FOURCLOT … Coagulation Analyser
USER MANUAL
FOUR CLOT
Coagulation Analyser
User Manual Version
With Incubator 1.005A
ROBONIK (INDIA) PVT. LTD.
Plot No.: A-374, TTC, MIDC Industrial Area,
Mahape, Navi Mumbai – 400 710. (INDIA)
Tel.: +91 (22) – 67829700, Fax: +91 (22) – 67829701
[email protected]
Website-www.robonik.in
Version No. : 1.005A
Effective Date: May 2012
Page 1 of 54
FOURCLOT … Coagulation Analyser
TABLE OF CONTENTS
USER MANUAL
1. GENERAL INFORMATION ................................................................................................... 4
1.1. Warranty Information:............................................................................................4
1.2. Technical Service: ..................................................................................................4
1.3. Disposal Instruction:..............................................................................................4
1.4. Contacts: ...................................................................................................................4
2. GENERAL SAFETY WARNINGS ........................................................................................ 5
2.1. Danger – warnings symbols: ...............................................................................5
2.2. Use of the Instrument: ...........................................................................................6
3. INTRODUCTION ..................................................................................................................... 7
3.1. DESCRIPTION ..........................................................................................................7
3.2. Special Features......................................................................................................7
3.3. Technical specification of FOUR CLOT:...........................................................8
4. PACKING, TRANSPORT AND STORAGE ....................................................................................10
4.1. GENERAL WARNINGS:.......................................................................................10
4.2. PACKING:................................................................................................................10
4.3. INSTRUMENT TRANSPORTATION :.................................................................10
4.4. STORAGE OF INSTRUMENT: ............................................................................10
5. INSTRUMENT DESCRIPTION ...........................................................................................11
5.1. Touch Sensitive screen .......................................................................................11
5.2. Perspective View...................................................................................................12
5.3. KEYPAD...................................................................................................................13
5.4. Printer.......................................................................................................................13
6. Installation and start-up instructions.............................................................................14
6.1. Placing the instrument ........................................................................................14
6.2. Power supply ..........................................................................................................14
6.3. Protective Grounding...........................................................................................14
6.4. Start up Instructions: ...........................................................................................14
7. PRECAUTIONS.....................................................................................................................15
8. MAIN MENU...........................................................................................................................16
8.1. TEST OPERATIONS: ............................................................................................16
A) Programming/ADDING a NEW TEST: ..................................................................16
B) Recalling a Saved TEST:..........................................................................................24
C) Running a Saved TEST (Operating the programmed Tests) ...........................
D) List Tests: .....................................................................................................................28
8.2. TECHNICAL DIAGNOSIS:....................................................................................29
8.2.1. Clear Memory .............................................................................................................................. 29
8.2.2. Temperature Setting.................................................................................................................... 30
8.2.3. Lamp Amplitude .......................................................................................................................... 30
8.2.4. Printer Self Test........................................................................................................................... 31
8.3. UTILITIES SCREEN:..............................................................................................31
8.3.1 To ENTER/CHANGE Date and Time.............................................................................................. 31
8.3.2. Password Utility: ......................................................................................................................... 32
8.3.3. To disable or enable the Printer: ................................................................................................. 33
8.3.4. To Enter Clinic Name: ................................................................................................................. 33
8.3.5. Lamp: ON .................................................................................................................................... 34
8.3.6. Stirrer: ON ................................................................................................................................... 34
Version No. : 1.005A
Effective Date: May 2012
Page 2 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
8.4. QUALITY CONTROL:............................................................................................35
8.5. View Stored Results/ Print Patient Report:....................................................36
8.6. Serial communication:.........................................................................................37
9. Programming Modes (PHOTOMETRIC).........................................................................38
9.1. ABSORBANCE.......................................................................................................38
9.2. FIXED TIME.............................................................................................................40
9.3. KINETIC....................................................................................................................44
9.4. END POINT..............................................................................................................48
10. Trouble Shooting: .............................................................................................................52
11. Decontamination ...............................................................................................................53
11.1. Decontamination Procedure ............................................................................53
11.2. Purpose of Decontamination ...........................................................................53
11.3. General Considerations ....................................................................................53
11.4. Procedure..............................................................................................................53
12. SAFETY CLEARANCE CERTIFICATE.........................................................................54
Version No. : 1.005A
Effective Date: May 2012
Page 3 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
1. GENERAL INFORMATION
1.1. Warranty Information:
Each Instrument is completely tested and guaranteed for twelve months from delivery. The
warranty applies to all the mechanical and electrical parts. It is valid only for proper installation,
use, and maintenance in compliance with the instructions given in this manual.
ROBONIK will at its discretion repair or replace parts, which may be found defective in the
warranty period. The warranty does not include any responsibility for direct or indirect personal
and material damages, caused by improper use or maintenance of the instrument.
Parts that are inherently subject to deterioration are excluded from the warranty. In case of
defects due to misuse of the instrument, any travel and man-hour expenses will be charged extra.
1.2. Technical Service:
ROBONIK is always accessible to the customers for any kind of information about installation,
use, maintenance, etc. When asking for service, please refer to this manual, and report the data
reported on the identification label (serial number).
Only qualified technicians are entitled to fix the instrument; the user, as described in this manual,
should carry out ordinary maintenance.
ROBONIK technical service or an authorized service center with specialized technicians, with
suitable instrumentation and original spare parts only are always available for extraordinary
maintenance (repair), under a yearly maintenance contract or on specific demand.
1.3. Disposal Instruction:
In case of removal or disposal of instrument, following instructions need to be followed
1) Do not dispose in municipal waste; follow local regulations for instrument disposal.
2) Plastic parts, Electronic PCBs and components can be recycled, so return back the
instrument to manufacturer.
1.4. Contacts:
MANUFACTURER:
ROBONIK (INDIA) PVT LTD
A-374, TTC, MIDC Industrial Area,
MAHAPE, NAVI MUMBAI –400710
INDIA
Tel: +91-22-67829700,
Fax: +91-22-67829701
Email: [email protected]
Website: www.robonik.in
Version No. : 1.005A
Effective Date: May 2012
Page 4 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
2. GENERAL SAFETY WARNINGS
2.1. Danger – warnings symbols:
The following symbols are used to inform the user of the safety rules.
This symbol indicates generic danger. It means that, serious damage can occur to
the operator if described precautions are not observed.
This symbol indicates HIGH ELECTRIC VOLTAGE. It is dangerous to touch any
part having this label. Only qualified operators can access these components, after
unplugging the instrument from the Supply.
This symbol indicates that the instrument involves the handling of samples, which
can be infected (urine or human serum).
In this condition, infection or
contamination might occur. Pay attention to the general safety warnings when in
presence of such biological substances. Use Protective clothes, gloves and
glasses.
This symbol in the user manual indicates that damages to the instrument or
erroneous results could occur if the given warnings are not followed.
This symbol indicates a portion, which is particularly important, and should be
studied carefully.
This symbol indicates a Protective Earth or Ground terminal.
General Symbols
Symbol for “Manufacturer”
Symbol for “ IN VITRO DIAGNOSTIC MEDICAL DEVICE”
Version No. : 1.005A
Effective Date: May 2012
Page 5 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
2.2. Use of the Instrument:
The instrument has to be used for the designed purposes under specified conditions,
following proper procedures and safety rules, by qualified personnel.
THIS MANUAL CONTAINS INSTRUCTIONS FOR OPERATION BY QUALIFIED PERSONNEL.
1) A qualified user has to make sure that environmental condition is suitable, the installation
is correct, the use and maintenance are proper, according to the general safety rules as
well as to the particular precautions described in the manual. (However, he is not entitled
to repair the instrument).
2) A qualified technician is entitled to maintain and fix the instrument, according to the
instructions given, using the original spare parts. Maintain room temperature and
Humidity as specified in the manual.
3) The instrument has to be used as described in this manual. If it is not use the protection
provided by the instrument may be impaired.
4) Alterations to the instrument are prohibited. The user is liable for any improper
modification to the instrument, and for the deriving consequences.
5) Should the instrument need extraordinary maintenance, contact MANUFACTURER
service or authorized service center. Specialized technicians who will be able to repair
the instrument using original spare parts will carry out the maintenance.
6) This IVD equipment complies with the emission and immunity requirements as per
IEC61326 series.
7) Warning: This equipment has been designed and tested to CISPER11 Class A. In a
domestic environment it may cause radio interference, in which case, you may need
to take measures to mitigate the interference."
8) An advisory that the electromagnetic environment should be evaluated prior to
operation of the device.
9) Warning: Do not use this device in close proximity to sources of strong electromagnetic
radiation (e.g. unshielded international RF sources), as these may interfere with the proper
operation.
Version No. : 1.005A
Effective Date: May 2012
Page 6 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
3. INTRODUCTION
3.1. DESCRIPTION
FOURCLOT is a programmable Coagulometer with a sophisticated on board software and
user-friendly touch screen. Its versatile and unique software supports most of the calculation
required for interpretation of results. It is a reliable, high precision machine. It is intended for in
vitro diagnostic use. It has a user-friendly program and capacity of storing the programmed
analytical methods and the QC results. It is intended for in vitro diagnostic use.
3.2. Special Features
Coagulation
•
•
•
•
•
•
•
•
•
•
•
Four independent programmable coagulation channels for faster throughput.
Display on line results of all four channels simultaneously on single screen.
Sample averaging up to 4 coagulation runs in PT, APTT, TT, Fibrinogen.
Designed to perform all routine tests such as PT, APTT, TT, Fibrinogen, Factor II, V, VII, IX,
X, XI, XII.
Uniform mixing of Sample & Reagent with magnetic stirrer.
Reporting of results in second, INR, Ratio and %,
Automatic sensing of sample/reagent dispensing to start the measurement
Incubation for 24 positions for samples and reagent.
Built in printer and option to connect external 80 column printer.
Low reagent/sample volume for testing
Multipoint Calibration curve with user entered standard dilution %, in FIB and Factor Assay
Photometer
•
•
•
Independent channel for Photometric assays
Multipoint calibration curve with max.7 calibrators for Photometric assays.
