Pin 1, and the positive to both pins 2 and 3. Set the CAL data field to match the voltage
being supplied and click the SET data field.
ÿ Calibrate the SPL Meter. Connect an SPL calibrator to the Toolbox microphone, or, if
none is available, use a high quality SPL meter and the pink noise output from the
Toolbox. Either way, generate a known SPL field, as near to 94dBA SPL as possible. Set
the SPL LOW CAL field to match the SPL sound field level being generated and click the
SET data field. Do the same for the SPL HIGH CAL data field.
These steps will calibrate a Toolbox, but note that the factory constants will not be altered
by these calibrations. The only way to change the factory constants (which are a backup set
of constants that you can always return to) is by using a TerraSonde factory calibrator.