Download ProLine pH meter B210
Transcript
Description of the instrument 3. Description of the instrument 3.1 Display 1 2 3 4 Battery status Auto-off override during battery operation Data transfer to PC/Printer Electrode condition slope: 95 -105 % offset: ± (0 -15) mV Electrode is in good condition 5 Meter self-diagnosis Self-diagnosis indicator 6 7 8 9 10 11 12 13 4 slope: 90 - 94 % offset: ± (15 - 35) mV Electrode needs cleaning Indication to press key Error index pH/mV-reading Auto/Manual Temperature compensation Endpoint stability/autom endpoint Endpoint stability A Temperature Calibration point Buffer groups Electrode offset and slope slope: 85 - 89 % offset: ± (>35) mV Electrode is faulty Self-diagnosis passed Auto endpoint ProLine pH