Chromgenic assays antithrombin III (AT-III), protein C (PC), protein S (PS), D-dimer.
Common Features
•
•
•
•
•
•
•
•
•
•
•
Effective temperature regulation system.
Robust system with built in stabilizer.
Latest technology with battery back up for 250 tests with QC and more than 2500 results,
independent memory for Coagulation and Photometric mode.
Alpha numeric Patients ID entry.
Combine report printing for coagulation and photometer test with unique patient ID.
Robust in built 20 column thermal printer with 384 stationary heads.
Unique circuitry for long lamp life.
Human machine interface : Touchpad, Keypad
Built in Incubator.
Quality Control with Levy Jennings and standard deviation graphs.
Access to test by touch of key.
Version No. : 1.005A
Effective Date: May 2012
Page 7 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
3.3. Technical specification of FOUR CLOT:
Coagulation:
Measuring System
Photometric
Optical measurement
Photodiode
Measurement Volume
150 µl
Sample & reagent mixing
With magnetic stirrer
Dry block incubator
Number of cuvette Holders
24 positions for sample and reagent cuvette
incubation.
Temperature
37 c
Timer
Programmable
0
Light source
LED/Laser
Analysis mode
PT, APTT, TT, Fibrinogen, Factor assay (Factor II,
V, VII, IX, X, XI, XII).
Results displayed in
Seconds, Ratio, INR, Concentration and %
Photometer:
Linear measurement range
Photometric Accuracy
Drift
Photometric Linearity
Optical measurement
Filters
Type of filter
Wave Length
Light source
Warm up Time
Analysis Mode
Version No. : 1.005A
Effective Date: May 2012
0.000 to 3.000 Absorbance Units (A).
± 2% or 0.007 whichever is higher, from 0 to 1.5
A
± 3% from 1.5 A to 3.0 A
<0.007 A/hr
2.2 A
Photodiode
Narrow band Interference
405, 630 nm (optional)
Tungsten Halogen
90 Sec
Photometric mode
Absorbance
Kinetic
End Point
Fixed Time
Page 8 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
Common:
Human Machine Interface
Display
TOUCH PANEL / KEYPAD
5” Graphic LCD, Negative Blue, STN
Printer
Built – in thermal printer, 20 column
Memory
32 KB Non Volatile RAM with Battery backup
Storage Capacity
Independent capacity of 250 open tests with 30
QC results (Normal and Abnormal controls) and
More than 2500 patient results with PID, for
Coagulation and Photometer.
External interface
RS232
Power
Wattage
50 Watts
Voltage
115 – 230 Volts ± 50 / 60 Hz.
Operating position
On horizontal flat, rigid and vibration free surface
Operating conditions
0
0
Temperature
From + 18 c to 35 c
Relative humidity
Up to 85 %
Storage conditions
0
0
Temperature
From -10 c to 60 c
Relative humidity
Up to 85 %
Enclosure
ABS Fire retardant
Size (cms)
34.5 X 30 X 14.7 (l x b x h)
Weight (Approx)
5.4 Kg (Approx.)
Version No. : 1.005A
Effective Date: May 2012
Page 9 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
4. PACKING, TRANSPORT AND STORAGE
4.1. GENERAL WARNINGS:
Instrument has to be decontaminated before packing for transportation.
4.2. PACKING:
Packaging is needed whenever the instrument is to be transported or shipped by
courier or other means.
To pack the instrument follows the instructions below:
1) Decontaminate the instrument as explained on decontamination chapter of this
manual.
2) Put the instrument into the original packaging box. Instrument has to be properly
protected by plastic protective material. Put copy of safety clearance certificate
(copy of Safety clearance certificate is attached at the end of this manual).
3) Mark the package with address, instrument identification and warning labels
4.3. INSTRUMENT TRANSPORTATION:
The transportation of the instrument in unpacked condition must be limited within the room
where it is used to avoid damage.
4.4. STORAGE OF INSTRUMENT:
Before storing the instrument for a long period, pack it carefully as described above and
store indoors.
Relative humidity has to be less than 85%, and temperature between -10ºc and -60ºC.
Version No. : 1.005A
Effective Date: May 2012
Page 10 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
5. INSTRUMENT DESCRIPTION
5.1. Touch Sensitive screen
FOURCLOT provides a Touch sensitive LCD panel and a KEYPAD for easy user interface. The
Menus are displayed; the text of the parameter forms the TOUCH ZONE.
Touch screen Layout
For Example: In PT Mode.
Name:
Mode: PT
ISI index: 1.000
Normal: N
MNPT: 0.000
%dilution:
Sample Avg: 4
QC: N
Incub. Time: 0
ESC
SAVE
Above is the generic representation of a Test Screen. The Highlighted zones are TOUCH
ZONES, which are active. On touching the “Touch Zone” of a parameter, a sub menu/menu is
displayed or the requested action is carried out, and rest of the “Touch Zone” is deactivated.
For Example:- To activate the selection.
•
To enter the ISI Index, touch any point in the shaded area “ISI Index” on the LCD screen. On
proper selection the analyser responds with blinking of the parameter text and also the TOUCH
ZONE and a submenu is displayed.
•
To enter Test Name: Touching the “Name ” touch zone provides a alphanumeric screen. Enter
the Test name by touching the Touch zone of that variable. The selected value blinks and is
displayed next to the parameter.
Selection Indicator:
Selected onscreen item is shown in a shaded background. When the screen first displays, the default
selection is shown. Pressing a selection either highlights that item or activates it.
Version No. : 1.005A
Effective Date: May 2012
Page 11 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
5.2. Perspective View
Front View:
Printer
Touch Screen
Incubator
Keypad
Photometric
Cuvette chamber
Coagulation Cuvette chamber
Rear View:
SMPS
Fan
Contrast POT
ON / OFF Switch
External Printer (Optional )
RS232 port
Version No. : 1.005A
Effective Date: May 2012
Cooling Fan
R-RED
Page 12 of 54
SMPS
FOURCLOT … Coagulation Analyser
USER MANUAL
5.3. KEYPAD
5.4. Printer
A) Internal Printer (Thermal Printer)
FOURCLOT comes with a built in 20 column Thermal Printer. User has to take proper
care to handle this delicate instrument.
TIPS FOR CAREFUL USAGE OF PRINTER
1) Do not pull the paper when loaded.
2) Lift the paper lever carefully and load the paper.
3) Keep the instrument clean and dust free.
FOURCLOT gives line feed automatically wherever it is necessary for clear reading like
a. While powering on
b. In between character lines
(Note: User may operate the instrument by disabling the printer from the utilities menu.)
ü How to insert the paper
Insert the thermal paper roll by placing the sensitive side facing down. The sensitive paper
side is recognizable by its smoother face.
Before inserting the paper
Version No. : 1.005A
Effective Date: May 2012
After inserting the paper
Page 13 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
B) External Printer (Optional)
Switch off the analyzer before connecting the printer.
Disconnect the small cable from the external printer port (Ref Diagram in 5.3) provided on
the rear of analyzer. User can then connect the external printer using the standard
communication cable to external printer port.
(Note : If an external printer is connected then the internal printer is disabled.)
6. Installation and start-up instructions
While installing and setting up the instrument, the safety warnings and general
precautions described in section 7 must be observed.
6.1. Placing the instrument
Place the instrument on a flat working surface or bench top capable of supporting the
weight of the instrument. A clearance of at least 3 inches around the instrument is
required to assure optimal ventilation. Room temperature should be between
18ºC and 35 ºC with a relative humidity below 85%. Protect it from direct sunshine and
maintain the instrument in a clean, relatively dust free environment to ensure maximum
performance.
6.2. Power supply
Once the instrument has been placed, plug it into a power source by using the locally
available approved plug-in cable. Power cord should be CE, CSA and UL marked
Voltage : 115 - 230 Volts ± 10%, 60- 50 Hz
6.3. Protective Grounding
Warning: Make sure that electrical power source is properly grounded.
6.4. Start up Instructions:
1) Switch on the instrument.
2) The instrument initializes all the parameters internally, and carries out a power on
self-test and then displays the following screen
FOURCLOT
ROBONIK (INDIA) PVT. LTD.
Plot No. A-374, TTC,
MIDC Industrial Area, Mahape
Navi Mumbai– 400 710
Tel: +91-22-67829700,
Fax: +91-22-67829701
Web:- www.robonik.india.in
Email:- [email protected]
Select Mode
1. Coagulation
2. Photometric
FOURCLOT TOUCH
CLINIC NAME
MAIN MENU
1. Test Operations
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Serial communication
User can select the option to run coagulation test or
photometric test.
Version No. : 1.005A
Effective Date: May 2012
Page 14 of 54
20/7/10
ESC
FOURCLOT … Coagulation Analyser
USER MANUAL
1) If a printer is enabled, Model Name ,Version Number ,Clinic name / Serial Number,
Current time and date will be printed .(Note: If a printer is not enabled, “Disable printer” message is
displayed. Touch YES to disable or NO to proceed. Refer for Printer settings)
2) Once initialization is over, a lamp located within the instrument will glow. This lamp
requires 90 seconds for stabilization.
3) After the instrument completes the above steps, a TEST MENU SCREEN / MAIN MENU
Screen appears.
4) The instrument is now in IDLE mode, and ready for use.
7. PRECAUTIONS
v Keep the place dry and clean.
v Check all the grounding wires properly.
v Repeat the readings, if Absorbance is more than 2.0 A.
v Use original Packing for transportation.
v Use clean Cuvettes. Check the blank absorbance of the cuvette at regular intervals.
v Check the temperature of cuvette block at regular intervals.
v Check the linearity of the instrument at regular intervals using standards.
v Do not take any reading when the lid is open.
v Incubate the cuvettes at set temperature for at least 10 minutes before using.
v Incubate the reagents at set temperature for at least 10 minutes before using.
v Do not use any sharp objects on the Touch Screen. Always use the STYLUS
provided to operate the touch panel.
Version No. : 1.005A
Effective Date: May 2012
Page 15 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
8. MAIN MENU
8.1. TEST OPERATIONS:
A) Programming/ADDING a NEW TEST:
When the unit is first turned on, Mode selection screen is displayed.
