ICS-1000 Ion Chromatography System B.20.2 Pressurizing the Eluent Reservoir (Optional) NOTE Do not pressurize the ICS-1000 until after the reservoir is plumbed. 1. Pull out the air regulator knob and turn it fully counterclockwise to ensure there is no pressure to the reservoir when the gas source is turned on. 2. Turn on the gas source. 3. Pull out the air regulator knob and turn it clockwise to pressurize the eluent reservoir. 4. Adjust the pressure to between 0.03 and 0.04 MPa (5 and 6 psi). Push the regulator knob back in. Do not pressurize the eluent reservoir above 0.07 MPa (10 psi). Ne mettez jamais les réservoirs d'éluants sous une pression supérieure à 0,07 MPa (10 lb/po²). Setzen Sie den Eluentbehälter auf keinen Fall einem Druck über 0,07 MPa aus. B-52 Doc. 031879-02 9/05