4.20 Multiple receivers
P-259
3 p1 ph1
d0
p0 ph2
INDEX
go=2 ph31
# 1 "mc_line 20 file expanding mc command in line "
MCWRK wr #0 if #0 zd ip1 id0
,1'(;
'21(
lo to LBLF1 times td1
# 21 ""
exit
ph1=0 2 2 0 1 3 3 1
ph2=0 2 0 2 1 3 1 3
ph31=0 2 2 0 1 3 3 1
4.20 Multiple receivers
If a spectrometer is equipped with multiple receivers, the number of the receiver
(1-8) from where data should be acquired can be appended to the following commands, e.g. go5=label (if no number is present, 1 is assumed, e.g. go is equivalent
to go1):
go, gonp, gosc, goscnp, adc, rcyc, rcycnp, eosc, eoscnp, ze, zd, st, st0, aq, dw,
dwov, recph, wr, if.
Parameters for the n’th RCU are taken from data set n of the data set list DSLIST.
The following parameters are relevant:
AQ_mod, DECIM, DIGMOD, DIGTYP, DR, DSPFIRM, DSPFVS, FTLPGN,
NBL, OVERFLW, SEOUT, SFO1, SW, SW_h, TD.
4.21 Real time outputs
The TCU provides a number of real time outputs which are used to control the various spectrometer components, such as gating and blanking the transmitters. Please
refer to your hardware documentation for information on which output is connected to a particular device. The pulse program compiler will select the correct
output automatically, e.g. for a command like p1:f2.