Select Mode
1 Coagulation
2 Photometric
To add and Run coagulation test select 1. Coagulation.
Main Menu Screen appears on the display.
ü
Main Menu screen
Main Menu screen consist of Menus as shown below
FOURCLOT
19/7/10
CLINIC NAME
MAIN MENU
1. Test Operations
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Serial communication
Test Operations screen
TEST OPERATIONS
1. Add New Test
2. List Test
3. Run Std / Cntr
4. Run sample
5. Run QC
ESC
Select Menu “Test Operations”
ESC
Select Menu “Add New Test”
Select Test Mode
1) PT
2) APTT
3) TT
4) FIB
5) FA
ESC
1
0
1
0
0
On Select Test Mode screen all the test modes along with the no. of test entered are displayed
Select test mode in which test is to be entered
Coagulation consists of five different modes. They are as follows1. PT mode
2. APTT mode
3. TT mode
4. FIB mode
5. FA mode
Let us see how to create a New Test in each mode.
Version No. : 1.005A
Effective Date: May 2012
Page 16 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
1. PT mode
Name:
Name:
Mode: PT
ISI index: 0.000
Normal: N
MNPT: 0.000
% DILUTION: N
Sample Avg:4
QC: N
Incub. Time: 0
Name:
Mode: PT
ISI index: 0.000
Normal: N
MNPT: 0.000
% DILUTION: N
Sample Avg:4
QC: N
Incub. Time: 0
Mode: PT
ESC
SAVE
ESC
SAVE
A
I
Q
Y
3
B C
J K
R S
Z ,
4 5
D E
L M
T U
- +
6 7
F
N
V
_
8
G
O
W
1
9
Enter Normal Values_
Name:
Mode: PT
ISI index: 0.000
Normal: N
MNPT: 0.000
% DILUTION: N
Sample Avg:4
QC: N
Incub. Time: 0
Normalize: N
Ctrl. Time: N
SAVE
Low: 0.000
High: 0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
Numeric Values:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
ENTR
CLR
Select “Normalize” if you want to run
the standards in PT mode.
Normalize: N
Ctrl. Time: N
Numeric Values:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
Name:
Mode: PT
ISI index: 0.000
Normal: N
MNPT: 0.000
% DILUTION: N
Numeric Value:
1 2 3 4 5 6
ENTR
7 8 9 0 .
CLR
Name:
Mode: PT
ISI index: 0.000
ESC
Normal: N
MNPT: 0.000
SAVE
% DILUTION: N
Sample Avg:4
QC: N
Normal
YES
NO
ESC
H
P
X
2
0
ENTR
CLR
Select “Ctrl. Time” if you directly want
to run the samples in PT mode.
Version No. : 1.005A
Effective Date: May 2012
Page 17 of 54
FOURCLOT … Coagulation Analyser
Name:
Mode: PT
ISI index: 0.000
Normal: N
MNPT: 0.000
% DILUTION: N
Sample Avg: 4
QC: N
Incub. Time: 0
ESC
SAVE
No. of Dil: 2
Concen
Time
0.000
0.000
0.000
0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
ESC
SAVE
Name:
Mode: PT
ISI index: 0.000
Normal: N
MNPT: 0.000
% DILUTION: N
Numeric Values:
1 2 3 4 5 6
ENTR
7 8 9 0 .
CLR
From this screen you can enter the
“Sample Avg” values. Max Samp Avg=4
Name:
Mode: PT
ISI index: 0.000
Normal: N
MNPT: 0.000
% DILUTION: N
Sample Avg: 4
QC: N
Incub. Time: 0
Name:
Mode: PT
ISI index: 0.000
Normal: N
MNPT: 0.000
% DILUTION: N
Sample Avg: 4
QC: N
% DILUTION
ESC
SAVE
YES
NO
No. of Dil: 2
Concen
Time
0.000
0.000
0.000
0.000
In this screen you have to enter the
concentration value & the time related to it.
Name:
Mode: PT
ISI index: 0.000
Normal: N
MNPT: 0.000
% DILUTION: N
Sample Avg: 4
QC: N
Incub. Time: 0
USER MANUAL
ESC
SAVE
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
Enter the Number of Dilutions.
Here, Min. number of dilution= 2 and
Max. number of dilution=6
QCNormal Values_
QCNH: 0.000
QCNL: 0.000
QCAbnormal Values_
QCABH: 0.000
QCABL: 0.000
ESC
QCNormal Values_
QCNH: 0.000
QCNL: 0.000
QCAbnormal Values_
QCABH: 0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
In this screen you can enter both
QCNormal (H /L) as well as QCAbnormal
(H /L) values.
Name:
Mode: PT
ISI index: 0.000
Normal: N
MNPT: 0.000
% DILUTION: N
Numeric Values:
1 2 3 4 5 6
ENTR
7 8 9 0 .
CLR
From this screen you can enter the
Incubation Time in seconds.
Version No. : 1.005A
Effective Date: May 2012
Page 18 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
2. APTT mode
Name:
Mode: APTT
Sample Avg: 1
ESC
Normal: N
QC: N
SAVE
Incub. Time: 0
Name:
Mode: APTT
Sample Avg: 1
ESC
Normal: N
QC: N
SAVE
Incub. Time: 0
Normal
YES
NO
Name:
A
I
Q
Y
3
B C
J K
R S
Z ,
4 5
Mode: APTT
D E
L M
T U
- +
6 7
F
N
V
_
8
G
O
W
1
9
H
P
X
2
0
ENTR
CLR
Name:
Mode: APTT
Sample Avg: 1
Normal: N
QC: N
Incub. Time: 0
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
Enter Normal Values_
Low: 0.000
High: 0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
Name:
Mode: APTT
Sample Avg: 1
ESC
Normal: N
QC: N
SAVE
Incub. Time: 0
QC Normal Values_
QCNH: 0.000
QC NL: 0.000
QC Abnormal Values_
QCABH: 0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
QC Normal Values_
QCNH: 0.000
QCNL: 0.000
QC Abnormal Values_
QCABH: 0.000
QCABL: 0.000
ESC
In this screen you can enter both QC
Normal (H /L) as well as QC Abnormal (H /L)
values.
ENTR
CLR
Name:
Mode: APTT
Sample Avg: 1
ESC
Normal: N
QC: N
SAVE
Incub. Time: 0
Name:
Mode: APTT
Sample Avg: 1
Normal: N
QC: N
Incub. Time: 0
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
This screen allows you to enter the
Incubation time in seconds.
Version No. : 1.005A
Effective Date: May 2012
Page 19 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
3. TT mode
Name:
Mode: TT
Sample Avg: 1
Normal: N
QC: N
Incub. Time: 0
Name:
Mode: TT
Sample Avg: 1
Normal: N
QC: N
Incub. Time: 0
Normal
Name:
Mode: TT
ESC
SAVE
ESC
SAVE
A
I
Q
Y
3
B C
J K
R S
Z ,
4 5
D E F
L M N
T U V
- + _
6 7 8
G
O
W
1
9
ENTR
CLR
Name:
Mode: TT
Sample Avg: 1
Normal: N
QC: N
Incub. Time: 0
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
YES
NO
H
P
X
2
0
ENTR
CLR
Enter Normal Values_
Low: 0.000
High: 0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
Name:
Mode: TT
Sample Avg: 1
Normal: N
QC: N
Incub. Time: 0
ESC
SAVE
QCNormal Values_
QCNH: 0.000
QCNL: 0.000
QCAbnormal Values_
QCABH: 0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
QCNormal Values_
QCNH: 0.000
QCNL: 0.000
QCAbnormal Values_
QCABH: 0.000
QCABL: 0.000
ESC
In this screen you can enter both
QCNormal (H /L) as well as QCAbnormal
(H /L) values.
ENTR
CLR
Name:
Mode: TT
Sample Avg: 1
Normal: N
QC: N
Incub. Time: 0
ESC
SAVE
Name:
Mode: TT
Sample Avg: 1
Normal: N
QC: N
Incub. Time: 0
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
This screen allows you to enter the
Incubation time in seconds.
Version No. : 1.005A
Effective Date: May 2012
Page 20 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
4. FIB mode
Name:
Mode: FIB
Unit: mg/dl
ESC
Normal: N
QC: N Incub. Time: 0
SAVE
Concen
Sample Avg: 1
Graph: PT TO PT
Y vs X: Time vs LOGCONC
Name:
Name:
Mode: FIB
Unit: mg/dl
Normal: N
QC: N
Incub. Time: 0
Concen
Sample Avg: 1
Graph: PT TO PT
Normal
Enter Normal Values_
A
I
Q
Y
3
B C
J K
R S
Z ,
4 5
Mode: FIB
D
L
T
6
E F
M N
U V
+ _
7 8
G
O
W
1
9
H
P
X
2
0
ENTR
CLR
ESC
SAVE
YES
NO
Low: 0.000
High: 0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
Enter the values of Normal in High / Low range
QCNormal Values_
QCNH: 0.000
QCNL: 0.000
QCAbnormal Values_
QCABH: 0.000
QCABL: 0.000
ESC
QCNormal Values_
QCNH: 0.000
QCNL: 0.000
QCAbnormal Values_
QCABH: 0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
Name:
Mode: FIB
Unit: mg/dl
ESC
Normal: N
QC: N Incub. Time: 0
SAVE
Concen
Sample Avg: 1
Graph: PT TO PT
Y vs X: Time vs LOGCONC
Name:
Mode: FIB
Unit: mg/dl
Normal: N
QC: N Incub. Time: 0
Concen
Numeric Values:
1 2 3 4 5 6
7 8 9 0 .
Name:
Mode: FIB
Unit: mg/dl
Normal: N
QC: N Incub. Time: 0
Concen
Sample Avg: 1
Graph: PT TO PT
Y vs X: Time vs LOGCONC
ENTR
CLR
ESC
SAVE
In this screen you can enter both QCNormal
(H /L) as well as QCAbnormal (H /L) values.
No. of Dil: 6
Concen
Time
0.000
0.000
0.000
0.000
0.000
0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
Select “Concen”. Enter the number of Dilution.
