Calibration - Section 5 CA L I BRA T I ON MA N U A L K - F A C T OR se l ec t : 2 o f 3 MANUA L K - F AC TOR p l ease en t e r 1. 00 ! Manual K-Factor This submenu is used to manually calibrate the instrument by entering a new K-Factor. This factor is a multiplication or correction value used to calibrate the response of the instrument to a specific gas concentration. The factory default K-Factor for all gases is 1.00. Use the <set> key to enter. The instrument will display the actual KFactor. <Actual K-Factor> Use the cursor keys <left> or <right> to reach the desired position. Use the cursor keys <up> and <down> to select the figures you wish to enter. The KFactor must be in the range of 0.20 and 5.00. After a new value has been entered, actuate the <set> key to confirm or <esc> to exit. !Note: Individual settings for the Manual K-Factor will automatically revert to the default value 1.00, whenever the sensor is replaced by a new sensor. Only instruments using catalytic sensors: Catalytic sensors are calibrated for methane. When monitoring for other combustible gases, a K-factor must be entered. Satellite XT 4-20 mA Versions 5-4