good fit, and turn airflow on to resume therapy. Inlet blocked. Check filter. Notification Blocked airway Blockage at device inlet. Check device air inlet is not obstructed. Check air filter(s) are installed properly and are clean; replace if needed. Low Leak: Check Mask and Tube Notification Blocked airway Blockage at tube or mask. Check tube is not crushed or folded such that air flow is restricted. Check mask is attached properly and without any obstruction. Status Displayed when Check Mask Fit function is enabled from Patient Menu. n/a This alert can be cleared by pressing the Control Dial. Otherwise it will time out after 60 seconds. Loading Language and Rebooting Status Displayed when a new n/a language is selected from the menu. No action needed. Times out when complete. Busy Status Displayed when the device is temporarily inaccessible due to data communication. n/a No action needed. Status Displays last 3 nights of hourly use. n/a Press Control Dial to acknowledge and clear the screen. Otherwise message times out after 30 seconds. Check Mask Fit “Sleep Progress” User Manual n/a n/a 19