Here you can enter Max 6 Concen, Min 2 Concen
Graph Type: PT TO PT
Y vs X: Time vs LOGCONC
Type
LINEAR
PT TO PT
(Note: The UNIT in FIB mode is fixed)
Version No. : 1.005A
Effective Date: May 2012
ENTR
CLR
Page 21 of 54
ESC
Scale
LogTime vs CONC
Time vs LOGCONC
LogitTime vs Log CONC
Time vs CONC
LogTime vs Log CONC
FOURCLOT … Coagulation Analyser
Name:
Mode: FIB
Unit: mg/dl
ESC
Normal: N
QC: N Incub. Time: 0
SAVE
Concen
Sample Avg: 1
Graph: PT TO PT
Y vs X: Time vs LOGCONC
USER MANUAL
Name:
Mode: FIB
Unit: mg/dl
Normal: N
QC: N Incub. Time: 0
Concen
Numeric Values:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
This screen allows you to enter the
Incubation time in seconds.
5. FA mode
Name:
Mode: FA
PT:
ESC
APTT:
QC: N Incub. Time: 0
SAVE
Concen
Normal: N
Graph: LINEAR
Y vs X: LogTime vs LogCONC
Name:
Name:
Mode: FA
PT:
ESC
APTT:
QC: N
Incub. Time: 0
SAVE
Concen
Normal: N
Graph: LINEAR
Y vs X: LogTime vs LogCONC
FA_II
FA_V
FA_VII
FA_X
FA_VIII
FA_IX
FA_XI
FA_XII
Name:
Mode: FA
PT: FA_V
ESC
APTT:
QC: N Incub. Time: 0
SAVE
Concen
Normal: N
Graph: LINEAR
Y vs X: LogTime vs LogCONC
A
I
Q
Y
3
B C
J K
R S
Z ,
4 5
Mode: FIB
D E F
L M N
T U V
- + _
6 7 8
Name:
Mode: FA
PT:
ESC
APTT: FA_XI
QC: N Incub. Time: 0
SAVE
Concen
Normal: N
Graph: LINEAR
Y vs X: LogTime vs LogCONC
Version No. : 1.005A
Effective Date: May 2012
Page 22 of 54
G
O
W
1
9
H
P
X
2
0
ENTR
CLR
FOURCLOT … Coagulation Analyser
Name:
Mode: FA
PT:
ESC
APTT: FA_XI
QC: N
Incub. Time: 0
SAVE
Concen
Normal: N
Graph: LINEAR
Y vs X: LogTime vs LogCONC
No. of Dil: 6
Concen
Time
0.000
0.000
0.000
0.000
0.000
0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
USER MANUAL
QCNormal Values_
QCNH: 0.000
QCNL: 0.000
QCAbnormal Values_
QCABH: 0.000
QCABL: 0.000
ESC
QCNormal Values_
QCNH: 0.000
QCNL: 0.000
QCAbnormal Values_
QCABH: 0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
Select “Concen”. Enter the number of Dilution.
Here you can enter Max 6 Concen, Min 4 Concen
Name:
Mode: FA
PT: FA_V
APTT:
QC: N Incub. Time: 0
Concen
Normal: N
Graph: LINEAR
Normal
ENTR
CLR
In this screen you can enter both QC Normal
(H /L) as well as QC Abnormal (H /L) values.
Enter Normal Values_
ESC
SAVE
YES
NO
Low: 0.000
High: 0.000
A
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
Enter the values of Normal in High / Low range
Name:
Mode: FIB
Unit: mg/dl
ESC
Normal: N
QC: N
Incub. Time: 0
SAVE
Concen
Sample Avg: 1
Graph: PT TO PT
Y vs X: Time vs LOGCONC
Graph Type: PT TO PT
Y vs X: Time vs LOGCONC
Name:
Mode: FA
PT:
ESC
APTT:
QC: N Incub. Time: 0
SAVE
Concen
Normal: N
Graph: LINEAR
Y vs X: LogTime vs LogCONC
Name:
Mode: FA
PT:
APTT:
QC: N Incub. Time: 0
Concen
Type
LINEAR
PT TO PT
ESC
Scale
LogTime vs CONC
Time vs LOGCONC
LogitTime vs Log CONC
Time vs CONC
LogTime vs Log CON
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ESC
SAVE
ENTR
CLR
This screen allows you to enter the
Incubation time in seconds.
Version No. : 1.005A
Effective Date: May 2012
Page 23 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
B) Recalling a Saved TEST:
The Programmed / Saved Test can be recalled/selected by List Tests
When the unit is first turned on, the screen is displayed as shown below
Select Mode
1. Coagulation
2. Photometric
Select Coagulation, Main menu screen is displayed as shown below
Main Menu screen
FOURCLOT
CLINIC NAME
MAIN MENU
1. Test Operations
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Serial communication
19/7/10
TEST OPERATIONS
1. Add new test
2. List test
3. Run Std / ctrl
4. Run Sample
5. Run QC
ESC
Select Test
1. TT
2. PT
ESC
ESC
MENU
SEL
PRNT
NXT
PREV
DEL
2. PT
List test displays max. of 10 Tests/ Page .
A “List tests” screen would look like the screen shown above. Touch the test name to select the test.
Suppose we have selected the test in PT mode then it will display the screen of PT mode. In this screen
you can Edit the test parameters.
Name: PT
ISI index: 9.000
Normal: Y
MNPT: 0.000
%dilution:
Sample Avg: 4
QC: Y
Incub. Time: 10
Mode: PT
ESC
PRNT
EDIT
Similarly, you can Edit the test parameters in APTT, TT, FIB, FA mode by recalling the particular test.
C) Running a Saved TEST (Operating the programmed Tests)
1. Run Std / Ctrl
2. Run Sample
Version No. : 1.005A
Effective Date: May 2012
Page 24 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
1) Run Std / Ctrl :
In case of test with standard/ control without running the standard / control the test sample can not be run.
a. To run the control of PT test:When the unit is first turned on, the screen is displayed as shown below:
Select Mode
1. Coagulation
2. Photometric
Select Coagulation, Main menu screen is displayed as shown belowSelect Test Operations
Main Menu screen
To run the STD of FIB / FA or Normalization Controls of PT.
Step1: Select Menu TEST OPERATIONS.
FOURCLOT
19/7/10
CLINIC NAME
MAIN MENU
1. Test Operations
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Serial communication
ESC
Step 2: Select Run Std / Ctrl.
TEST OPERATIONS
1. Add new test
2. List test
3. Run Std / ctrl
4. Run Sample
5. Run QC
ESC
After selecting Run Std / Ctrl menu, screen is displayed to check the channel temperature.
Initially if the channel temperature is
30.43°C then this screen is displayed till
Channel temperature attend 37°C. To abort
Temperature, Select ESC
ESC
MENU
SEL
PRNT
NXT
PREV
DEL
Select test
1. PT
2. APTT
3. TT
4. FIB
5. FA
4. FIB
FIB Std Res
FIB
CH-1 Dilut:
Incub. Time: 10.25
ESC
Run
CH-2 Dilut:
Incub. Time: 10.15
ESC
Run
CH-3 Dilut:
Incub. Time: 10.09
ESC
Run
CH-4 Dilut:
Incub. Time: 10.20
ESC
ESC
Run
ESC
Set temperature :37.00
Channel temperature:30.43
ESC to abort temperature
After reaching the temperature or selecting
ESC, test mode screen is displayed as
shown aside. Select the test mode of which the
standard is to be run
After selecting the particular test, it
displays the RUN screen. On
selecting the RUN option, it will run
the Incubation time if and only if you
select Incub. Time in test parameter.
Again select RUN option in order to
run the standard for the required test.
A
Version No. : 1.005A
Effective Date: May 2012
Page 25 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
A
ESC
Run
FIB Std Res
CH-1 Dilut:
Add Sample:
CH-2 Dilut:
Load Cuv + Rgt
ESC
Run
CH-2 Dilut:
Add Sample:
ESC
Run
CH-3 Dilut:
Load Cuv + Rgt
ESC
Run
CH-3 Dilut:
Add Sample:
ESC
Run
CH-4 Dilut:
Load Cuv + Rgt
ESC
ESC
Run
CH-4 Dilut:
Add Sample:
ESC
ESC
Run
FIB Std Res
CH-1 Dilut:
Load Cuv + Rgt
FIB
FIB
ESC
Run
After loading Cuv +Rgt , it will run the standards in FIB mode. Thus you can get the standard values for the test
saved in FIB mode. After running the standards you have to run the samples in FIB mode.
2) Run Sample:
When the unit is first turned on, the screen is displayed as shown below
Select Mode
1. Coagulation
2. Photometric
Select Coagulation main menu screen is displayed as shown below
Select Test Operation
Main Menu screen
To run the Sample of added test.
Step 1: Select menu TEST OPERATIONS.
FOURCLOT
CLINIC NAME
MAIN MENU
1. Test Operations
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Serial communication
19/7/10
ESC
Step 2: Select Run Sample.
TEST OPERATIONS
1. Add new test
2. List test
3. Run Std / ctrl
4. Run Sample
5. Run QC
ESC
A
Version No. : 1.005A
Effective Date: May 2012
Page 26 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
After selecting Run sample menu, screen is displayed to check the channel temperature.
A
ESC
Initially if the channel temperature is
30.43°C then this screen is displayed till
Channel temperature attend 37°C. To abort
Temperature, Select ESC
Set temperature :37.00
Channel temperature:30.43
ESC to abort temperature
After attending the temperature or Selecting
ESC, Program Channels screen is displayed
as shown . Select the channel in which you
want to run the sample.
Program Channels
Test Name Mode
CH-1=
CH-2=
CH-3=
CH-4=
MENU
ESC
RUN
PRNT
On selecting the channel it will display “Select Test” screen through which we can select
the tests whose samples we want to run.
Suppose we have selected test in PT mode whose Sample AVG is 4, then it will assign
the samples in all the four channels.
Program Channels
Test Name Mode
CH-1= ` PT
P
CH-2= PT
P
CH-3= PT
P
CH-4= PT
P
Version No. : 1.005A
Effective Date: May 2012
MENU
ESC
RUN
PRNT
Page 27 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
D) List Tests:
List Test screen Displays max. of 10 Tests/ Page .
