A summary screen appears displaying the settings for the low and high BG targets.
HIGH TARGET BG
LOW TARGET BG
150 mg/dL
100 mg/dL
OK
Hourly Delivery Limit
Hourly Delivery Limit is a safety feature that limits the total amount of basal and bolus
insulin you can receive within an hour. When the combined basal and bolus delivery
reaches the limit within an hour, you will receive the DELIVERY LIMIT REACHED alarm.
Acknowledge the alarm to resume insulin delivery. Discuss this setting with your
healthcare professional. For information on the alarm, see the “Delivery Limit
Reached” alarm on page 119.
1
Main menu > SETUP MENU > DELIVERY LIMIT
2
Press
or
to set the hourly delivery limit value.
SET HOURLY
DELIVERY LIMIT
SAVE
3
20
U/H
EXIT
Press SAVE.
Button Guard
The Button Guard feature locks all access to the menus to prevent the accidental
pressing of a pump button.
Locking the Buttons
1
Main menu > SETUP MENU > BUTTON GUARD
2
Press
to turn ON the setting.
ON
SET BUTTON
GUARD
SAVE
EXIT
Setup
107