Download User Guide: SelexION Technology for 5500 and 6500
Transcript
Introduction to DMS Technology Instrument Tuning ® ® Before installing the SelexION or SelexION spectrometer is tuned and calibrated. + technology on the mass spectrometer, make sure that the mass DMS Off Operation Mode ® ® + When the SelexION or SelexION technology is installed on the mass spectrometer, users might need to ® ®+ acquire spectral data without using the SelexION or SelexION technology. Instead of removing the technology ® from the system, users can use the DMS Off operation mode to acquire data. The Analyst software provides a DMS Off operation mode, which allows the user to work in an environment similar to when the technology is not installed on the mass spectrometer. However, in the DMS Off mode, signal intensities will be significantly decreased as compared to the signal intensities of data acquired with the technology physically removed from the mass spectrometer. To work in the DMS Off operation mode, select the DMS Off check box available on the MS tab in the Acquisition method window or on the MS tab in the Tune Method Editor window. Figure 1-2 DMS Off Check Box in the Tune Method Editor In the DMS Off operation mode, the following changes occur: Analyst® 1.6.3 Software 8 of 77 User Guide RUO-IDV-05-0275-C