A “List tests” screen would look like the screen shown below
Select Test
1 TT
2 PT
3 PT_1
4 FIB
5 FA
ESC
MENU
SEL
PRNT
NXT
PREV
DEL
“ESC “: To escape.
User can select a saved test by first touching the required
TEST NAME on the screen and then touch the “SEL”
option to carry out further operation on the test.
“DEL”: Deletes the selected test.
If the number of tests is more than 10, “NEXT “and
“PREV” can be used for browsing through the list pages
“PRINT”: For printing the “LIST TESTS “
Main Menu screen
FOURCLOT
22/7/10
CLINIC NAME
MAIN MENU
1. Test Operations
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Serial communication
Select Test
1 TT
2 PT
3 PT_1
4 FIB
5 FA
ESC
ESC
MENU
SEL
PRNT
NXT
PREV
DEL
TEST OPERATIONS
1. Add New Test
2. List Tests
3. Run STD/ Control
4. Run Sample
5. Run QC
Select Test
1 TT
2 PT
3 PT_1
4 FIB
5 FA
1 TT
Touch “SEL” to operate the test.
(The selected test is displayed at the bottom
of the screen)
Version No. : 1.005A
Effective Date: May 2012
Touch “TT” to select the test.
Page 28 of 54
ESC
ESC
MENU
SEL
PRNT
NXT
PREV
DEL
FOURCLOT … Coagulation Analyser
USER MANUAL
8.2. TECHNICAL DIAGNOSIS:
Ref.
Ref.
Ref.
Ref.
8.2.1
8.2.2
8.2.3
8.2.4
TECHNICAL DIAGNOSIS
1 Clear Memory
2 Printer Self test
3 Set Temperature
4 Lamp Amplitude
ESC
8.2.1. Clear Memory
To clear memory. Tests and results are stored separately.
TECHNICAL DIAGNOSIS
1 Clear Memory
2 Printer Selftest
3 Set Temperature
4 Lamp Amplitude
ESC
TECHNICAL DIAGNOSIS
1 Clear Memory
2 Printer Selftest
3 Set Temperature
4 Lamp Amplitude
Clear Results ?
ESC
YES
NO
“Clearing memory …”
[Clears all the stored Results only]
TECHNICAL DIAGNOSIS
1 Clear Memory
2 Printer Selftest
3 Set Temperature
4 Lamp Amplitude
ESC
Clear Tests Records?
YES
NO
“Clearing memory ….”
[Clears all the stored tests]
Warning: If you select “YES” option then all the Test Results and Test Records will get deleted.
Version No. : 1.005A
Effective Date: May 2012
Page 29 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
8.2.2. Temperature Setting
To verify the set temperature.
FOURCLOT
22/7/10
CLINIC NAME
MAIN MENU
1. Test Operations
ESC
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Serial Communication
TECHNICAL DIAGNOSIS
1 Clear Memory
2 Printer Self Test
3 Set Temperature
4 Lamp Amplitude
ESC
Set Channel Temp.
ESC
Temperature=36.86
Temperature Setting
1 Set Channel Temp.
2 Set Incubation Temp.
ESC
Ref UP. 785
DOWN.
6. Serial communications
When the required temperature is selected, it
displays the cuvette temperature.
Set Incubation Temp.
ESC
Temperature=36.46
Ref UP. 785
DOWN.
8.2.3. Lamp Amplitude
To check lamp amplitude of the filters
FOURCLOT
22/7/10
CLINIC NAME
MAIN MENU
1. Test Operations
ESC
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Serial communications
TECHNICAL DIAGNOSIS
1 Clear Memory
2 Printer Selftest
3 Set Temperature
4 Lamp Amplitude
ESC
Displays the Lamp Amplitude of all the filters.
Note: The amplitude ideally should be in the range of 1500 to 2400 for all filters.
Version No. : 1.005A
Effective Date: May 2012
Page 30 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
8.2.4. Printer Self Test
Routine test to check printer. When selected,
FOURCLOT
22/7/10
CLINIC NAME
MAIN MENU
1. Test Operations
ESC
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Serial communication
TECHNICAL DIAGNOSIS
1 Clear Memory
2 Printer Selftest
3 Set Temperature
4 Lamp Amplitude
ESC
If printer is ON, and paper is properly
inserted, the analyzer prints the
following two lines to check printing.
TECHNICAL DIAGNOSIS
1 Clear Memory
8.3. UTILITIES SCREEN:
UTILITIES
8.3.1.
8.3.2.
8.3.3.
8.3.4.
8.3.5.
8.3.6.
Ver (Version No)
1 Date
2 Password
3 Printer On/ Off: ON
4 Enter Clinic Name
5 Lamp: OFF
6 Stirrer: OFF
ESC
8.3.1. To ENTER/CHANGE Date and Time
FOURCLOT
22/7/10
CLINIC NAME
MAIN MENU
1. Test Operations
ESC
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Serial communication
UTILITIES
Ver (Version No)
ESC
1. Date.
2. Password
3. Printer On/Off: ON
4. Enter Clinic Name
5. Lamp: OFF
6. Stirrer: OFF
Date : DD / MM / YY
ESC
27 / 12 / 06
Time: Hrs
: Min
: Sec
13 : 46 : 40
To enter /change the date and time, touch on
the shaded zones accordingly.
A numeric submenu screen will be displayed.
Enter the current date and time.
On completion, Touch ESC to escape.
Version No. : 1.005A
Effective Date: May 2012
Page 31 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
8.3.2. Password Utility:
FOURCLOT
CLINIC NAME
22/7/10
ESC
PASSWORD UTILITY
J K
ESC
1. Enable / Disable
E
F
G
H
L M
N
O
P
U
V
W
X
ENTR
CLR
Q R S T
1. Date
2. Password
3. Printer On/ Off: ON
4. Enter Clinic Name
5. Lamp: OFF
6. Stirrer: OFF
PASSWORD UTILITY
ADMIN 0000000
I
Ver (Version No)
ESC
MAIN MENU
1. Test Operations
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Serial communication
A B C D
UTILITIES
Y
Z
,
-
+
-
1
2
3
4
5
6
7
8
9
0
2. Set Password
3. LOGIN
4. LOGOUT
To enable password, Enter “PASSWORD”
PASSWORD UTILITY
Touch ENTR on completion
USER 0000000
(A message” password enabled “will be
displayed at the bottom of the screen)
A B C D
I
J K
E
F
G
H
L M
N
O
P
V
W
X
ENTR
CLR
Q R S T
U
Y
Z
,
-
+
-
1
2
3
4
5
6
7
8
9
0
To create user, enter required user name For
example “DAVID” Touch ENTR on
completion
PASSWORD
UTILITY
1. Enable / Disable
2. Set Password
3. LOGIN
4. LOGOUT
ESC
PASSWORD UTILITY
USER 0000000
A B C D
I
J K
E
F
G
H
L M
N
O
P
V
W
X
ENTR
CLR
Q R S T
U
Y
Z
,
-
+
-
1
2
3
4
5
6
7
8
9
0
To Login/ enable user, enter the name.
In this example its “DAVID”
Version No. : 1.005A
Effective Date: May 2012
Page 32 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
8.3.3. To disable or enable the Printer:
It is a toggle to disable or enable the printer.
(Toggle : Any instruction that works first one way and then the other; it turns something on the first time it is used
and then turns it off the next time)
Step 1: Go to “Utilities”
Step 2: Select “Printer On/Off” to enable.
FOURCLOT
22/7/10
CLINIC NAME
MAIN MENU
1. Test Operations
ESC
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Serial communication
UTILITIES
Ver (Version No)
ESC
1. Date
2. Password
3. Printer On/Off: OFF
4. Enter Clinic Name
5. Lamp: OFF
6. Stirrer: OFF
UTILITIES
Ver (Version No)
ESC
1. Date
2. Password
3. Printer On/Off: ON
4. Enter Clinic Name
5. Lamp: OFF
6. Stirrer: OFF
Step 3: Pinter is enabled. To disable
the printer select “Printer On/Off
8.3.4. To Enter Clinic Name:
Step 1: Go to “Utilities”
Step 2: Touch “Enter Clinic Name”
FOURCLOT
22/7/10
CLINIC NAME
MAIN MENU
1. Test Operations
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Serial communication
UTILITIES
Ver (Version No)
ESC
ESC
1. Date
2. Password
3. Printer On/Off: ON
4. Enter Clinic Name
5. Lamp: OFF
6. Stirrer: OFF
UTILITIES
(Version No)
CLINICUU UUU UUU
A B C D E F G H
I J K L M N O P
Q R S T U V W X
Y Z , - + - 1 2
3 4 5 6 7 8 9 0
ENTR
CLR
Enter Clinic Name (Max. Up to 19 characters)
Touch “ENTR”, upon completion.
Touch “CLR”, to clear a wrong entry.
Version No. : 1.005A
Effective Date: May 2012
Page 33 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
8.3.5. Lamp: ON
It is a toggle to switch the lamp ON and OFF only in case of Photometric mode. Whereas in Coagulation mode,
the LAMP is always in OFF condition.
(Toggle: any instruction that works first one way and then the other; it turns something on the first time it is used
and then turns it off the next time)
Step 1: Go to “Utilities”
FOURCLOT
22/7/10
CLINIC NAME
MAIN MENU
1. Test Operations
ESC
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Serial communication
Step 2: Touch “Lamp”
UTILITIES
(Version No)
ESC
1. Date
2. Password
3. Printer On/Off: ON
4. Enter Clinic Name
5. Lamp: OFF
6. Stirrer: OFF
UTILITIES
(Version No)
ESC
1. Date
2. Password
3. Printer On/Off: ON
4. Enter Clinic Name
5. Lamp: ON
6. Stirrer: OFF
Lamp is enabled. To disable the lamp touch
the Lamp option once again.
8.3.6. Stirrer: ON
It is a toggle to switch the lamp ON and OFF.
