ICS-1000 Ion Chromatography System Outlet Check Valve Pressure Transducer Primary Pump Head Secondary Pump Head Priming Valve Inlet Check Valve Waste Valve Figure 2-12. ICS-1000 Pump Components To open the priming valve, turn the knob one-quarter to one-half turn counterclockwise. When the priming valve is open, liquid can flow into and out of the primary pump head via the port in the front of the valve. NOTE The priming valve must be open when the pump is being primed with a syringe or with isopropyl alcohol. For detailed priming instructions, see Section B.14. Secondary Pump Head The secondary pump head delivers eluent to the remainder of the chromatography system (the injection valve, column, and detector). The waste valve is located on the front of the secondary pump head (see Figure 2-12). To open the waste valve, turn the knob one-quarter to one-half turn counterclockwise. When the waste valve is in the open position, all output is directed to waste. NOTE The waste valve must be open when the pump is being primed using the Prime button. For detailed priming instructions, see Section B.14. 2-22 Doc. 031879-02 9/05