Problem
Possible Cause
Solution
...continued
The spectrometer is overheated.
(You can check whether the
temperature of the electronics
board is within the acceptable
range by clicking the
electronics indicator on the
Diagnostics tab of the
Experiment Setup dialog box.)
Make sure the cooling vents on the
back of the spectrometer are not
blocked. Make sure the ambient
temperature is between 16° and
27°C (60° and 80°F).
A sampling accessory is not
installed properly or is not
aligned.
Make sure the accessory is installed
correctly and, if necessary, align it.
See the instructions that came with
the accessory.
Detector and beamsplitter are
not compatible.
142
Thermo Electron Corporation
Make sure that selected detector and
beamsplitter are compatible. Change
detector or beamsplitter if necessary.