(Toggle: any instruction that works first one way and then the other; it turns something on the first time it is used
and then turns it off the next time)
Step 1: Go to “Utilities”
FOURCLOT
22/7/10
CLINIC NAME
MAIN MENU
1. Test Operations
ESC
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Serial communication
Step 2: Touch “Lamp”
UTILITIES
(Version No)
ESC
1. Date
2. Password
3. Printer On/Off: ON
4. Enter Clinic Name
5. Lamp: OFF
6. Stirrer: OFF
UTILITIES
(Version No)
ESC
Stirrer is enabled. To disable the Stirrer
touch the Stirrer option once again.
Version No. : 1.005A
Effective Date: May 2012
1. Date
2. Password
3. Printer On/Off: ON
4. Enter Clinic Name
5. Lamp: OFF
6. Stirrer: OFF
Page 34 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
8.4. QUALITY CONTROL:
Quality Control is a process that checks an instrument or testing site to make sure it is reporting
accurate results on patients. The reproducibility of a result from a testing site or instrument should fall within a
certain range. Control solutions of known values are often times used for checking quality control. An institution
may choose how often control solutions are run depending on the accrediting body and test complexity the
analyte falls under. Levy Jennings charts are often used identify problems with QC results.
A levy Jennings chart is a graph that quality control data is plotted on to give a visual indication whether a
laboratory test is working well.
SD Standard Deviation: A measure of variability representing an average distance of the data from the mean.
The greater the standard deviation, the greater the difference between the individual determinations and the less
the precision of the method.
For example:
FOURCLOT
22/7/10
CLINIC NAME
MAIN MENU
1. Test Operations
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Serial communication
Select test
1 PT
6 PT_N1
2 APTT
7 FIB_2
3 TT_1
8 APTT_2
4 FIB
9 TT_2
5 FA
10 PT_N2
QC NORMAL
ESC
ESC
MENU
SEL
PRNT
NXT
PREV
DEL
YES
NO
Version No. : 1.005A
Effective Date: May 2012
Select test
1 PT
2 APTT
3 TT_1
4 FIB
5 FA
6 PT_N1
7 FIB_2
8 APTT_2
9 TT_2
10 PT_N2
ESC
ESC
MENU
SEL
PRNT
NXT
PREV
DEL
1 PT
(Touch “NO” for QC Abnormal)
(Touch “YES” for QC Normal)
QC Normal
Test Name: PT
H: 5.000
L: 2.000
No.
Date
Result
1.
20/02/06
1555.2
2.
20/02/06
1678
3.
20/02/06
1457.3
4.
20/02/06
1567
5.
20/02/06
154
QC
1. Levy Jennings.
2. Standard Deviation.
QC ABNORMAL
Test Name: PT
H: 4.000
L: 1.000
No.
Date
Result
1.
20/02/06
1555.2
2.
20/02/06
1678.6
3.
20/02/06
1457.3
4.
20/02/06
1567.4
5.
20/02/06
1545.2
ESC
NXT
PREV
PLOT
PRNT
Page 35 of 54
ESC
NXT
PREV
PLOT
PRNT
FOURCLOT … Coagulation Analyser
USER MANUAL
8.5. View Stored Results/ Print Patient Report:
Displays the latest 10 stored tests results.
The system can store 2500 results in its database. The moment a sample is RUN, the result is stored in the
system database. The variables stored in the database are as follows
MODEL NAME
CURRENT DATE
Date of Testing
Name
M
Result
PT
P
2.454
TT
T
66.59
FIB
F
272.6
Units
sec
sec
ESC NEXT PREV DEL
ID
101
27
PAT ID
A B C
I J K
Q R S
Y Z ,
3 4 5
D
L
T
6
E
M
U
+
7
F
N
V
8
G
O
W
1
9
H
P
X
2
0
ENTR
CLR
PRINT PID
Enter Patient’s identification number.
Test Name:
H:
No. Date
1. 20/02/06
2. 20/02/06
3. 20/02/06
4. 20/02/06
5. 20/02/06
6. 20/02/06
CREATNINE
L:
Result
1555.2
1678
1457.3
1567
1545.2
1595.2
Enter Patient’s Name:
ESC
NEXT
PREV
PLOT
PRINT
PAT ID
A B C
I J K
Q R S
Y Z ,
3 4 5
D
L
T
6
E
M
U
+
7
F
N
V
8
G
O
W
1
9
H
P
X
2
0
ENTR
CLR
“PRINT”: Prints the Patient report with the patient name and identification number
(Note: The patient’s name and identification number (PID) can be entered up to six places)
“DEL”: To delete the selected result.
M: Mode (2ND Column): A-Absorbance, F-Fixed time, K-Kinetic, E- Endpoint,
D-Differential, R-Ratio
Version No. : 1.005A
Effective Date: May 2012
Page 36 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
8.6. Serial communication:
The instrument is equipped with an RS232 serial port for PC configuration (user-computer interface.)
A cable is available to link the instrument to PC
Communication will only start when both ends detect the presence of an active terminal or device.
RS232 port settings in a windows Operating system
PORT SETTINGS
Bits per second
Data Bits
Parity None
Stop Bits
Flow control
: 9600
: 8
: None
: 1
: None
FOURCLOT
22/7/10
CLINIC NAME
MAIN MENU
1. Test Operations
ESC
2. Technical Diagnosis
3. Utilities
4. Quality control
5. View Stored Results
6. Serial communication
Version No. : 1.005A
Effective Date: May 2012
SERIAL COMM.
ESC
1.
Result
Page 37 of 54
Send Results
YES
NO
FOURCLOT … Coagulation Analyser
USER MANUAL
9. Programming Modes (PHOTOMETRIC)
9.1. ABSORBANCE
The instrument measures the Blank transmittance and the Sample transmittance, subtracts them and calculates
the sample absorbance (monochromatic mode).It is possible to repeat the measure with a different wavelength
(Filter 2 / secondary filter) and consider the difference between the two measures. (Bichromatic mode)
PROGRAMMING / ADDING a NEW TEST
Name:
Pri. : 405
Mod: ABS
Sec: 0
ESC
ADD
SAVE
ABSORBANCE Mode screen before programming would look like the above screen
Refer chapter11 for entering the test parameters .On completion, touch “SAVE” to save the programmed test
Name:
Pri. : 405
Mod: ABS
Sec: 0
Filter 1 :
Filter 2 :
Name:
ESC
ADD
SAVE
A
I
Q
Y
3
B C
J K
R S
Z ,
4 5
ESC
Select Filter Val :
0
630
405
Version No. : 1.005A
Effective Date: May 2012
Page 38 of 54
Mode: FIB
D E F
L M N
T U V
- + _
6 7 8
G
O
W
1
9
H
P
X
2
0
ENTR
CLR
FOURCLOT … Coagulation Analyser
USER MANUAL
Ø Recalling /Running (Operating ) a pre-programmed test
For example:
Name: Test Name, Mod: ABS
Pri. : 405
, Sec.:0
ESC
PRNT
EDIT
RUN
QCR
No_Unt
To change values press EDIT
Test Name: ABS
ESC
PID
RUN
RUN QC
Set temperature: 37.00
Channel temperature: 30.43
Test Name: ABS
ESC
PID
RUN
RunQC
Close the Lid & Press Run
ESC to abort temperature
Test Name: ABS
Test Name: ABS
ESC
PID
RUN
RunQC
ESC
PID
RUN
RunQC
.
Referencing.........
Lag Time: 5
Read Sample
Test Name: ABS
Sr. No. : 1
Result: 0.705
ESC
PID
RUN
RunQC
Sample Readings
Result: 0.705
Remove Sample Cuvette
After inserting the Cuvette in the Cuvette
block in the photometric mode, the sample
get run by selecting “RUN” option.
Version No. : 1.005A
Effective Date: May 2012
Page 39 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
9.2. FIXED TIME
Change in absorbance of sample is taken at programmed time interval and concentration is calculated either from
the factor fed by the user or using standard.
PROGRAMMING / ADDING a NEW TEST
Name:
Mod: FIX
Pri. : 405
, Sec.: 0
KF: 1.000
Unit: No – Unt
QC: N
Lag: 0
, Read: 0
Blk: N,
Normal: N
Std: N,
Factor: 0.000
Limit Set:
ESC
PRNT
EDIT
RUN
QCR
FIXED TIME Mode screen before programming would look like the above screen
Refer chapter11 for entering the test parameters .On completion, touch “SAVE” to save the programmed test in
memory.
Name:
Mod: FIX
Pri. : 405
, Sec.: 0
KF: 1.000
Unit: No – Unt
QC: N
Lag: 0
, Read: 0
Blk: N,
Normal: N
Std: N,
Factor: 0.000
Limit Set:
Filter 1 :
Filter 2 :
Select Filter Val :
0
630
405
Version No. : 1.005A
Effective Date: May 2012
Name:
ESC
PRNT
EDIT
RUN
QCR
ESC
A
I
Q
Y
3
B C
J K
R S
Z ,
4 5
Mode: FIX
D E
L M
T U
- +
6 7
F
N
V
_
8
G
O
W
1
9
Name:
Mode: FIX
Pri: 630
,Sec: 0
KF: 1.000
Unit: No_Unit
QC: N
Lag: 0
,Read: 0
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
Page 40 of 54
H
P
X
2
0
ENTR
CLR
ENTR
CLR
FOURCLOT … Coagulation Analyser
Name:
Mode: FIX
Pri: 630
,Sec: 0
KF: 1.000
Unit: No_Unit
QC: N
Lag: 0
,Read: 0
Blk: N
Normal: N
Std: N
Factor: 1.000
Limit Set:
ESC
ADD
SAVE
Set Times_
Lag Time: 0
Read Time: 0
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ESC
CLR
USER MANUAL
UNITS_ _
No_Unt
U/ L
mU/ ml
MIL
U/ml
mEq/ L
mmo/ L
Umo/ L
nmo/ L
%
g/ l
mg/ dl
Ug/ dl
g/ dl
Ug/ ml
Name:
Mod: FIX
Pri. : 405
, Sec.: 0
KF: 1.000
Unit: No – Unt
QC: N
Lag: 0
, Read: 0
Blk : N,
Normal: N
Std: N,
Factor: 0.000
Blank:
YES
NO
Touch “YES”, if you want to select Blank.
