Download Skyscan 1176 User Manual
Transcript
SkyScan 1176 In vivo Microtomograph 5.8.3. Actions menu: X-ray start / stop X-RAY START / STOP Use this command to switch on or off the x-ray source. After more than 100 hours interrupt in the source emission the aging procedure with gradual increasing of voltage and current will be started first. And following indication will start on the screen: Using this command once more x-ray emission can be stopped To open specimen chamber not necessary to push the X-RAY ON/OFF button to stop the emission: the X-ray source will stopped automatically after pressing SPECINEN CHAMBER button: After closing specimen chamber the x-ray source can be started by pushing X-RAY ON/OFF button or it will be started automatically after selection for any imaging mode: or In the closing program x-ray source will stopped automatically. It can take certain time. Don‟t switch off computer before X-ray source will be stopped and control program will be properly closed. Instruction Manual 79