Download Model 450 Gaussmeter User`s Manual, 2002
Transcript
Lake Shore Model 450 Gaussmeter User’s Manual 5.3.3 Probe Specifications Terminology used in Figures 5-1 thru 5-4 are defined as follows: A +B HST HSE UHS Definition of Probe Terminology Distance from tip of probe to center line of active area Magnetic flux density vector (for positive gaussmeter reading) High Stability Probe High Sensitivity Probe Ultra-High Sensitivity Probe = = = = = Usable Full-Scale Ranges Versus Probe Type HST-1 HST-2 HSE-1 UHS-1 300 G 300 G 30 G 300 mG 3 kG 3 kG 300 G 3G 30 kG 30 kG 3 kG 30 G 300 kG — 30 kG — Type Usable on Ranges TRANSVERSE +B L 2.5" A W Model No. L MMT-6J02-VH 2" ±0.125 MMT-6J04-VH 4" ±0.125 MMT-6J08-VH 8" ±0.125 T T 0.36 ±0.030" dia. Cable length = 6.6 feet W A Active Area 0.061" 0.180 max. ±0.005" Stem Material Freq. Range Alum. DC HSE-1 MMT-6J18-VH 18" ±0.125 2" ±0.125 0.045" 0.150 MNT-4E04-VH 4" ±0.125 max. ±0.005" 0.150 ± MMT-6J02-VG 2" ±0.125 0.050" MMT-6J04-VG 4" ±0.125 0.061" 0.180 MMT-6J08-VG 8" ±0.125 max. ±0.005" MNT-4E02-VH 0.040" dia. approx. Rigid Glass Epoxy DC, 10 to 400 Hz Alum. DC MMT-6J18-VG 18" ±0.125 Rigid Glass Epoxy MNT-4E02-VG 2" ±0.125 0.045" 0.150 MNT-4E04-VG 4" ±0.125 max. ±0.005" MCT-3160-WN 60" ±0.05 0.25" dia. ±0.010 Type 0.210" ±0.050 Stainless Steel DC, 10 to 400 Hz Corrected Accuracy (% of rdg) Op. Temp. Range ±0.25% to 30 kG Temperature Coefficient (Max.) Zero Calibration ±0.09 Gauss per °C 0.015% per °C 0 °C to +75 °C HST-2 ±0.15% to 30 kG HST-1 ±2% to 100 kG ±0.13 Gauss per °C 1.5 K to 350 K 0.005% per °C 0.010% per °C C450-5-1.eps Figure 5-1 Definition of Lake Shore Transverse Probes 5-4 Accessories and Probes