Name:
Mod: FIX
Pri. : 405
, Sec.: 0
KF: 1.000
Unit: No – Unt
QC: N
Lag: 0
, Read: 0
Blk: N,
Normal: N
Std: N,
Factor: 0.000
Standard
YES
NO
Name:
Mod: FIX
Pri. : 405
, Sec.: 0
KF: 1.000
Unit: No – Unt
QC: N
Lag: 0
, Read: 0
Blk: N,
Normal: N
Std: Y,
Concen:
Touch “YES”, if you want to Run the Standards in
Fixed Time mode. Touch “Concen” in order to
enter the number of concentrations.
Std Num: 0
Con1: 25
Con2: 50
Con3:
Con4:
Con5:
Con6:
Con7:
Del1:
Del2:
Del3:
Del4:
Del5:
Del6:
Del7:
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
(Touch “NO” if you want to select “Factor”)
Name:
Mod: FIX
Pri. : 405
, Sec.: 0
KF: 1.000
Unit: No – Unt
QC: N
Lag: 0
, Read: 0
Blk: N,
Normal: N
Std: N,
Factor: 0.000
Limit Set:
Here you have to enter the number of standards,
after which you can enter it’s concentration.
Name:
Mod: FIX
Pri. : 405
, Sec. : 0
KF: 1.000
Unit: No – Unt
QC: N
Lag: 0
, Read: 0
Numeric Value:
1 2 3 4 5 6
ENTR
7 8 9 0 .
CLR
Version No. : 1.005A
Effective Date: May 2012
Page 41 of 54
FOURCLOT … Coagulation Analyser
Name:
Mode: FIX
Pri: 630
,Sec: 0
KF: 1.000
Unit: No_Unit
QC: N
Lag: 0
,Read: 0
Blk: N
Normal: N
Std: N
Factor: 1.000
Limit Set:
USER MANUAL
Init. OD: 0.000
Max Delta: 0.000
Rgnt. Linearity: 0.000
ESC
ADD
SAVE
Init. OD: 0.000
Max Delta: 0.000
Rgnt. Linearity: 0.000
ESC
Init. OD: 1.000
Max Delta: 0.000
Rgnt. Linearity: 0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
ENTR
CLR
Min Init. OD= 0, Max Init. OD= 4
Init. OD: 1.000
Max Delta: 1.500
Rgnt. Linearity: 0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
After entering all the parameters in FIXTIME mode, touch the SAVE option present on the screen.
Version No. : 1.005A
Effective Date: May 2012
Page 42 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
Ø Recalling /Running (Operating) a pre-programmed test.
For example: (Running screen of FIXED TIME mode when Factor=1)
Name:
Mode: FIX
Pri: 630
,Sec: 0
KF: 1.000
Unit: No_Unit
QC: N
Lag: 0
,Read: 0
Blk: N
Normal: N
Std: N
Factor: 1.000
Limit Set:
To change values press EDIT
Test Name: FIX
ESC
PRNT
EDIT
RUN
QCR
ESC
PID
RUN
RUN QC
Test Name: FIX
ESC
PID
RUN
RUNQC
Set temperature: 37.00
Channel temperature: 30.43
Close the Lid & Press Run
ESC to abort temperature
Test Name: FIX
Test Name: FIX
ESC
PID
RUN
RunQC
Referencing.........
Lag Time: 5
ESC
PID
RUN
RunQC
Read Sample
Insert sample Cuvette & select Run
Test Name: FIX
Factor: 1.000
Sr. No. : 1
Init. OD: 0.710
0.710
Delta: 0.001
Result: 0.001 Umo/ L
Sample Readings
Result: 0.001 Umo/ L
Remove Sample Cuvette
.
Version No. : 1.005A
Effective Date: May 2012
Page 43 of 54
ESC
PID
RUN
RunQC
FOURCLOT … Coagulation Analyser
USER MANUAL
9.3. KINETIC
Multiple readings are taken at set temperature, at regular intervals and change in absorbance per minute is
calculated. Concentration is calculated from the factor fed by the user or by using standard.
PROGRAMMING / ADDING a NEW TEST
Name:
Pri: 630
KF: 1.000
Unit: mg/ dl
Lag: 0
Blk: N
Std: N
Limit Set:
Mode: KIN
,Sec: 0
QC: N
,Read: 0
Normal: N
Factor: 1.000
ESC
ADD
SAVE
KINETIC Mode screen before programming would look like the above screen
Name:
Pri: 630
KF: 1.000
Unit: mg/ dl
Lag: 0
Blk: N
Std: N
Limit Set:
Mode: KIN
,Sec: 0
QC: N
,Read: 0
Normal: N
Factor: 1.000
Filter 1 :
Filter 2 :
Name:
ESC
ADD
SAVE
ESC
Select Filter Val :
0
630
405
A
I
Q
Y
3
B C
J K
R S
Z ,
4 5
Mode: KIN
D E F
L M N
T U V
- + _
6 7 8
G H
O P
W X
1 2
9 0
Name:
Mode: FIX
Pri: 630
,Sec: 0
KF: 1.000
Unit: No_Unit
QC: N
Lag: 0
,Read: 0
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
ENTR
CLR
Select the Primary & Secondary filters.
Name:
Pri: 630
KF: 1.000
Unit: mg/ dl
Lag: 0
Blk: N
Std: N
Limit Set:
Mode: KIN
,Sec: 0
QC: N
,Read: 0
Normal: N
Factor: 1.000
ESC
ADD
SAVE
Set Times_
Lag Time: 0
Read Time: 0
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ESC
CLR
UNITS_ _
No_Unt
U/ L
mU/ ml
MIL
U/ml
mEq/ L
mmo/ L
Umo/ L
nmo/ L
%
g/ l
mg/ dl
Ug/ dl
g/ dl
Ug/ ml
Name:
Mod: KIN
Pri. : 405
, Sec.: 0
KF: 1.000
Unit: No – Unt
QC: N
Lag: 0
, Read: 0
Blk : N,
Normal: N
Std: N,
Factor: 0.000
Blank:
YES
NO
Touch “YES”, if you want to select Blank.
Version No. : 1.005A
Effective Date: May 2012
Page 44 of 54
FOURCLOT … Coagulation Analyser
Name:
Mod: KIN
Pri. : 405
, Sec. : 0
KF: 1.000
Unit: No – Unt
QC: N
Lag: 0
, Read: 0
Blk: N,
Normal: N
Std: N,
Factor: 0.000
Standard
YES
NO
USER MANUAL
Name:
Mod: KIN
Pri. : 405
, Sec.: 0
KF: 1.000
Unit: No – Unt
QC: N
Lag: 0
, Read: 0
Blk: N,
Normal: N
Std: Y,
Concen:
Touch “YES”, if you want to Run the Standards in
Fixed Time mode. Touch “Concen” in order to
enter the number of concentrations.
Std Num: 0
Con1: 25
Con2: 50
Con3:
Con4:
Con5:
Con6:
Con7:
D/ M1:
D/ M2:
D/ M3:
D/ M4:
D/ M5:
D/ M6:
D/ M7:
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
(Touch “NO” if you want to select “Factor”)
Name:
Mod: KIN
Pri. : 405
, Sec.: 0
KF: 1.000
Unit: No – Unt
QC: N
Lag: 0
, Read: 0
Blk : N,
Normal: N
Std: Y,
Factor: 0.000
Limit Set:
ENTR
CLR
Here you have to enter the number of standards,
after which you can enter it’s concentration.
Name:
Mod: KIN
Pri. : 405
, Sec.: 0
KF: 1.000
Unit: No – Unt
QC: N
Lag: 0
, Read: 0
Numeric Value:
1 2 3 4 5 6
ENTR
7 8 9 0 .
CLR
Name:
Pri: 630
KF: 1.000
Unit: mg/ dl
Lag: 0
Blk: N
Std: N
Limit Set:
Mode: KIN
,Sec: 0
QC: N
,Read: 0
Normal: N
Factor: 1.000
ESC
ADD
SAVE
Name:
Pri: 630
KF: 1.000
Unit: mg/ dl
Lag: 0
Blk: N
Std: N
Normal
Mode: KIN
,Sec: 0
ESC
ADD
SAVE
QC: N
,Read: 0
Normal: N
Factor: 1.000
YES
NO
Enter Normal Values_
Low: 0.000
High: 0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
Version No. : 1.005A
Effective Date: May 2012
Page 45 of 54
ENTR
CLR
FOURCLOT … Coagulation Analyser
Name:
Mode: KIN
Pri: 630
,Sec: 0
KF: 1.000
Unit: No_Unit
QC: N
Lag: 0
,Read: 0
Blk: N
Normal: N
Std: N
Factor: 1.000
Limit Set:
USER MANUAL
Init. OD: 0.000
Max Delta: 0.000
Rgnt. Linearity: 0.000
ESC
ADD
SAVE
Init. OD: 0.000
Max Delta: 0.000
Rgnt. Linearity: 0.000
ESC
Init. OD: 1.000
Max Delta: 0.000
Rgnt. Linearity: 0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
ENTR
CLR
Init. OD: 1.000
Max Delta: 1.500
Rgnt. Linearity: 0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
After entering all the parameters in KINETIC mode, touch the SAVE option present on the screen.
Version No. : 1.005A
Effective Date: May 2012
Page 46 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
Ø Recalling /Running (Operating ) a pre-programmed test
For example:
Name:
Mode: KIN
Pri: 630
,Sec: 0
KF: 1.000
Unit: mg/ dl
QC: N
Lag: 0
,Read: 0
Blk: N
Normal: N
Std: N
Factor: 1.000
Limit Set:
To change values press EDIT
Test Name: KIN
ESC
PRNT
EDIT
RUN
QCR
ESC
PID
RUN
RUN QC
Set temperature: 37.00
Channel temperature: 30.43
ESC to abort temperature
Test Name: KIN
Test Name: KIN
ESC
PID
RUN
RunQC
ESC
PID
RUN
RUNQC
Referencing.........
