Download Model 450 - Lake Shore Cryotronics, Inc.
Transcript
Lake Shore Model 450 Gaussmeter User’s Manual 3.16 FAST DATA MODE In normal operation, the instrument updates the display, computer interfaces, and the corrected analog output at a rate of 5 readings per second. Fast Data Mode increases the data rate when operating with either the IEEE-488 or Serial Interface. While the corrected analog output update rate does correspond to the Fast Data Mode, the front panel display will not operate in this mode. In Fast Data Mode, the front panel screen displays the message below. Fast D a ta Mode To place the instrument in Fast Data Mode, use the interface command: FAST 1. To leave fast data mode, use this command: FAST 0. To query the status of Fast Data Mode, use this command: FAST? The unit returns 0 if Fast Data Mode is Off, and 1 if On. NOTE: Fast Data Mode activation disables the following Model 450 functions: Relative, Max Hold, Alarms, and Autorange. Temperature compensation (if applicable) is based on the last temperature reading prior to FAST DATA MODE activation. The temperature is not updated during FAST DATA MODE. Use the normal interface command to query field measurement data. Without display overhead, the instrument can take 18 new readings each second. An efficiently written IEEE-488 program can return all 18 readings without slowing the instrument down. When using the IEEE-488 Interface, never try to read faster than 18 readings a second. The additional overhead associated with Serial Communication slows instrument Serial Interface communications to a maximum of 15 readings per second at 9600 Baud. When using the Serial Interface, never try to read faster than 15 readings a second. 3-16 Operation