Download 20006233_4 AWS Installation&Configuration Manual - Nova
Transcript
CONFIGURING THE PLANMED NUANCE ACQUIRE APPLICATION 7.13 Cropping images NOTE Always adjust the image cropping on the RCC image. Image Cropping is made by adjusting the cropping values in the windows registry NOTE Always use even numbers for the image cropping. Minimum value is 2 . NOTE Factory default image cropping values will be set with the software upgrade / install. Cropping values are in HEX format (decimal in parenthesis) 7.13.1 Image Cropping Modes The following new image cropping modes are introduced. 1. Screening (no cropping needed, minimum value 2 used) 2. Magnification 3. Coned Down Magnification (spot filter) 4. Coned down (contact – no cropping needed, minimum value 2 used) 5. Biopsy 7.13.2 Cropping for Contact View No additional cropping is necessary; the minimum value of 2 has been set. [HKEY_LOCAL_MACHINE\SOFTWARE\ISG\VR\5000\ImageCropping] "Left"= dword:00000002 (2) "Right"=dword:00000002 (2) "Top"= dword:00000002 (2) "Bottom"= dword:00000002 (2) 7.13.3 Cropping for LMAM Cropping for Coned Down view (Spot paddle, Bucky) No additional cropping is necessary; the minimum value of 2 has been set. [HKEY_LOCAL_MACHINE\SOFTWARE\ISG\VR\5000\ImageCroppingCD] [HKEY_LOCAL_MACHINE\SOFTWARE\ISG\VR\5000\ImageCroppingCD] "Enabled"=dword:00000001 (1) "Left"=dword:00000002 (2) "Right"=dword:00000002 "Top"=dword:00000002 "Bottom"=dword:00000002 Installation & Configuration Manual Planmed Nuance / Nuance Excel AWS B-53