Lag Time: 5
Close the Lid & Press Run
Test Name: KIN
Factor: 1.000
Read Sample
Version No. : 1.005A
Effective Date: May 2012
Page 47 of 54
ESC
PID
RUN
RunQC
FOURCLOT … Coagulation Analyser
USER MANUAL
9.4. END POINT
The instrument reads absorbance of the sample and calculates concentration using fed Factor or calculates the
factor from concentration of the standard.
PROGRAMMING / ADDING a NEW TEST
Name:
Mode: END
Pri: 630
,Sec: 0
KF: 1.000
Unit: mg/ dl
QC: N
Lag: 0
,Read: 0
Blk: N
Normal: N
Std: N
Factor: 1.000
Limit Set:
ESC
ADD
SAVE
END POINT Mode screen before programming would look like the above screen
Name:
Mode: END
Pri: 630
,Sec: 0
KF: 1.000
Unit: mg/ dl
QC: N
Lag: 0
,Read: 0
Blk: N
Normal: N
Std: N
Factor: 1.000
Limit Set:
Filter 1 :
Filter 2 :
Name:
ESC
ADD
SAVE
ESC
Select Filter Val :
0
630
405
A
I
Q
Y
3
B C
J K
R S
Z ,
4 5
Mode: END
D E F
L M N
T U V
- + _
6 7 8
G
O
W
1
9
H
P
X
2
0
ENTR
CLR
Name:
Mode: END
Pri: 630
,Sec: 0
KF: 1.000
Unit: No_Unit
QC: N
Lag: 0
,Read: 0
Numeric Value:
1 2 3 4 5 6
ENTR
7 8 9 0 .
CLR
Select the Primary & Secondary filters.
Name:
Pri: 630
KF: 1.000
Unit: mg/ dl
Lag: 0
Blk: N
Std: N
Limit Set:
Mode: END
,Sec: 0
QC: N
,Read: 0
Normal: N
Factor: 1.000
Name:
Mod: END
Pri. : 405
, Sec.: 0
KF: 1.000
Unit: No – Unt
QC: N
Lag: 0
, Read: 0
Blk : N,
Normal: N
Std: N,
Factor: 0.000
Blank:
ESC
ADD
SAVE
UNITS_ _
No_Unt
U/ L
mU/ ml
MIL
U/ml
mEq/ L
mmo/ L
Umo/ L
nmo/ L
%
g/ l
mg/ dl
Ug/ dl
g/ dl
Ug/ ml
Set Times_
Lag Time: 0
Read Time: 0
YES
NO
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
Touch “YES”, if you want to select Blank.
Version No. : 1.005A
Effective Date: May 2012
Page 48 of 54
ESC
CLR
FOURCLOT … Coagulation Analyser
Name:
Pri: 630
KF: 1.000
Unit: mg/ dl
Lag: 0
Blk: N
Std: N
Limit Set:
Mode: END
,Sec: 0
QC: N
,Read: 0
Normal: N
Factor: 1.000
Name:
Mod: END
Pri. : 405
, Sec.: 0
KF: 1.000
Unit: No – Unt
QC: N
Lag: 0
, Read: 0
Blk: N,
Normal: N
Std: N,
Factor: 0.000
Standard
USER MANUAL
QCNormal Values_
QCNH: 0.000
QCNL: 0.000
QCAbnormal Values_
QCABH: 0.000
ESC
ADD
SAVE
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
Name:
Mod: END
Pri. : 405
, Sec.: 0
KF: 1.000
Unit: No – Unt
QC: N
Lag: 0
, Read: 0
Blk: N,
Normal: N
Std: Y,
Factor: 0.000
Limit Set:
YES
NO
Touch “NO” if you want to select “Factor”
Name:
Mod: END
Pri. : 405
, Sec.: 0
KF: 1.000
Unit: No – Unt
QC: N
Lag: 0
, Read: 0
Blk: N,
Normal: N
Std: Y,
Concen:
Name:
Mod: END
Pri. : 340
, Sec.: 0
KF: 1.000
Unit: No – Unt
QC: N
Lag: 0
, Read: 0
Numeric Value:
1 2 3 4 5 6
ENTR
7 8 9 0 .
CLR
Touch “YES”, if you want to Run the Standards in
Fixed Time mode. Touch “Concen” in order to
enter the number of concentrations.
Std Num: 0
Con1: 25
Con2: 50
Con3:
Con4:
Con5:
Con6:
Con7:
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
Ab1:
Ab2:
Ab3:
Ab4:
Ab5:
Ab6:
Ab7:
ENTR
CLR
Here you have to enter the number of standards,
after which you can enter it’s concentration.
Version No. : 1.005A
Effective Date: May 2012
Page 49 of 54
FOURCLOT … Coagulation Analyser
Name:
Pri: 630
KF: 1.000
Unit: mg/ dl
Lag: 0
Blk: N
Std: N
Limit Set:
Mode: END
,Sec: 0
ESC
ADD
SAVE
QC: N
,Read: 0
Normal: N
Factor: 1.000
Name:
Mode: END
Pri: 630
,Sec: 0
KF: 1.000
Unit: No_Unit
QC: N
Lag: 0
,Read: 0
Blk: N
Normal: N
Std: N
Factor: 1.000
Limit Set:
Init. OD: 0.000
Max Delta: 0.000
Rgnt. Linearity: 0.000
USER MANUAL
Name:
Pri: 630
KF: 1.000
Unit: mg/ dl
Lag: 0
Blk: N
Std: N
Normal
Mode: END
,Sec: 0
QC: N
,Read: 0
Normal: N
Factor: 1.000
Enter Normal Values_
ESC
ADD
SAVE
Low: 0.000
High: 0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ESC
ENTR
CLR
Init. OD: 1.000
Max Delta: 0.000
Rgnt. Linearity: 0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
Init. OD: 0.000
Max Delta: 0.000
Rgnt. Linearity: 0.000
Init. OD: 1.000
Max Delta: 1.500
Rgnt. Linearity: 0.000
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
YES
NO
ENTR
CLR
Numeric Value:
1 2 3 4 5 6
7 8 9 0 .
ENTR
CLR
Enter Reagent Linearity in the Numeric screen.
Touch “ENTR”, upon completion.
Touch “CLR”, to clear a wrong entry.
After entering all the parameters in ENDPOINT mode, touch the SAVE option present on the screen.
Version No. : 1.005A
Effective Date: May 2012
Page 50 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
Ø Recalling /Running (Operating) a pre-programmed test .
For example:
Name: ABC
Mode: END
Pri: 630
,Sec: 0
KF: 2.000
Unit: mg/ dl
QC: Y
Lag: 0
,Read: 0
Blk: N
Normal: Y
Std: N
Factor: 1.000
Limit Set:
To change values press EDIT
Test Name: END
ESC
PRINT
EDIT
RUN
QCR
Set temperature: 37.00
Channel temperature: 30.43
ESC to abort temperature
Test Name: END
Test Name: END
ESC
PID
RUN
RunQC
Referencing.........
Lag Time: 5
Close the Lid & Press Run
Test Name: END
Factor: 1.000
ESC
PID
RUN
RunQC
Read Sample
Version No. : 1.005A
Effective Date: May 2012
ESC
PID
RUN
RUN QC
Page 51 of 54
ESC
PID
RUN
RUNQC
FOURCLOT … Coagulation Analyser
USER MANUAL
10. Trouble Shooting:
MESSAGES
CAUSE /CORRECTIVE ACTION
Remove Cuvette
This error will come in coagulation mode while
referencing. Remove the cuvette.
Invalid Assay
In Multi standard mode if standard are not
proper then this error will appear. Check the
calibrators and rerun the test.
Memory Full
If Number of saved tests exceeds the memory
limit then delete the unwanted tests and save
the test.
Version No. : 1.005A
Effective Date: May 2012
Page 52 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
11. Decontamination
11.1. Decontamination Procedure
If the instrument is to be shipped after being exposed to potentially hazardous material, it should be
decontaminated. The following procedure outlines the method of decontaminating the instrument
before packaging and shipment.
11.2. Purpose of Decontamination
Decontamination minimizes the risk to all who come in contact with the instrument during shipping,
handling, and servicing.
11.3. General Considerations
Any laboratory instrument that has been used for clinical analysis is considered a biohazard and
should be decontaminated prior to handling. Intact skin is generally considered an effective barrier
against infectious Organisms; however, small abrasions and cuts may not be always be visible.
Prophylactic gloves must be worn when handling instruments that have not been decontaminated.
Gloved hands should be considered contaminated at all times and must be kept away from eyes,
mouth and nose at all times.
Mucous membranes are considered prime entry routes for infectious agents. Wear eye protection
and a surgical mask when there is a possibility of aerosols.
Eating and drinking while decontaminating instruments is not advisable.
11.4. Procedure
A solution of .5% Sodium Hypo Chlorite (NaOCL) solution (Bleach) is used. Commercial bleach is
5% NaOCL; household bleach is 3% NaOCL. When using commercial bleach, use a 10:1 mixture; if
using household bleach, a 6:1 mixture is required. This is a caustic solution. It is important to wear
gloves and eye protection when handling it.
Wipe down the carrier and all exposed surfaces of the unit with the bleach solution. Remove the top
shroud of the instrument and wipe down the top surface of the instrument base, as well as the inside
of the top shroud.
Reassemble the unit and discard the used gloves and towels.
Version No. : 1.005A
Effective Date: May 2012
Page 53 of 54
FOURCLOT … Coagulation Analyser
USER MANUAL
12. SAFETY CLEARANCE CERTIFICATE
Please complete all information requests on this form prior to returning the instrument to the manufacturer or
your local distributor for servicing, repairs or return. Thank you for your co-operation.
Customer
Contact
Address
Position
Dept
Tel:
Country
Fax:
Post Code
Model No.
Serial No.
Accessories
Returned
Date of Purchase (if known)
Complaint
Has the equipment been exposed to any of the following:
(*delete as applicable)
a) Blood, body fluids, pathological specimens
If YES, please specify
*YES/NO
b) Other Biohazard
if YES, Please specify
*YES/NO
Version No. : 1.005A
Effective Date: May 2012
Page 54 of 54