Resolution:
UPLOAD CkSum ERR
CMS Upload experienced a Check Sum Error
At the completion of an upload of a CMS script the Check Sum signature
from the CMS system did not match the signature calculated by the pump
A second attempt to upload the script should be made. If the second
attempt fails, then a new script should be created for the desired
prescription. (See CMS User’s Manual).
131
Troubleshooting
LCD Display:
Alarm:
Cause: